Skip to content

Exacta-Mix 2400 pharmacy compounding system recallZ-0210-04 · Baxa Corporation · initiated Nov 7, 2003

Terminated All recalls posted in July 2004

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxa Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

If device stops due to an alarm or operater pause, inlet valves will remain open, allowing some unwanted mixing of ingredients in the valve chamber..

FDA root cause category: Other

What the company is doing

All users were contacted by phone and fax on 11/7/03 with instructions to safely use the device.

Recall details

Recall number
Z-0210-04
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Nov 7, 2003
Posted by the FDA
Jul 20, 2004
Recalling firm
Baxa Corporation
Quantity
231 systems
Product code
NEP
FDA submissions
K002705
Distribution
Nationwide. Foreign distribution to UK. No government or military consignees.

Product description

Exacta-Mix 2400 pharmacy compounding system.

Code information (models, lots, serial numbers)

Catalog Number EM2400, all serial numbers.

Other devices under the same FDA submission

7 devices cleared under K002705 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page