Skip to content

Dimension TSPA (Total Prostate Specific Antigen) Flex… recallZ-0873-03 · Dade Behring Inc · initiated Apr 21, 2003

Terminated All recalls posted in May 2003

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Dade Behring Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

test results may be inaccurate by 0.2 to 0.5 ng/mL

FDA root cause category: Other

What the company is doing

The recalling firm telephone each customer on 4/24-28/03 to determine if any customer used the lot (CD4085). If so, the customer was instructed to retest all samples with an alternate lot. All customers were instructed to discard any remaining inventory. As a follow up to the phone calls, the recalling firm issued a recall letter dated 4/25/03 informing the customer of the problem and to discard any remaining product.

Recall details

Recall number
Z-0873-03
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 21, 2003
Posted by the FDA
May 29, 2003
Recalling firm
Dade Behring Inc
Quantity
662 cartons
Product code
LTJ
FDA submissions
P000021
Distribution
The product was shipped to hospitals and clinic laboratories nationwide. There are eight government/military accounts; one of which is foreign.

Product description

Dimension TSPA (Total Prostate Specific Antigen) Flex Reagent Cartridge. The product is shipped 4 cartridges per carton.

Code information (models, lots, serial numbers)

Product code number RF451. Lot number CD4085 exp 3/25/04

Other devices under the same FDA submission

2 devices cleared under P000021 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.