Generic Name: Edwards Commander M Delivery System recallZ-0287-2027 · Edwards Lifesciences, LLC · initiated Oct 5, 2026

Class I Open, Classified All recalls posted in November 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Edwards Lifesciences, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to a manufacturing issue which may result in cardiovascular balloon rupturing.

FDA root cause category: Process change control

What the company is doing

On 10/05/2026, an "URGENT: MEDICAL DEVICE RECALL (PRODUCT REMOVAL)" Letter was sent via various delivery methods (e.g., electronic delivery (e-mail), FedEx Priority Overnight mail, or on-site visit) to customers informing them that, due to complaints certain manufacturing lots of the SAPIEN M3 Commander M delivery systems potentially contain a manufacturing defect, which may result in balloon burst during use. Customers are instructed to: " Review this notice carefully and share it with the appropriate clinical staff at their site. " Immediately examine their inventory and quarantine units subject to this recall. Do not use these affected units. Units with other lot numbers are not impacted by this recall and can continue to be used. " Complete the attached Customer Acknowledgment Form and return it to Edwards TMTT Representative or via email to [email protected] " Contact Customer Service at 1-800-424-3278 or email: [email protected] to obtain a Returned Goods Authorization (RGA) number and return any impacted product to Edwards. Edwards asks that customers complete the last two actions listed above within 15 days of receipt of this notification. Customers' assistance is crucial to ensuring this recall is executed thoroughly and promptly. This notice has been communicated by Edwards to the appropriate Regulatory Authorities. For questions and further assistance, contact an Edwards representative who can assist with product return.

Recall details

Recall number
Z-0287-2027
Recall class
Class I (most serious)
Status
Open, Classified
Initiated by the company
Oct 5, 2026
Posted by the FDA
Nov 2, 2026
Recalling firm
Edwards Lifesciences, LLC
Quantity
164
Product code
NPU
FDA submissions
P250019
Distribution
U.S.: AZ, CA, CO, FL, GA, IL, KS, LA, MA, MI, MN, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, UT, VA and WA. O.U.S.: N/A

Product description

Generic Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery System Model/Catalog Number: 9880CM29A Software Version: Not applicable. Product Description: Prosthesis, mitral valve, percutaneously delivered Component: No

Code information (models, lots, serial numbers)

Lot Code: Model No.: 9880CM29A; UDI-DI: 00690103219777; Recalled Lot Number(s): 67718561, 67804014, 67887043, 67887044.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under P250019 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page