FHC microTargeting Power Assist System used with the micro… recallZ-0179-06 · FHC, Inc. · initiated Oct 31, 2005

Terminated All recalls posted in November 2005

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from FHC, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Power Assist microTargeting System may advance beyond intended target

FDA root cause category: Other

What the company is doing

FHC Inc. and Medtronic notified customers by letter and instructs the customer to return their Motor Option for upgrade to include the following components: 1) Handheld Remote Control, 2) Motor Controller Unit with Display, 3) Motor, and 4) Medical Power Adapter. (See pictures at the end of this communication). Until the Motor Option has been updated, instructions to customers are: Prior to each clinical use: Ensure the microTargeting Drive system is calibrated before each clinical use per the User Guide. Check the microTargeting Drive system functionality over a range of at least 10mm movement. Should any performance indicating either a stall or unintended movement be observed, immediately unplug the 12V Medical Power Adapter, remove the MicroTargeting Drive Motor Option from the system, and continue using the microTargeting Drive in the manual mode. During clinical use: Should any performance indicating either a stall or unintended movement be observed, immediately unplug the 12V Medical Power Adapter, remove the MicroTargeting Drive Motor Option from the system, and continue using the microTargeting Drive in the manual mode. Every 100 uses or annually: Have your microTargeting Drive system checked by the manufacturer per the User Guide. Have your equipment support department check or tighten the MicroTargeting Drive Motor Option Remote Control Knob set screws, tighten as needed, and then re-calibrate.

Recall details

Recall number
Z-0179-06
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 31, 2005
Posted by the FDA
Nov 23, 2005
Recalling firm
FHC, Inc.
Quantity
70 units
Product code
HAW
FDA submissions
K011992
Distribution
Nationwide Foreign: South Africa, Taiwan, Hong Kong, China, France

Product description

FHC microTargeting Power Assist System used with the micro Targeting Drive System, models: 66-DA-ME; 66EL-MC; 66EL-MS;66EL-MS-02; 66-EL-MS-03; 66-EL-RM; MT-DS-AM; MT-DS-MD.

Code information (models, lots, serial numbers)

N/A

Other devices under the same FDA submission

2 devices cleared under K011992 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 33946).