Stratus CS STAT Fluorometric Analyzer CKMB TestPak (CCKMB)… recallZ-0875-06 · Dade Behring , Inc. · initiated Apr 3, 2006

Terminated All recalls posted in May 2006

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Dade Behring , Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Foil sealing of the TestPak may occlude the pipet tip resulting in insufficient aspiration of fluids. Erroneous test results may be generated without an associated error code. Test results may be falsely elevated or depressed, and the magnitude of the inaccuracy may vary based on the degree of occlusion.

FDA root cause category: Other

What the company is doing

Customers were provided clear written instructions (04/06/2006) to discontinue use and discard remaining inventory of products over 2 months old. In addition, each customer is being contacted by phone as an effectiveness verification. Recalling firm is also instructing Stratus CS customers to process all samples and dilutions in duplicate.

Recall details

Recall number
Z-0875-06
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 3, 2006
Posted by the FDA
May 19, 2006
Recalling firm
Dade Behring , Inc.
Quantity
3229 cartons of 100 TestPaks
Product code
JHX
FDA submissions
K984067
Distribution
Nationwide to hospitals and government accounts in addition to International accounts in Canada, Saudi Arabia, Bangladesh, UAE, India, Japan, Denmark, and Australia.

Product description

Stratus CS STAT Fluorometric Analyzer CKMB TestPak (CCKMB) in vitro diagnostic.

Code information (models, lots, serial numbers)

Lot #245276002; 245289002; 245304002; 245318002; 245332002; 245361002; 246023002; 246037002; 246051002; 246065002; and 246079002

Other devices under the same FDA submission

1 device cleared under K984067 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.