Model Number: 712010 recallZ-0288-2027 · Philips North America · initiated Sep 18, 2026

Class II Open, Classified All recalls posted in October 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for the device to fail immediately after an exposure, causing the system to remain stationary.

FDA root cause category: Software design

What the company is doing

Philips Medical Systems notified consignees on about 09/18/2026 via "URGENT Medical Device Correction" letter. Consignees were instructed that they may continue to use the affected system in accordance with its intended use, however if the no-drive issue occurs and prevents repositioning of the system, system movement may be restored through one or more of the following options: - Option 1: System reboot: Power OFF the unit, wait at least 10 seconds, and then power system ON While system startup may take up to 4 minutes, drive functionality will be available within approximately 8 seconds after the unit is powered on. Confirm that drive functionality has been restored before continuing clinical use. - Option 2: Use alternate registered RFID card: Use a different registered RFID card than the one most recently used to log into the system (for example, the backup RFID card kept with the system). Hold the RFID card over the RFID reader until an audible beep is heard (typically within 1 to 3 seconds). Confirm that drive functionality has been restored before continuing clinical use. - Option 3: Manually unlock the wheels using tools provided on the system to free the movement of the system. Consignees were requested to circulate the notification all users of the device for their awareness, retain the notification with the system until the correction is complete and complete and return the provided response form. Philips will be implementing a software correction to address the issue. A Philips representative will contact consignees to schedule implementation when the correction becomes available. If you have any questions, contact the Customer Care Solution Center at 800-722-9377.

Recall details

Recall number
Z-0288-2027
Recall class
Class II
Status
Open, Classified
Initiated by the company
Sep 18, 2026
Posted by the FDA
Oct 26, 2026
Recalling firm
Philips North America
Quantity
563 units
Product code
IZL
FDA submissions
K233678
Distribution
Worldwide Distribution: (US Nationwide) and International distribution to Angola, Australia, Austria, Canada, Chile, Cook Islands, Finland, France, Germany, Ghana, Gibraltar, Guadeloupe, Honduras, Israel, Italy, Latvia, Lithuania, Macedonia, Namibia, Netherlands, New Zealand, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Switzerland, United Arab Emirates, United Kingdom.

Product description

Model Number: 712010.

Code information (models, lots, serial numbers)

Show the full code information (18k characters)

