Skip to content

ACT diff 2 Analyzer recallZ-0339-05 · Beckman Coulter Inc · initiated Dec 3, 2004

Terminated All recalls posted in December 2004

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Erroneous results issue when MONOJECT blood collection tubes are used during analysis after tubes are uncapped and recapped then analysed in the closed vial mode. Results are lower than if cap had not been removed.

FDA root cause category: Other

What the company is doing

A letter issued to users informing them of the incompatability of the Coulter ACT diff 2 Analyzer with the use of Monoject blood collection tubes made by Tyco Healthcare Group/Kendall Healthcare. Customers choosing to use these tubes are instructed to uncap the tubes and analyze the samples,as open vial samples.

Recall details

Recall number
Z-0339-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Dec 3, 2004
Posted by the FDA
Dec 21, 2004
Recalling firm
Beckman Coulter Inc
Quantity
2709
Product code
GKZ
FDA submissions
K990352
Distribution
Nationwide and Canada

Product description

ACT diff 2 Analyzer (Automated differential cell counter)

Code information (models, lots, serial numbers)

All codes

Other devices under the same FDA submission

1 device cleared under K990352 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page