Advisor Vital Signs Monitor model 9200, catalog number… recallZ-0532-05 · Smiths Medical PM, Inc. · initiated Jan 5, 2005
Terminated All recalls posted in February 2005
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical PM, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An issue with the pace detect function was discovered on a 9200 II monitor. Could result in a falsely high or low heart rate reading.
FDA root cause category: Other
What the company is doing
On 12/22/04 distribution of new units was placed on hold and customers were notified to recommend that they do not use the pace detect function. On 12/27/03 an email was sent to customers of the solution. On 12/29/04 a Technical Worksheet describing how to replace a resistor at R8 was emailed to customers. Between 12/29/04 and 01/03/05 kits containing instructions and new resistors were sent to customers to begin reworking of the devices.
Recall details
- Recall number
- Z-0532-05
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jan 5, 2005
- Posted by the FDA
- Feb 10, 2005
- Recalling firm
- Smiths Medical PM, Inc.
- Quantity
- 2
- Product code
- DPS
- FDA submissions
- K030407, K982279, K011177
- Distribution
- Internationally only.
Product description
Advisor Vital Signs Monitor model 9200, catalog number 92E654220. item Description 9200 II 3LD/RSP/SP02 MDD MEDEXL. The basic monitoring package includes ECG (3 lead / 5 lead), impedance respiration, non-invasive blood pressure, pulse oximetry, two invasive blood pressures, and tow temperature channels. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals.
Code information (models, lots, serial numbers)
serial numbers: 706459000 and 706459001
Other devices under the same FDA submission
2 devices cleared under K030407, K982279, K011177 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- BCI WW1100Icu Medical, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall2 reports
- BCI 920654230AIcu Medical, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 30811).
- Z-0531-05 · TerminatedAdvisor Vital Signs Monitor model 9200, catalog number 925454325. Item Description 9200 II 3LD/RP/02/IP/TP/BAT MDD. The basic monitoring package includes ECG (3 lead / 5 lead), …
- Z-0533-05 · TerminatedAdvisor Vital Signs Monitor model 9200, catalog number 92E654325. Item Description 9200 II 3LD/RP/O2/IP/TP/BATT ME. The basic monitoring package includes ECG (3 lead / 5 lead), …
- Z-0534-05 · TerminatedAdvisor Vital Signs Monitor model 9200, catalog number 92E654335. Item Description 9200II 3LD/RP/O2/IP/TP/PR/B ME. The basic monitoring package includes ECG (3 lead / 5 lead), i…
- Z-0535-05 · TerminatedAdvisor Vital Signs Monitor model 9200, catalog number 925474325. Item Description 9200II 3LD/RP/SO/O2/IP/TP BMDD. The basic monitoring package includes ECG (3 lead / 5 lead), i…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page