The Arrow Trerotola Percutaneous Thrombolytic Device… recallZ-0842-05 · Arrow International Inc · initiated Mar 18, 2005
Terminated All recalls posted in May 2005
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Arrow International Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The orange Pebax lumen located within the catheter basket may separate from the catheter basket sleeve permitting the lumen to become dislodged during use.
FDA root cause category: Other
What the company is doing
The firm began notification of consignees on 3/23/2005 via certified letter.
Recall details
- Recall number
- Z-0842-05
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 18, 2005
- Posted by the FDA
- May 13, 2005
- Recalling firm
- Arrow International Inc
- Quantity
- 2418 UNITS
- Product code
- DXE
- FDA submissions
- K011056
- Distribution
- Product was distributed nationwide to dealers, hospitals, sales representatives, and subsidiaries (distributors). The product was also distributed to Australia, Canada, Colombia, Czech Republic, France, Germany, Israel, Italy, Japan, Netherlands, Russia, South Africa, South Korea, Spain, Sweden, Taiwan, Turkey, United Kingdom. There are no government accounts.
Product description
The Arrow Trerotola Percutaneous Thrombolytic Device (Product Codes: PT-65709-WC, PT-65709-W, PT-65709-HFWC, and PT-12709-WC) consists of a catheter with an expandable fragmentation basket attached to a through lumen torque cable and an outer catheter sheath. The catheter is used in conjunction with a low-speed, battery operated rotator unit to treat thrombosed hemodialysis grafts and AV fistulas. The macerating mechanism is done by 9 mm basket (made up of four stranded stainless steel wires) that rotates at 3000 rpm when rotator is activated. The current device is 7 Fr in size and comes in two lengths -- 65 cm and 120 cm. This device is used by inserting the catheter into the occluded graft or fistula via introducer sheath and then activating the rotator unit.
Code information (models, lots, serial numbers)
All units with Lot Numbers beginning with: MF311; MF312; MF401; MF402; MF403; MF404; MF405; MF406; MF407; MF408; MF409; MF410; MF411; MF412; MF501.
Other devices under the same FDA submission
11 devices cleared under K011056 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ARROW IPN035902Teleflex Incorporated · CATHETER, EMBOLECTOMY1 recall12 reports
- ARROW IPN036253Teleflex Incorporated · CATHETER, EMBOLECTOMY1 recall1 report
- ARROW IPN036254Teleflex Incorporated · CATHETER, EMBOLECTOMY1 recall1 report
- ARROW IPN036249Teleflex Incorporated · CATHETER, EMBOLECTOMY1 recall0 reports
- ARROW IPN036250Teleflex Incorporated · CATHETER, EMBOLECTOMY1 recall0 reports
- ARROW IPN036252Teleflex Incorporated · CATHETER, EMBOLECTOMY1 recall0 reports
- ARROW IPN036255Teleflex Incorporated · CATHETER, EMBOLECTOMY1 recall0 reports
- ARROW IPN920363Teleflex Incorporated · Catheter, embolectomy1 recall0 reports
- ARROW IPN920364Teleflex Incorporated · Catheter, embolectomy1 recall0 reports
- ARROW IPN920365Teleflex Incorporated · Catheter, embolectomy1 recall0 reports
- ARROW IPN914602Teleflex Incorporated · Catheter, embolectomy1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page