Model Number: 712010. UDI: (01)00884838115521(21)26000155, (01)00884838115521(21)26000154, (01)00884838115521(21)26000151, (01)00884838115521(21)26000150, (01)00884838115521(21)26000149, (01)00884838115521(21)26000148, (01)00884838115521(21)26000147, (01)00884838115521(21)26000146, (01)00884838115521(21)26000145, (01)00884838115521(21)26000144, (01)00884838115521(21)26000143, (01)00884838115521(21)26000142, (01)00884838115521(21)26000141, (01)00884838115521(21)26000140, (01)00884838115521(21)26000139, (01)00884838115521(21)26000138, (01)00884838115521(21)26000137, (01)00884838115521(21)26000136, (01)00884838115521(21)26000135, (01)00884838115521(21)26000134, (01)00884838115521(21)26000133, (01)00884838115521(21)26000132, (01)00884838115521(21)26000129, (01)00884838115521(21)26000125, (01)00884838115521(21)26000126, (01)00884838115521(21)26000131, (01)00884838115521(21)26000130, (01)00884838115521(21)26000124, (01)00884838115521(21)26000123, (01)00884838115521(21)26000122, (01)00884838115521(21)26000121, (01)00884838115521(21)26000120, (01)00884838115521(21)26000119, (01)00884838115521(21)26000118, (01)00884838115521(21)26000116, (01)00884838115521(21)26000115, (01)00884838115521(21)26000113, (01)00884838115521(21)26000110, (01)00884838115521(21)26000109, (01)00884838115521(21)26000108, (01)00884838115521(21)26000107, (01)00884838115521(21)26000106, (01)00884838115521(21)26000117, (01)00884838115521(21)26000114, (01)00884838115521(21)26000111, (01)00884838115521(21)26000112, (01)00884838115521(21)26000105, (01)00884838115521(21)26000104, (01)00884838115521(21)26000103, (01)00884838115521(21)26000102, (01)00884838115521(21)26000090, (01)00884838115521(21)26000089, (01)00884838115521(21)26000088, (01)00884838115521(21)26000087, (01)00884838115521(21)26000086, (01)00884838115521(21)26000085, (01)00884838115521(21)26000084, (01)00884838115521(21)26000101, (01)00884838115521(21)26000095, (01)00884838115521(21)26000094, (01)00884838115521(21)26000093, (01)00884838115521(21)26000092, (01)00884838115521(21)26000100, (01)00884838115521(21)26000099, (01)00884838115521(21)26000098, (01)00884838115521(21)26000097, (01)00884838115521(21)26000096, (01)00884838115521(21)26000091, (01)00884838115521(21)26000080, (01)00884838115521(21)26000079, (01)00884838115521(21)26000083, (01)00884838115521(21)26000082, (01)00884838115521(21)26000081, (01)00884838115521(21)26000078, (01)00884838115521(21)26000076, (01)00884838115521(21)26000073, (01)00884838115521(21)26000070, (01)00884838115521(21)26000067, (01)00884838115521(21)26000048, (01)00884838115521(21)26000047, (01)00884838115521(21)26000046, (01)00884838115521(21)26000045, (01)00884838115521(21)26000074, (01)00884838115521(21)26000058, (01)00884838115521(21)26000057, (01)00884838115521(21)26000056, (01)00884838115521(21)26000055, (01)00884838115521(21)26000044, (01)00884838115521(21)26000043, (01)00884838115521(21)26000041, (01)00884838115521(21)26000040, (01)00884838115521(21)26000072, (01)00884838115521(21)26000071, (01)00884838115521(21)26000075, (01)00884838115521(21)26000065, (01)00884838115521(21)26000064, (01)00884838115521(21)26000068, (01)00884838115521(21)26000063, (01)00884838115521(21)26000006, (01)00884838115521(21)26000005, (01)00884838115521(21)26000004, (01)00884838115521(21)26000003, (01)00884838115521(21)26000002, (01)00884838115521(21)26000001, (01)00884838115521(21)26000042, (01)00884838115521(21)26000062, (01)00884838115521(21)26000061, (01)00884838115521(21)26000060, (01)00884838115521(21)26000059, (01)00884838115521(21)26000037, (01)00884838115521(21)26000054, (01)00884838115521(21)26000053, (01)00884