Skip to content

VERSAPOINT, Hysteroscopic system Connector Cable… recallZ-0493-05 · Ethicon, Inc · initiated Dec 16, 2004

Terminated All recalls posted in February 2005

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ethicon, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Deformation of the Verspoint Hysteroscopic System Connector Cable hand grips when autoclaved causing possible non-sterile electrodes.

FDA root cause category: Nonconforming Material/Component

What the company is doing

The recall letter was mailed to the Director of Material Management and the Director of Operating Room on December 14, 2004.

Recall details

Recall number
Z-0493-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Dec 16, 2004
Posted by the FDA
Feb 2, 2005
Recalling firm
Ethicon, Inc
Quantity
891electrodes
Product code
HIH
FDA submissions
K962482
Distribution
Ethicon distributed the devices to 166 hospitals in the US. There are no Government sales. Product sold to other countries was distributed directly from the manufacturerin England to Ethicon distributors outside the US. The Foreign countries with accounts are: Austria, Czech Republic, Greece, Italy, Benelux, Poland, Portugal, Slovenia, Spain, Sweden, France, Germany, Ireland, U.K. Canada, China, Mexico, Taiwan, Venezuela, Korea, Australia, Brazil, and India.

Product description

VERSAPOINT, Hysteroscopic system Connector Cable, Non-Sterile, Reusable (20 uses only). US Distribution/Manufactured for GYNECARE, A division of Ethicon, Inc., a Johnson & Johnson Company, Somerville, NJ 08876. The product subject of this recall is the connector cable which goes between the electrosurgical generator and the elctrode used for surgery. Specifically, the recall focuses on the hand grip on the electrode, formed by overmolding a polymer onto the cable, which deforms upon sterilization.

Code information (models, lots, serial numbers)

Product Code: 00480; Lot #s:  0404086, 0404121, 0404152, 0405010, 0405123, 0406014, 0406068, 0406130, 0407104, 0408031, 0409029, 0409080, 0410021, 0410103, 0411040.