Skip to content

ABX Diagnostics Minotrol 16 Hematology Controls, Normal and… recallZ-0992-04 · R & D Systems,Inc · initiated Apr 15, 2004

Terminated All recalls posted in July 2004

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from R & D Systems,Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Clots in hematology controls resulted in low red blood cell counts when used in blood analyzers because control values were outside the expected range due to the clotting which took place after manufacturing.

FDA root cause category: Other

What the company is doing

Recall letters dated April 15, 2004 instructed consignees not to use the products to verify their analyzers and stated that the recalling firm is working on providing replacement products.

Recall details

Recall number
Z-0992-04
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 15, 2004
Posted by the FDA
Jul 20, 2004
Recalling firm
R & D Systems,Inc
Quantity
29494 vials of the normal controls and 15196 of the high controls
Product code
JPK
FDA submissions
K843962
Distribution
The products were shipped to California, France and Japan.

Product description

ABX Diagnostics Minotrol 16 Hematology Controls, Normal and High, in vials distributed by ABX Diagnostics, Irvine, CA; and Horiba LC-TROL 16 (Hematology Reference Controls), Normal and High, in vials distributed by Horiba Ltd., Kyoto, Japan

Code information (models, lots, serial numbers)

Lots MX249N and MX249H of ABX Diagnostics Minotrol 16 Controls, Normal and High; and   Lots 040308ND and 040308HD of Horiba LC-TROL 16 (Hematology Reference Controls), Normal and High.

Other devices under the same FDA submission

1 device cleared under K843962 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page