HomeChoice Automated Peritoneal Dialysis Systems recallZ-1018-05 · Baxter Healthcare Renal Div · initiated Jun 21, 2005
Terminated All recalls posted in July 2005
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Renal Div or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The screws that secure the Power Entry Module (PEM) to the base of the HomeChoice device are not reliably grounded to the device's ground system, posing an electric shock hazard.
FDA root cause category: Other
What the company is doing
Baxter sent Urgent Device Correction letters worldwide to their customers on 6/21/05 and home patients on 6/22/05. The accounts were informed of teh risk of electrical shock from screws used to secure the Power Entry Module to the back of the HomeChoice system, and were instructed to always connect the power cord to the power entry module first, before plugging the power cord into a grounded electrical outlet. The users were advised to use caution to avoid contact the power entry module screws while powering on or powering off the device, and were informed that a new mounting bracket that providews more reliable grounding was installed in instruments processed through the Baxter service organization as of May 2005. All remaining instruments will receive the new mounting bracket by August 2006, and Baxter will contact the accounts to arrange for the upgrade. Any questions were directed to Baxter Global Technical Services at 1-800-553-6898, prompt 1.
Recall details
- Recall number
- Z-1018-05
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jun 21, 2005
- Posted by the FDA
- Jul 19, 2005
- Recalling firm
- Baxter Healthcare Renal Div
- Quantity
- 52,617 units
- Product code
- FKX
- FDA submissions
- K923065
- Distribution
- Nationwide, including Puerto Rico, and internationally to Argentina, Austria, Australia, Belgium, Canada, Chile, China, Colombia, Denmark, Ecuador, France, Germany, Guatemala, Hong Kong, Ireland, India, Italy, Korea, Mexico, Malaysia, Nicaragua, the Netherlands, New Zealand, Peru, the Philippines, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, the United Kingdom, Uruguay and Venezuela.
Product description
HomeChoice Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A.; catalog numbers 5C4471, 5C4471R, 5C4474, 5C4474R, T5C4441, T5C4441R
Code information (models, lots, serial numbers)
all serial numbers
Other devices under the same FDA submission
3 devices cleared under K923065 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- HOMECHOICE APD SET 5C4469CBaxter International Inc. · SYSTEM, PERITONEAL, AUTOMATIC DELIVERY252 reports
- Effluent Sample Bag 5C4476Baxter International Inc. · SYSTEM, PERITONEAL, AUTOMATIC DELIVERY59 reports
- HOMECHOICE APD SET 5C4469CVantive US Healthcare LLC · System, peritoneal, automatic delivery19 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 32251).
- Z-1019-05 · TerminatedHomeChoice PRO Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A.; catalog numbers 5C8310, 5C8310R, R…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page