Skip to content

Proteus XR/a Radiographic X-ray System, models 2259973-1… recallZ-0219-05 · General Electric Med Systems LLC · initiated Dec 30, 2004

Terminated All recalls posted in February 2006

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from General Electric Med Systems LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Units have a defect in accordance with 21 CFR 1003.2 because an incorrect exposure can occur.

FDA root cause category: Other

What the company is doing

An 'Urgent: Radiation Safety Notice' letter was issued to hospitals with affected systems describing potential hazards, the planned modification and recommended action to limit the hazard until the planned modification is installed. A GE field Service Representative will install the modification.

Recall details

Recall number
Z-0219-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Dec 30, 2004
Posted by the FDA
Feb 14, 2006
Recalling firm
General Electric Med Systems LLC
Quantity
2716
Product code
KPR
FDA submissions
K993090
Distribution
globally

Product description

Proteus XR/a Radiographic X-ray System, models 2259973-1, 2259973-2, 2259973-3, 2259973-4. The system consists of a radiographic table, a overhead tube suspension, high frequency generator and power distribution unit, wall stand, dual focal spot x-ray tube and operator''s console.

Code information (models, lots, serial numbers)

all serial numbers

Other devices under the same FDA submission

1 device cleared under K993090 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page