Weck Hem-o-lok Endo 5 Automatic Clip Applier with Medium… recallZ-1099-06 · Teleflex Medical · initiated Apr 12, 2006

Terminated All recalls posted in June 2006

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Teleflex Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Hem-o-lok ligating clips may become dislodged following ligation of the renal artery after laparoscopic donor nephrectomy. Hem-o-lok ligating clips are now contraindicated for use in ligating the renal artery during laparascopic nephrectomies in living donor patients.

FDA root cause category: Other

What the company is doing

Teleflex Medical sent Important Product Safety Information letters dated 4/18/06 to all of their customers who have purchased Hem-o-lok Ligating Clips from Teleflex. The letters informed the healthcare providers that Teleflex has been made aware of incidents in which the ligating clips (sizes L and XL) were reported to have become dislodged following ligation of the renal artery after laparoscopic donor nephrectomy. Although none of the incidents appear to have involved any defect in or malfunction of the ligating clips, misapplication of the Hem-o-lok clips during such laparascopic procedures may not immediately be apparent and can have serious, even life-threatening consequences post-operatively. Therefore, Teleflex has added the following contraindication to the Instructions for Use accompanying the Hem-o-lok Polymer Ligating Clips: 'Contraindications': Hem-o-lok ligating clips are contraindicated for use in ligating the renal artery during laparoscopic nephrectomies in living donor patients. 'Caution': The clip must be latched to ensure proper ligation of the vessel or tissue. Inspect the ligation site after application to ensure proper closure of the clip. Security of the closure should be confirmed after ligation. The Hem-o-lok Polymer Ligating Clip is not designed for use as a tissue marker. Weck recommends that more than one clip be used to ligate the renal artery in procedures other than laparoscopic donor nephrectomy (see Contraindication, above). Application of more than one clip to all other vessels should be left to the surgeon's judgment.' Any questions concerning the letter were directed to Teleflex Medical at 800-334-9751, option #7, then option #5. Teleflex sent follow-up Important Product Correction Notice letters dated 8/2/06 to all of their Hem-o-lok Ligating Clips customers via registered mail on 8/7/06. The letters provide clarification of the 4/18/06 letter, reinforcing that the clips are contraindicated for use in ligating t

Recall details

Recall number
Z-1099-06
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 12, 2006
Posted by the FDA
Jun 9, 2006
Recalling firm
Teleflex Medical
Quantity
374,760 clips
Product code
FZP
FDA submissions
K021808
Distribution
Nationwide

Product description

Weck Hem-o-lok Endo 5 Automatic Clip Applier with Medium Polymer Ligating Clips; Rx, sterile, single use, disposable clip applier with medium size, non-absorbable polymer ligation clips; 15 clips per applier, 3 appliers per sales unit, 6 sales units per case; Mfg. by Weck Closure Systems, 2917 Weck Drive, Research Triangle Park, NC 27709; REF 543965

Code information (models, lots, serial numbers)

catalog number 543965, all lots

Other devices under the same FDA submission

1 device cleared under K021808 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.