Advisor Vital Signs Monitor (model 9200), catalog number… recallZ-0239-06 · Smiths Medical PM, Inc. · initiated Sep 1, 2005
Terminated All recalls posted in November 2005
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical PM, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the primary lead is set to Lead II, III or any of the augmented leads, AVL, AVR or AVF on the Advisor monitor.
FDA root cause category: Other
What the company is doing
Distirbutors and consignees were notified beginning 09/01/05 via email or phone about Safety Action Bulletin 05-004. Customers were asked to respond to Smiths Medical with a written acknowledgement that they received this Safety Action Bulletin (SAB). The SAB described the issue, actions they should take and instruction on how to have the monitor repaired.
Recall details
- Recall number
- Z-0239-06
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Sep 1, 2005
- Posted by the FDA
- Nov 30, 2005
- Recalling firm
- Smiths Medical PM, Inc.
- Quantity
- 1
- Product code
- DPS
- FDA submissions
- K010770, K011177, K030407, K982279
- Distribution
- AZ, CA, KS, KY, NY, TN, TX OUS to countries such as: Egypt, France, Iran, Mexico, Peru, Romania, Saudi Arabia and Spain
Product description
Advisor Vital Signs Monitor (model 9200), catalog number 92D754335, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main boards (P/N: 80002B1) 5-lead 60 Hz ECG, standard SPO2 option, invasive Pressure/Temperature, printer and battery.
Code information (models, lots, serial numbers)
serial number: AM05070143
Other devices under the same FDA submission
2 devices cleared under K010770, K011177, K030407, K982279 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- BCI WW1100Icu Medical, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall2 reports
- BCI 920654230AIcu Medical, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 33375).
- Z-0220-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 920634235, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0221-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 920654230, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0222-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), Catalog number 920654235, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0223-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 920654335, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0224-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 920655235, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0225-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 920754235, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0226-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 925454220, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0227-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 925454225, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0228-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 925454325, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0229-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 925454335, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0230-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 925474335, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0231-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 925534325, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0232-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 925654320, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0233-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 925654335, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0234-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 925674325, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0235-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 925754220, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0236-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 925754225, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0237-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 925754325, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0238-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 92D654335, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
- Z-0240-06 · TerminatedAdvisor Vital Signs Monitor (model 9200), catalog number 92E654335, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.