Yeastone Broth, 11ML, 10/BOX, used in antibiotic… recallZ-0355-2027 · Remel, Inc · initiated Oct 19, 2026
Class II Open, Classified All recalls posted in November 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Remel, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Affected lot did not meet specifications. Product may report false susceptibility to antifungal agents, specifically within the Azole family.
FDA root cause category: Under Investigation by firm
What the company is doing
The firm began notifying affected customers on October 19, 2026. Customers were instructed to notify any personnel who needed to be aware of the issue. Customers should review previously reported susceptibility results generated using the affected lot and assess whether retesting is appropriate. Customers should destroy any remaining inventory of the affected lot. Do not use it for any clinical laboratory testing. Customers should complete and return the acknowledgement form accompanying the letter. Firm will replace product or issue credit. If you have any questions, please contact our Technical Services Department at 800-255-6730 (US) or email [email protected]. You should complete the accompanying Acknowledgment Form upon receipt and return to [email protected]
Recall details
- Recall number
- Z-0355-2027
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Oct 19, 2026
- Posted by the FDA
- Nov 3, 2026
- Recalling firm
- Remel, Inc
- Quantity
- 33 US, 54 OUS
- Product code
- NGZ
- FDA submissions
- K090968
- Distribution
- Worldwide - US Nationwide distribution in the states of MA and CA. The countries of India, Saudi Arabia, UK.
Product description
Yeastone Broth, 11ML, 10/BOX, used in antibiotic susceptibility testing of mycological clinical specimens
Code information (models, lots, serial numbers)
YY3462 lot 440400 UDI-DI 848838018817
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NA Y3462Remel, Inc. · Susceptibility Test Plate, Antifungal2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Nov 27, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Nov 27, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Nov 27, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page