Why it was recalled

Firm wants users to be aware not to leave the units in standby mode for more than an hour in rooms in excess of 75 F due to risk of an alarm mode being generated due to overheating.

FDA root cause category: Other

What the company is doing

Firm will be sending letters of caution to all users that reemphasizes labeling beginning 7/25/2005. New software is being developed which will include a change to maintain STANDBY coolant temperature regulation indefinitely, removing any risk of failure.

Recall details

Recall number
Z-1397-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jul 25, 2005
Posted by the FDA
Aug 24, 2005
Recalling firm
Alsius Corporation
Quantity
207
Product code
NCX
FDA submissions
K014241, K030421
Distribution
Nationwide and in Uruguay, Netherlands, Switzerland, Spain, Poland, China, Norway, Malasia, Italy, Ireland, greece, Germany, France, Finland, Canada, Belgium, Austria, Bahamas,

Product description

Alsius CoolGard 3000 Temperature Regulation System

Code information (models, lots, serial numbers)

All units manufactured before April 12, 2005.

Other devices under the same FDA submission

2 devices cleared under K014241, K030421 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.