Cochlear Nucleus Implant recallZ-0349-2027 · Cochlear Limited · initiated Sep 30, 2026
Class II Open, Classified All recalls posted in November 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cochlear Limited or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a manufacturing deviation of incomplete electrode welds of cochlear implants and resulting in one or more open circuits
FDA root cause category: Under Investigation by firm
What the company is doing
On 09/30/2026, an "URGENT Field Safety Corrective Action Notification" was sent to U.S. customers informing them that a potential condition has been identified that could result in one or more open circuits affecting individual electrode pathways with the implant. Customers were instucted to not implant and return affected products to Cochlear Limited. For questions contact [email protected].
Recall details
- Recall number
- Z-0349-2027
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Sep 30, 2026
- Posted by the FDA
- Nov 3, 2026
- Recalling firm
- Cochlear Limited
- Quantity
- 4 device in the U.S.
- Product code
- MCM
- FDA submissions
- P970051S225
- Distribution
- U.S.: NJ, NY, and PA. O.U.S.: Not provided
Product description
Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, single use devices intended for long term implantation under the skin in the mastoid region on either side of the head. The cochlear implant system works by converting sound in the environment into electrical code. It transmits this code to the auditory nerve and on to the brain where it is interpreted as sound. The Cochlear Implant System consists of internal (implanted) and external components. The implanted component consists of a receiver/stimulator and an electrode array. The external components consist of a sound processor and their associated transmitting coils and cables and must be worn for the implant to work.
Code information (models, lots, serial numbers)
Model: CI1000 Product/Catalog: P1979576 UDI-DI code: 9321502069847 Serial Numbers: 1020210051972 1020210070461 1020210145046 1020210048286
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Cochlear® Nucleus® Nexa™ cochlear implant CI1012Cochlear Limited · Implant, cochlear1 recall32 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Nov 27, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Nov 27, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Nov 27, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page