Model 740 and 760 series Ventilators recallZ-0445-05 · Nellcor Puritan Bennett Inc · initiated Dec 17, 2004
Terminated All recalls posted in January 2005
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Nellcor Puritan Bennett Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Capacitor weakness in backup alarm system might fail causing lack of backup alarm to function in the event of total power loss.
FDA root cause category: Other
What the company is doing
Customers of record will be notified by letter sent FEDEX, a simple test instruction will be included to use to test the performance of the backup alarm.
Recall details
- Recall number
- Z-0445-05
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Dec 17, 2004
- Posted by the FDA
- Jan 28, 2005
- Recalling firm
- Nellcor Puritan Bennett Inc
- Quantity
- 13,010 worldwide, 3,948 in USA and territories.
- Product code
- CBK
- FDA submissions
- K964540, K984379
- Distribution
- Worldwide.
Product description
Model 740 and 760 series Ventilators.
Code information (models, lots, serial numbers)
All units ever manufactured. The product may be identified and tracked through serial numbers. This is a tracked device.
Other devices under the same FDA submission
1 device cleared under K964540, K984379 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Puritan Bennett G-061949-00Covidien LP · VENTILATOR, CONTINUOUS, FACILITY USE18 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page