838115521(21)26000052, (01)00884838115521(21)26000051, (01)00884838115521(21)26000050, (01)00884838115521(21)26000049, (01)00884838115521(21)26000038, (01)00884838115521(21)26000036, (01)00884838115521(21)26000035, (01)00884838115521(21)26000034, (01)00884838115521(21)26000032, (01)00884838115521(21)26000029, (01)00884838115521(21)26000028, (01)00884838115521(21)26000033, (01)00884838115521(21)26000031, (01)00884838115521(21)26000027, (01)00884838115521(21)26000026, (01)00884838115521(21)26000025, (01)00884838115521(21)26000020, (01)00884838115521(21)26000024, (01)00884838115521(21)26000023, (01)00884838115521(21)26000022, (01)00884838115521(21)26000021, (01)00884838115521(21)26000018, (01)00884838115521(21)26000019, (01)00884838115521(21)26000017, (01)00884838115521(21)26000016, (01)00884838115521(21)26000015, (01)00884838115521(21)26000014, (01)00884838115521(21)26000013, (01)00884838115521(21)26000012, (01)00884838115521(21)26000009, (01)00884838115521(21)26000008, (01)00884838115521(21)26000007, (01)00884838115521(21)26000011, (01)00884838115521(21)26000010, (01)00884838115521(21)25000233, (01)00884838115521(21)25000232, (01)00884838115521(21)25000234, (01)00884838115521(21)25000231, (01)00884838115521(21)25000230, (01)00884838115521(21)25000229, (01)00884838115521(21)25000223, (01)00884838115521(21)25000216, (01)00884838115521(21)25000228, (01)00884838115521(21)25000227, (01)00884838115521(21)25000222, (01)00884838115521(21)25000221, (01)00884838115521(21)25000226, (01)00884838115521(21)25000225, (01)00884838115521(21)25000224, (01)00884838115521(21)25000220, (01)00884838115521(21)25000219, (01)00884838115521(21)25000215, (01)00884838115521(21)25000213, (01)00884838115521(21)25000218, (01)00884838115521(21)25000217, (01)00884838115521(21)25000214, (01)00884838115521(21)25000210, (01)00884838115521(21)25000209, (01)00884838115521(21)25000212, (01)00884838115521(21)25000211, (01)00884838115521(21)25000208, (01)00884838115521(21)25000205, (01)00884838115521(21)25000206, (01)00884838115521(21)25000204, (01)00884838115521(21)25000207, (01)00884838115521(21)25000203, (01)00884838115521(21)25000202, (01)00884838115521(21)25000201, (01)00884838115521(21)25000200, (01)00884838115521(21)25000197, (01)00884838115521(21)25000199, (01)00884838115521(21)25000198, (01)00884838115521(21)25000187, (01)00884838115521(21)25000186, (01)00884838115521(21)25000195, (01)00884838115521(21)25000194, (01)00884838115521(21)25000193, (01)00884838115521(21)25000192, (01)00884838115521(21)25000191, (01)00884838115521(21)25000190, (01)00884838115521(21)25000189, (01)00884838115521(21)25000181, (01)00884838115521(21)25000188, (01)00884838115521(21)25000185, (01)00884838115521(21)25000184, (01)00884838115521(21)25000182, (01)00884838115521(21)25000180, (01)00884838115521(21)25000183, (01)00884838115521(21)25000174, (01)00884838115521(21)25000179, (01)00884838115521(21)25000176, (01)00884838115521(21)25000168, (01)00884838115521(21)25000178, (01)00884838115521(21)25000177, (01)00884838115521(21)25000175, (01)00884838115521(21)25000166, (01)00884838115521(21)25000173, (01)00884838115521(21)25000172, (01)00884838115521(21)25000171, (01)00884838115521(21)25000169, (01)00884838115521(21)25000167, (01)00884838115521(21)25000170, (01)00884838115521(21)25000165, (01)00884838115521(21)25000164, (01)00884838115521(21)25000163, (01)00884838115521(21)25000162, (01)00884838115521(21)25000160, (01)00884838115521(21)25000159, (01)00884838115521(21)25000158, (01)00884838115521(21)25000157, (01)00884838115521(21)25000156, (01)00884838115521(21)25000155, (01)00884838115521(21)25000161, (01)00884838115