Nellcor brand OxiMax Pulse Oximeter, model number N-595 recallZ-0664-05 · Nellcor Puritan Bennett · initiated Oct 18, 2004

Terminated All recalls posted in March 2005

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Nellcor Puritan Bennett or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The internal speaker components are failing and do not produce an audio alarm in the event of patient emergency.

FDA root cause category: Other

What the company is doing

Customers of record were notified of the advisory by letter sent via US Mail on October 18, 2004, via certified mail. A second letter was sent on December 9, 2004, advising customers that upgrades would be available. Upgrades would be available via returns to the firm''s service center, upgrade kits and speaker replacements via self-installation, and a third party service provider.

Recall details

Recall number
Z-0664-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 18, 2004
Posted by the FDA
Mar 30, 2005
Recalling firm
Nellcor Puritan Bennett
Quantity
42,451 units
Product code
DQA
FDA submissions
K012891
Distribution
There are 2,950 Direct US and US territory Customers. Product shave also been distributed to Canada, Europe, Japan, Latin America and Australia.

Product description

Nellcor brand OxiMax Pulse Oximeter, model number N-595 Made in Ireland; Distributed by Tyco Healthcare Group, LP, Nellcor Puritan Bennett, Inc., Pleasanton, CA 94588 U.S.A.

Code information (models, lots, serial numbers)

All units with serial numbers beginning with G01844386.

Other devices under the same FDA submission

27 devices cleared under K012891 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 31271).