Skip to content

Stanmore modular hip system recallZ-1286-03 · Biomet, Inc. · initiated Jun 18, 2003

Terminated All recalls posted in September 2003

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The labeling does not instruct the removal of the protective shipping cap, and one surgeon implanted with the cap still in place.

FDA root cause category: Other

What the company is doing

Distributors were sent a letter on June 18, 2003 via FedEx. This mailing contained labels and the distributors were instructed to affix the label to the inventory consigned to hospitals within their territory.

Recall details

Recall number
Z-1286-03
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jun 18, 2003
Posted by the FDA
Sep 24, 2003
Recalling firm
Biomet, Inc.
Quantity
28
Product code
JDI
FDA submissions
K911684
Distribution
United States

Product description

Stanmore modular hip system; Stanmore CoCr femoral size 1 straight stem; part 164251

Code information (models, lots, serial numbers)

lots 267824, 292736, 336747, 348666, 351182, 351944, 358183, 361071, 366677, 375324, 376152, 396782, 424534, 424535, 438547, 469956, 497678, 095912

Other devices under the same FDA submission

9 devices cleared under K911684 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 26618).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page