521(21)25000154, (01)00884838115521(21)25000153, (01)00884838115521(21)25000152, (01)00884838115521(21)25000151, (01)00884838115521(21)25000146, (01)00884838115521(21)25000150, (01)00884838115521(21)25000148, (01)00884838115521(21)25000149, (01)00884838115521(21)25000147, (01)00884838115521(21)25000145, (01)00884838115521(21)25000142, (01)00884838115521(21)25000121, (01)00884838115521(21)25000141, (01)00884838115521(21)25000144, (01)00884838115521(21)25000143, (01)00884838115521(21)25000140, (01)00884838115521(21)25000139, (01)00884838115521(21)25000133, (01)00884838115521(21)25000135, (01)00884838115521(21)25000134, (01)00884838115521(21)25000130, (01)00884838115521(21)25000138, (01)00884838115521(21)25000132, (01)00884838115521(21)25000137, (01)00884838115521(21)25000126, (01)00884838115521(21)25000136, (01)00884838115521(21)25000131, (01)00884838115521(21)25000120, (01)00884838115521(21)25000128, (01)00884838115521(21)25000127, (01)00884838115521(21)25000129, (01)00884838115521(21)25000125, (01)00884838115521(21)25000124, (01)00884838115521(21)25000122, (01)00884838115521(21)25000123, (01)00884838115521(21)25000119, (01)00884838115521(21)25000115, (01)00884838115521(21)25000114, (01)00884838115521(21)25000118, (01)00884838115521(21)25000112, (01)00884838115521(21)25000110, (01)00884838115521(21)25000116, (01)00884838115521(21)25000113, (01)00884838115521(21)25000109, (01)00884838115521(21)25000107, (01)00884838115521(21)25000106, (01)00884838115521(21)25000105, (01)00884838115521(21)25000104, (01)00884838115521(21)25000103, (01)00884838115521(21)25000102, (01)00884838115521(21)25000101, (01)00884838115521(21)25000117, (01)00884838115521(21)25000100, (01)00884838115521(21)25000099, (01)00884838115521(21)25000098, (01)00884838115521(21)25000097, (01)00884838115521(21)25000096, (01)00884838115521(21)25000095, (01)00884838115521(21)25000111, (01)00884838115521(21)25000108, (01)00884838115521(21)25000077, (01)00884838115521(21)25000094, (01)00884838115521(21)25000093, (01)00884838115521(21)25000092, (01)00884838115521(21)25000091, (01)00884838115521(21)25000090, (01)00884838115521(21)25000089, (01)00884838115521(21)25000088, (01)00884838115521(21)25000087, (01)00884838115521(21)25000084, (01)00884838115521(21)25000083, (01)00884838115521(21)25000082, (01)00884838115521(21)25000086, (01)00884838115521(21)25000085, (01)00884838115521(21)25000081, (01)00884838115521(21)25000080, (01)00884838115521(21)25000079, (01)00884838115521(21)25000078, (01)00884838115521(21)25000076, (01)00884838115521(21)25000075, (01)00884838115521(21)25000074, (01)00884838115521(21)25000073, (01)00884838115521(21)25000072, (01)00884838115521(21)25000071, (01)00884838115521(21)25000068, (01)00884838115521(21)25000067, (01)00884838115521(21)25000066, (01)00884838115521(21)25000061, (01)00884838115521(21)25000070, (01)00884838115521(21)25000069, (01)00884838115521(21)25000065, (01)00884838115521(21)25000064, (01)00884838115521(21)25000063, (01)00884838115521(21)25000062, (01)00884838115521(21)25000060, (01)00884838115521(21)25000058, (01)00884838115521(21)25000057, (01)00884838115521(21)25000050, (01)00884838115521(21)25000056, (01)00884838115521(21)25000055, (01)00884838115521(21)25000054, (01)00884838115521(21)25000053, (01)00884838115521(21)25000052, (01)00884838115521(21)25000051, (01)00884838115521(21)25000044, (01)00884838115521(21)25000059, (01)00884838115521(21)25000022, (01)00884838115521(21)25000049, (01)00884838115521(21)25000048, (01)00884838115521(21)25000047, (01)00884838115521(21)25000045, (01)00884838115521(21)25000046, (01)00884838115521(21)25000042, (01)00884838115521