Skip to content

Mammotome EX Holster recallZ-0440-05 · Ethicon Endo-Surgery · initiated Dec 10, 2004

Terminated All recalls posted in January 2005

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ethicon Endo-Surgery or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A capacitor of the wrong value was used. The capacitor used was 0.11 microfarad instead of 0.22 microfarad called for in the product specifications.

FDA root cause category: Other

What the company is doing

The recalling firm notified hospitals and distributors of the product by letter, dated 12/10/04.

Recall details

Recall number
Z-0440-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Dec 10, 2004
Posted by the FDA
Jan 27, 2005
Recalling firm
Ethicon Endo-Surgery
Quantity
120 devices
Product code
KNW
FDA submissions
K033700
Distribution
The product was distributed to consignees located in MI, TN, IL, NC, GA, IL, CA, FL, IN, WI, TX, AR, KS, IA, OK, NY, PA, MS, KY, ME, MO, MA, VA, NJ, LA, MN, SD, UT, CO, AL, and NH. There are no foreign consignees.

Product description

Mammotome EX Holster

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

HRDE39904, HRDE40502, HRDE40102, HRDE39920 HRDE40209, HRDE40204, HRDE39916, HRDE40220 HRDE39921, HRDE40114, HRDE40225, HRDE40508 HRDE40210, HRDE40222, HRDE39919, HRDE39906 HRDE40218, HRDE40724, HRDE40109, HRDE39903 HRDE40106, HRDE40118, HRDE40203, HRDE40113 HRDE40511, HRDE40206, HRDE39908, HRDE39910 HRDE40701, HRDE40513, HRDE40713, HRDE40720 HRDE40718, HRDE40710, HRDE40704, HRDE40716 HRDE40706, HRDE40520, HRDE40722, HRDE40708 HRDE40714 HRDE40522, HRDE40719, HRDE40705 HRDE40525, HRDE40527, HRDE40721, HRDE40703 HRDE40516, HRDE40725, HRDE40709, HRDE40702 HRDE40715, HRDE40518, HRDE40506, HRDE40519 HRDE40707, HRDE40712, HRDE40711, HRDE39922 HRDE40103, HRDE39901, HRDE40126, HRDE40120 HRDE39905, HRDE40723, HRDE40101, HRDE40205 HRDE40214, HRDE40526, HRDE40618, HRDE39902 HRDE40202, HRDE40122, HRDE40217, HRDE40221 HRDE40216, HRDE40624, HRDE39924, HRDE40717 HRDE40226, HRDE40503, HRDE40213, HRDE40107 HRDE40110, HRDE40524, HRDE40121, HRDE40514 HRDE39925, HRDE39909, HRDE39912, HRDE40505 HRDE40112, HRDE40212, HRDE40510, HRDE40208 HRDE40201, HRDE39907, HRDE39926, HRDE40124 HRDE40223, HRDE39923, HRDE40219, HRDE40111 HRDE40509, HRDE40515, HRDE40517, HRDE40108 HRDE39918, HRDE40521, HRDE40123, HRDE39914 HRDE39915, HRDE40501, HRDE40211 HRDE40224 HRDE40105, HRDE40207, HRDE39911, HRDE39911,  HRDE 39913

Other devices under the same FDA submission

2 devices cleared under K033700 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page