Skip to content

ADVIA 1200 Chemistry System, Automated Clinical Chemistry… recallZ-1449-06 · Bayer Healthcare, LLC (Diagnostics Division) · initiated Feb 17, 2006

Terminated All recalls posted in August 2006

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bayer Healthcare, LLC (Diagnostics Division) or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Low QC recovery observed on carbamazepine (CARB) results immediately following a Gentamicin (GENT) assay or a Digoxin (DIG) assay. When GENT or DIG precedes CARB, the CARB test result is artificially low (as much as -25%). PHNY was determined to also exhibit a similar carryover effect on CARB. All others were found not to produce the same reagent probe-based carryover effect.

FDA root cause category: Other

What the company is doing

Bayer HealthCare LLC issued a Support Bulletin to all affected Bayer HealthCare LLC Branches on/about 2/17/06 in order to inform them of the issue and to provide them with instructions for this ''field correction''. Customer Bulletins were also issued to the Branches both in the United States and elsewhere for communication with the affected customers.

Recall details

Recall number
Z-1449-06
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Feb 17, 2006
Posted by the FDA
Aug 29, 2006
Recalling firm
Bayer Healthcare, LLC (Diagnostics Division)
Quantity
1138 for all products
Product code
JGS
FDA submissions
K990346
Distribution
Worldwide including USA and Puerto Rico.

Product description

ADVIA 1200 Chemistry System, Automated Clinical Chemistry Analyzer - human serum, plasma and urine tests.

Code information (models, lots, serial numbers)

Part No. 094-A001; all Advia Chemistry System units and serial numbers are involved in this ''field correction'' as well as existing assay lots for CARB, PHNY, GENT and DIG.

Other devices under the same FDA submission

2 devices cleared under K990346 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 35880).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page