(21)25000038, (01)00884838115521(21)25000043, (01)00884838115521(21)25000041, (01)00884838115521(21)25000040, (01)00884838115521(21)25000039, (01)00884838115521(21)25000033, (01)00884838115521(21)25000031, (01)00884838115521(21)25000030, (01)00884838115521(21)25000009, (01)00884838115521(21)25000037, (01)00884838115521(21)25000036, (01)00884838115521(21)25000035, (01)00884838115521(21)25000034, (01)00884838115521(21)25000032, (01)00884838115521(21)25000018, (01)00884838115521(21)25000020, (01)00884838115521(21)25000028, (01)00884838115521(21)25000027, (01)00884838115521(21)25000016, (01)00884838115521(21)25000023, (01)00884838115521(21)25000005, (01)00884838115521(21)25000029, (01)00884838115521(21)25000026, (01)00884838115521(21)25000025, (01)00884838115521(21)25000024, (01)00884838115521(21)25000019, (01)00884838115521(21)25000017, (01)00884838115521(21)25000012, (01)00884838115521(21)25000006, (01)00884838115521(21)25000002, (01)00884838115521(21)25000015, (01)00884838115521(21)25000014, (01)00884838115521(21)24000209, (01)00884838115521(21)25000021, (01)00884838115521(21)25000013, (01)00884838115521(21)25000011, (01)00884838115521(21)25000003, (01)00884838115521(21)25000001, (01)00884838115521(21)25000007, (01)00884838115521(21)25000008, (01)00884838115521(21)25000004, (01)00884838115521(21)25000010, (01)00884838115521(21)24000198, (01)00884838115521(21)24000202, (01)00884838115521(21)24000199, (01)00884838115521(21)24000211, (01)00884838115521(21)24000210, (01)00884838115521(21)24000203, (01)00884838115521(21)24000208, (01)00884838115521(21)24000206, (01)00884838115521(21)24000205, (01)00884838115521(21)24000204, (01)00884838115521(21)24000197, (01)00884838115521(21)24000179, (01)00884838115521(21)24000178, (01)00884838115521(21)24000177, (01)00884838115521(21)24000213, (01)00884838115521(21)24000188, (01)00884838115521(21)24000187, (01)00884838115521(21)24000185, (01)00884838115521(21)24000212, (01)00884838115521(21)24000194, (01)00884838115521(21)24000193, (01)00884838115521(21)24000191, (01)00884838115521(21)24000190, (01)00884838115521(21)24000172, (01)00884838115521(21)24000171, (01)00884838115521(21)24000169, (01)00884838115521(21)24000192, (01)00884838115521(21)24000180, (01)00884838115521(21)24000174, (01)00884838115521(21)24000173, (01)00884838115521(21)24000170, (01)00884838115521(21)24000183, (01)00884838115521(21)24000182, (01)00884838115521(21)24000184, (01)00884838115521(21)24000195, (01)00884838115521(21)24000175, (01)00884838115521(21)24000189, (01)00884838115521(21)24000181, (01)00884838115521(21)24000163, (01)00884838115521(21)24000167, (01)00884838115521(21)24000162, (01)00884838115521(21)24000161, (01)00884838115521(21)24000160, (01)00884838115521(21)24000166, (01)00884838115521(21)24000176, (01)00884838115521(21)24000164, (01)00884838115521(21)24000158, (01)00884838115521(21)24000155, (01)00884838115521(21)24000157, (01)00884838115521(21)24000118, (01)00884838115521(21)24000168, (01)00884838115521(21)24000165, (01)00884838115521(21)24000154, (01)00884838115521(21)24000153, (01)00884838115521(21)24000152, (01)00884838115521(21)24000151, (01)00884838115521(21)24000156, (01)00884838115521(21)24000144, (01)00884838115521(21)24000138, (01)00884838115521(21)24000149, (01)00884838115521(21)24000148, (01)00884838115521(21)24000146, (01)00884838115521(21)24000145, (01)00884838115521(21)24000142, (01)00884838115521(21)24000126, (01)00884838115521(21)24000125, (01)00884838115521(21)24000150, (01)00884838115521(21)24000147, (01)00884838115521(21)24000143, (01)00884838115521(21)24000141, (01)00884838115521(21)24000140, (01)00884838115521(21