Skip to content

Hemochron Jr. Whole Blood Microcoagulation system Low Range… recallZ-1381-05 · International Technidyne Corp. · initiated Apr 27, 2005

Terminated All recalls posted in August 2005

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from International Technidyne Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Validation had been initiated for a molded component of the finished device but not completed. Also, the device did not meet mean clotting time using Normal Donor Blood as media.

FDA root cause category: Other

What the company is doing

Recall letter was sent out on 4/27/2005 via First Class Mail.

Recall details

Recall number
Z-1381-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 27, 2005
Posted by the FDA
Aug 23, 2005
Recalling firm
International Technidyne Corp.
Quantity
924 boxes x 45 cuvettes per box
Product code
JBP
FDA submissions
K960749
Distribution
Products were distributed domestically to customers nationwide. Products were also distributed internationally to customers in the following countries: Italy, Sweden, Norway, Germany, and Japan. Products were distributed to the VA Medical Center, Los Angeles, CA 90073.

Product description

Hemochron Jr. Whole Blood Microcoagulation system Low Range Activated Clotting Time. ACT-LR Cuvettes. In vitro diagnostic use, non-sterile, Packaged in individual foil pouches. 45 foil pouches per shelf carton box. Mfg. by International Technidyne Corp, Edison, NJ 08820.

Code information (models, lots, serial numbers)

Catalog No. JACT-LR Lot No''s:  M4JLR164, Exp. 3/2006; M4JLR165, Exp. 3/2006; A5JLR001, Exp. 4/2006; A5JLR002, Exp. 4/2006; A5JLR003, Exp. 4/2006; C5JLR032, Exp. 6/2006.

Other devices under the same FDA submission

1 device cleared under K960749 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page