Skip to content

Cook brand Zenith Flex AAA Endovascular graft with the… recallZ-0512-05 · Cook, Inc. · initiated Dec 20, 2004

Terminated All recalls posted in February 2005

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cook, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm has received complaints of difficulty in releasing the barbed suprarenal stent from the top cap that constrains it, which may result in misplacement of the graft.

FDA root cause category: Other

What the company is doing

Recall letter dated 12/20/04 was issued to each consignee via FedEx instructing them to cease using the products and to return them to the firm.

Recall details

Recall number
Z-0512-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Dec 20, 2004
Posted by the FDA
Feb 3, 2005
Recalling firm
Cook, Inc.
Product code
MIH
FDA submissions
P020018
Distribution
Nationwide and Canada.

Product description

Cook brand Zenith Flex AAA Endovascular graft with the H&L-B One-Shot Introduction System( A= 26mm, B=141mm, C=111mm, D=12mm, E= 18.0f, F=.035'', G =40cm); Reorder number TFBB-26-111.

Code information (models, lots, serial numbers)

All units distributed on or before 12/17/04.

Other devices under the same FDA submission

36 devices cleared under P020018 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 30721).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page