)24000139, (01)00884838115521(21)24000137, (01)00884838115521(21)24000133, (01)00884838115521(21)24000131, (01)00884838115521(21)24000130, (01)00884838115521(21)24000123, (01)00884838115521(21)24000107, (01)00884838115521(21)24000119, (01)00884838115521(21)24000129, (01)00884838115521(21)24000122, (01)00884838115521(21)24000117, (01)00884838115521(21)24000115, (01)00884838115521(21)24000114, (01)00884838115521(21)24000112, (01)00884838115521(21)24000111, (01)00884838115521(21)24000110, (01)00884838115521(21)24000086, (01)00884838115521(21)24000116, (01)00884838115521(21)24000109, (01)00884838115521(21)24000090, (01)00884838115521(21)24000108, (01)00884838115521(21)24000104, (01)00884838115521(21)24000097, (01)00884838115521(21)24000105, (01)00884838115521(21)24000106, (01)00884838115521(21)24000102, (01)00884838115521(21)24000101, (01)00884838115521(21)24000100, (01)00884838115521(21)24000099, (01)00884838115521(21)24000089, (01)00884838115521(21)24000087, (01)00884838115521(21)24000103, (01)00884838115521(21)24000098, (01)00884838115521(21)24000085, (01)00884838115521(21)24000088, (01)00884838115521(21)24000080, (01)00884838115521(21)24000079, (01)00884838115521(21)24000078, (01)00884838115521(21)24000077, (01)00884838115521(21)24000096, (01)00884838115521(21)24000095, (01)00884838115521(21)24000094, (01)00884838115521(21)24000084, (01)00884838115521(21)24000093, (01)00884838115521(21)24000092, (01)00884838115521(21)24000091, (01)00884838115521(21)24000083, (01)00884838115521(21)24000082, (01)00884838115521(21)24000081, (01)00884838115521(21)24000069, (01)00884838115521(21)24000076, (01)00884838115521(21)24000075, (01)00884838115521(21)24000074, (01)00884838115521(21)24000072, (01)00884838115521(21)24000073, (01)00884838115521(21)24000071, (01)00884838115521(21)24000065, (01)00884838115521(21)24000057, (01)00884838115521(21)24000068, (01)00884838115521(21)24000067, (01)00884838115521(21)24000064, (01)00884838115521(21)24000063, (01)00884838115521(21)24000062, (01)00884838115521(21)24000061, (01)00884838115521(21)24000060, (01)00884838115521(21)24000059, (01)00884838115521(21)24000053, (01)00884838115521(21)24000058, (01)00884838115521(21)24000052, (01)00884838115521(21)24000066, (01)00884838115521(21)24000055, (01)00884838115521(21)24000054, (01)00884838115521(21)24000051, (01)00884838115521(21)24000049, (01)00884838115521(21)24000056, (01)00884838115521(21)24000034, (01)00884838115521(21)24000035, (01)00884838115521(21)24000050, (01)00884838115521(21)24000033, (01)00884838115521(21)24000048, (01)00884838115521(21)24000044, (01)00884838115521(21)24000041, (01)00884838115521(21)24000040, (01)00884838115521(21)24000043, (01)00884838115521(21)24000042, (01)00884838115521(21)24000036, (01)00884838115521(21)24000032, (01)00884838115521(21)24000031, (01)00884838115521(21)24000030, (01)00884838115521(21)24000038, (01)00884838115521(21)24000037, (01)00884838115521(21)24000047, (01)00884838115521(21)24000046, (01)00884838115521(21)24000045, (01)00884838115521(21)24000039, (01)00884838115521(21)24000028, (01)00884838115521(21)24000025, (01)00884838115521(21)24000027, (01)00884838115521(21)24000026, (01)00884838115521(21)24000020, (01)00884838115521(21)24000029, (01)00884838115521(21)24000014, (01)00884838115521(21)24000024, (01)00884838115521(21)24000023, (01)00884838115521(21)24000022, (01)00884838115521(21)24000021, (01)00884838115521(21)24000019, (01)00884838115521(21)24000018, (01)00884838115521(21)24000015, (01)00884838115521(21)24000013, (01)00884838115521(21)24000017, (01)00884838115521(21)24000016.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page