Skip to content

FHC 66-ZD-MD microTargeting Drive System: System for… recallZ-0895-05 · FHC, Inc. · initiated Mar 14, 2005

Terminated All recalls posted in June 2005

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from FHC, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for non-sterile pin to contaminate sterile field

FDA root cause category: Other

What the company is doing

FHC notified customers by letter dated 3/14/05 and 3/25/05. The firm has issued a temporary set of instructions which include covering the non-sterile engagement pin with sterile tape and an amended Instruction For Use page to all customers. A redesigned pulley cover which will be sent to all customer when it becomes available in April 2005.

Recall details

Recall number
Z-0895-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Mar 14, 2005
Posted by the FDA
Jun 16, 2005
Recalling firm
FHC, Inc.
Quantity
169 units
Product code
HAW
FDA submissions
K003776, K011775, K011992
Distribution
Natiowide

Product description

FHC 66-ZD-MD microTargeting Drive System: System for Stereotactic Positioning Used with the Power Assist (66-DA-ME) or Display Assembly (66-DA-EN)

Code information (models, lots, serial numbers)

All codes

Other devices under the same FDA submission

8 devices cleared under K003776, K011775, K011992 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page