Skip to content

BC Thrombin Reagent . Thrombin Time Test recallZ-0007-05 · Dade Behring , Inc. · initiated Jul 21, 2004

Terminated All recalls posted in October 2004

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Dade Behring , Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The product does not demonstrate the expected prolonged thrombin times for specimens obtained from patients recieving heparin

FDA root cause category: Other

What the company is doing

The recalling firm issued a recall letter dated 7/21/04 to their accounts informing them of the problem and the need to discard the product.

Recall details

Recall number
Z-0007-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jul 21, 2004
Posted by the FDA
Oct 5, 2004
Recalling firm
Dade Behring , Inc.
Quantity
3942 kits
Product code
GJA
FDA submissions
K970645
Distribution
The products were shipped to medical facilities in CT, FL, IA, IL, NJ, PA, and WI. The products were also shipped to Austria, Bosnia, Croatia, Hungry, Germany, Poland, Spain, RussiaFrance, Italy, Australia, Brazil, Canada, Saudi Arabia, El Salvador, and Turkey.

Product description

BC Thrombin Reagent . Thrombin Time Test

Code information (models, lots, serial numbers)

Lot 540624, Lot 540625, and Lot 540626

Other devices under the same FDA submission

1 device cleared under K970645 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates