Skip to content

Halyard MINOR PACK. Model/Catalog Number: VASF105-08 recallZ-0331-2027 · AVID Medical, Inc. · initiated Sep 8, 2026

Class II Open, Classified All recalls posted in November 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from AVID Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for open seals on header bags, compromising sterility.

FDA root cause category: Process control

What the company is doing

AVID Medical notified consignees on about 09/08/2026 via emailed "URGENT: MEDICAL DEVICE CORRECTION" letter. Consignees were instructed to Segregate, Quarantine, and Inspect all kits identified. If the header bag seal appears to be compromised, discard the entire kit. Consignees were also requested to complete and return the provided Recall Response Form. Distributors were instructed to notify customers if affected kits were further distributed. On 10/06/2026 AVID Medical sent an updated "URGENT: MEDICAL DEVICE CORRECTION" letter to consignees via email. The strategy was updated and consignees were instructed not to use the affected kits and that they should either discard or return them. If returning, they should contact customer service for a Return Goods Authorization. Consignees were again requested to complete and return the provided Recall Response Form. Distributors were again instructed to notify customers if affected kits were further distributed.

Recall details

Recall number
Z-0331-2027
Recall class
Class II
Status
Open, Classified
Initiated by the company
Sep 8, 2026
Posted by the FDA
Nov 3, 2026
Recalling firm
AVID Medical, Inc.
Quantity
108 kits
Product code
LRO
Distribution
U.S Nationwide distributor in the state of Illinois.

Product description

Halyard MINOR PACK. Model/Catalog Number: VASF105-08.

Code information (models, lots, serial numbers)

Model/Catalog Number: VASF105-08. UDI-DI: 10809160451960. Lot Number: 1659384. Expiration Date: 9/30/2027.