Skip to content

Brand name recallZ-0310-05 · Medtronic Xomed, Inc. · initiated Oct 29, 2004

Terminated All recalls posted in December 2004

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Xomed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm received six reports of tubes with wires 'out of channel' two of which involved injuried patients. In one case in Germany it was reported that the patient had soreness of the throat and was discharged after four days. In the next case the patient had a tracheal wound. It was not clear what caused the wound.

FDA root cause category: Other

Recall details

Recall number
Z-0310-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 29, 2004
Posted by the FDA
Dec 15, 2004
Recalling firm
Medtronic Xomed, Inc.
Quantity
22569 EMG Tubes
Product code
ETN
FDA submissions
K925640
Distribution
Distributed nationwide and worldwide. to countries in the five continents with the highest populations.

Product description

Brand name: NIM EMG Endotracheal Tube REF 82-29106, 82-29107, 82-29108 NIM EMG Reinforced Endotracheal Tube REF 82-29306, 82-29307, 82-29308 [All products are packaged one unit per box.]

Code information (models, lots, serial numbers)

Product Code: 77 ETN  Lot Numbers: 32423300 through 35505400, and 35542200, 35542300, 35542400, 35542500, 35542600, 35543400, 35597200, 35602100, 35633400, 35633600, 35638200, 35638400

Other devices under the same FDA submission

12 devices cleared under K925640 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page