Skip to content

Life Pulse High Frequency Ventilator (HFV) recallZ-0893-06 · Bunnell, Inc. · initiated Apr 4, 2005

Terminated All recalls posted in May 2006

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bunnell, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential loss of air flow or change in cycling of the ventilator. Valve Drive Regulator Board may not function properly at elevated temperatures. The increase in temperature could cause a reduction of power, causing the solenoid in the valve to function sporadically.

FDA root cause category: Other

What the company is doing

Notified consignees by letter on 04/04/2005.

Recall details

Recall number
Z-0893-06
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 4, 2005
Posted by the FDA
May 23, 2006
Recalling firm
Bunnell, Inc.
Quantity
297 units
Product code
LSZ
FDA submissions
P850064
Distribution
Nationwide, Military medical center, Canada, and Czech Republic

Product description

Life Pulse High Frequency Ventilator (HFV)

Code information (models, lots, serial numbers)

Serial numbers 2585 and below.

Other devices under the same FDA submission

3 devices cleared under P850064 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.