Locator-Wand Cover that is sold in a shelf carton (unit… recallZ-0804-06 · Smiths Medical MD, Inc. · initiated Mar 29, 2006
Terminated All recalls posted in April 2006
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical MD, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
One lot of Locator-Wand Cover sterile pouches may have an insufficient seal and product sterility could be compromised. The Locator-Wand Cover is supplied with the PORT-A-CATH II AND P.A.S. PORT systems. The packaged Wand-Cover is placed on the outside of the PORT-A-CATH II AND P.A.S. PORT tray.
FDA root cause category: Other
What the company is doing
An Urgent Product Safety & Recall Notification, dated 03/29/06, was mailed to domestic customers and emailed to international customers. The letter described the issue, identified re-order number and lot numbers affected and gave 3 options. The options include to use only the sterile tray and discard the wand cover from the shelf box; remove and discard Locator-Wand Cover pouches and request replacement Locator-Wand Covers; or return the entire device for a credit or replacement.
Recall details
- Recall number
- Z-0804-06
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 29, 2006
- Posted by the FDA
- Apr 27, 2006
- Recalling firm
- Smiths Medical MD, Inc.
- Quantity
- 506 trays
- Product code
- MMP
- FDA submissions
- K871132, K931073, K942024
- Distribution
- Nationwide. AR, CA, DE, FL, IL, KY, LA, MA, MD, MI, MN, MS, NJ, NY, OK, PA, TN, TX, VA, WI. Germany, Italy, Netherlands, Singapore, Spain
Product description
Locator-Wand Cover that is sold in a shelf carton (unit packaging) separate to the sterile tray containing and identified as PORT-A-CATH II Fluoro-Free, Low Profile Venous Access System, PolyFlow Polyurethane Catheter, 1.9mm O.D. tray. REF 21-4683-24. Contents: Portal with ULTRA-LOCK Connector, Radiopaque PolyFlow Polyurethane Catheter with CATH-FINDER Sensor Wire, Locator-Wand cover, Nonvented Luer Cap, 20G PORT-A-CATH needle, 18G Blunt Needle, 20G GRIPPER PLUS Non-Coring Safety needle, 18G Extra Thin Wall Introducer Needle, 6 Fr Introducer, 2 ''J'' Guidewires, Syringes, Vein Pick, Tunneling Tool, 20G 90-degree PORT-A-CATH needle, 3 Point-Lok Sharps Safety Devices, Product Information. EtO Sterilized. Single Use. Deltec, Inc., St. Paul, MN 55112 U.S.A.
Code information (models, lots, serial numbers)
Re-order number 21-4683-24, Lot numbers: M41020, M41851, M42231, M42434, M42880, M43365
Other devices under the same FDA submission
2 devices cleared under K871132, K931073, K942024 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Deltec 21-4083-24Icu Medical, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR28 reports
- Deltec 21-4071-24Icu Medical, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR6 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 35148).
- Z-0802-06 · TerminatedLocator-Wand Cover that is sold in a shelf carton (unit packaging) separate to the sterile tray containing and identified as PORT-A-CATH P.A.S. PORT T2 Fluoro-Free, Venous Access…
- Z-0803-06 · TerminatedLocator-Wand Cover that is sold in a shelf carton (unit packaging) separate to the sterile tray containing and identified as PORT-A-CATH P.A.S. PORT T2 Fluoro-Free, Venous Access…
- Z-0805-06 · TerminatedLocator-Wand Cover that is sold in a shelf carton (unit packaging) separate to the sterile tray containing and identified as PORT-A-CATH II Fluoro-Free, Low Profile Venous Access…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page