Skip to content

Style 363 Saline-Filled BIOCELL textured Breast Implants recallZ-0851-05 · Inamed Corp · initiated May 5, 2005

Terminated All recalls posted in May 2005

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Inamed Corp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Wrong size implant was mislabeled.

FDA root cause category: Employee error

What the company is doing

Firm sent letter by overnite mail on May 5, 2005. Follow-up phone calls were made to confirm reciept. Doctors are to review surgery records and evaluate ammount of saline used to determine the risk for folding occuring and if found a visit by the patient is recommended to further evaluate for folding.

Recall details

Recall number
Z-0851-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
May 5, 2005
Posted by the FDA
May 25, 2005
Recalling firm
Inamed Corp
Quantity
6
Product code
FWM
FDA submissions
P990074
Distribution
WI, TN, TX, NY, AR

Product description

Style 363 Saline-Filled BIOCELL textured Breast Implants

Code information (models, lots, serial numbers)

UF6328, UF6330, UF6331, UM5056, UM5057, UM5059

Other devices under the same FDA submission

103 devices cleared under P990074 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.