Skip to content

HomeChoice PRO Automated Peritoneal Dialysis Systems recallZ-0397-05 · Baxter Healthcare Renal Div · initiated Dec 23, 2004

Terminated All recalls posted in January 2005

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Renal Div or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A software anomaly may allow concentrated solution to be infused into the patient when the HomeChoice unit is used with Physioneal Clear-Flex PD products.

FDA root cause category: Other

What the company is doing

Baxter notified the Danish Ministry of Health of the problem with the HomeChoice Pro not prompting the patient when using the Physioneal Clear-Flex PD products on 12/23/04. The patients in Denmark and Sweden will be re-evaluate and re-train each patient as needed using enhanced training materials, which will include the importance of and the rationales for breaking the peel-seal on teh Physioneal ClearFlex prior to use and what to do if mis-infusion occurs. Larger, adhesive, colored stickers will be affixed to all applicable HomeChoice Pro machines to further remind patients of the need to break all peel-seals prior to use. Patients will receive a new Pro-Card when their machine is exchanged, and the exchange will be documented. The HomeChoice Pro software will be updated to correct the software anomaly, validated and installed on the HomeChoice Pro machines.

Recall details

Recall number
Z-0397-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Dec 23, 2004
Posted by the FDA
Jan 15, 2005
Recalling firm
Baxter Healthcare Renal Div
Quantity
165 units
Product code
FKX
FDA submissions
K012988
Distribution
Those HomeChoice units in Denmark and Sweden only, where the Physioneal Clear-Flex products are approved for use.

Product description

HomeChoice PRO Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A.; catalog numbers 5C8310, 5C8310R, R5C8320, R5C8320R

Code information (models, lots, serial numbers)

Units with software version 8.7 used with Physioneal Clear-Flex PD products

Other devices under the same FDA submission

1 device cleared under K012988 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page