Deltec branded Administration Sets for use with Deltec 3000… recallZ-0491-05 · Smiths Medical MD, Inc. · initiated Dec 3, 2004
Terminated All recalls posted in February 2005
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical MD, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Administration Sets which are used with selected infusion pump systems may have a defect within the fluid path above the upper chamber of the set cassette which does not allow the pump valve to close completely and can allow fluid to leask past and may cause an inaccurate infusion rate..
FDA root cause category: Other
What the company is doing
Customers were sent a product Safety & Recall notification via email or regular mail begining 12/03/04 through 12/08/04. A Product Recall form requested the customer for feedback as to Quantity Used, Quantity to be Retained and Quantity to be returned . The Customer Recall Notication states that if no Administration Sets from alternative lots are available and the use of the volumetric pump is essential, follow the instructions in the attached Customer Information Bulletin, which describes how to identify a set which does not have the defect. Customers who find product subject to the recall are to contact Smith Medical Customer Service Department to make arrangements to have the affected products returned to Smiths.
Recall details
- Recall number
- Z-0491-05
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Dec 3, 2004
- Posted by the FDA
- Feb 2, 2005
- Recalling firm
- Smiths Medical MD, Inc.
- Quantity
- 17,760 units (444 cases) Internationally. Each case contains 40 units.
- Product code
- FPA
- FDA submissions
- K811933, K896161, K991599
- Distribution
- worldwide.
Product description
Deltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number RC2124 (Reorder No. 21-0441-25) Specialty Administration Set described as 109 inch length/277 cm, 26ml priming volume, 20 drops/ml and 1 Needless Injection site, 15 micron filter.
Code information (models, lots, serial numbers)
FEB04T05, MAR04T30 & APR04T20 distributed Internationally.
Other devices under the same FDA submission
1 device cleared under K811933, K896161, K991599 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CADD 21-7081-24Icu Medical, Inc. · Set, administration, intravascular21 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 30603).
- Z-0464-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number 8C390 (Reorder No. 21-0325-01) 121 in…
- Z-0465-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number 8C220 (Reorder No. 21-0323-01) 107 in…
- Z-0466-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. 3 Model Numbers are used for this same Burette Adm…
- Z-0467-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. 2 Model Numbers are used for this same Specialty A…
- Z-0468-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. 2 Model Numbers are used for this same Burette Adm…
- Z-0469-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number AK2304 (Reorder No. 21-0308-01) 118 i…
- Z-0470-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number CC2102 (Reorder No. 21-0309-01) 112 i…
- Z-0471-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number CJ4303 (Reorder No. 21-0313-01)110 in…
- Z-0472-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. 2 Model Numbers are used for this same Primary Adm…
- Z-0473-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. 2 Model Numbers are used for this same Primary Che…
- Z-0474-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number AC2029 (Reorder No. 21-0330-01) Speci…
- Z-0475-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number CC1201 (Reorder No. 21-0333-01) Prima…
- Z-0476-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number CC1301 (Reorder No. 21-0334-01) Prima…
- Z-0477-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number CC2204 (Reorder No. 21-0336-01) Speci…
- Z-0478-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number CC3217 (Reorder No. 21-0338-01) Speci…
- Z-0479-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number CC4203 (Reorder No. 21-0339-01) 105 i…
- Z-0480-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number CF4203 (Reorder No. 21-0340-01) 105 i…
- Z-0481-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number LC4203 (Reorder No. 21-0341-01) 105 i…
- Z-0482-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number NJ4303 (Reorder No. 21-0426-01) descr…
- Z-0483-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number NK4303 (Reorder No. 21-0427-01) descr…
- Z-0484-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number NC1301 (Reorder No. 21-0428-01) Prima…
- Z-0485-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number NJ2304 (Reorder No. 21-0433-01) Buret…
- Z-0486-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number 8C4209 (Reorder No. 21-0327-01) Prima…
- Z-0487-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number I8C820 (Reorder No. 21-0346-25) Prima…
- Z-0488-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number I8C825 (Reorder No. 21-0347-25) Speci…
- Z-0489-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number I8C4240J (Reorder No. 21-0352-25) Spe…
- Z-0490-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number RC4106 (Reorder No. 21-0402-25) Prima…
- Z-0492-05 · TerminatedDeltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. Model Number RC1201 (Reorder No. 21-0442-25) Prima…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page