Boston Scientific Cutting Balloon Ultra2 Monorail Device… recallZ-1512-06 · Boston Scientific · initiated Jul 21, 2006
Terminated All recalls posted in September 2006
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Lack of assurance of sterility (pre-sterilization bioburden limits exceeded)
FDA root cause category: Other
What the company is doing
On July 21, 2006, Boston Scientific initiated a voluntary recall of 45 lots/batches of Cutting Balloon Ultra Monorail Devices due to their determination that the lots do not meet their sterilization requirements. There is a remote possibility of infection in the patient. An infection, if it occurred, could be recognized and treated effectively in these patients. Boston Scientific is not aware of any complaints related to this product issue. Letters dated July 21, 2006, were sent to their customers with instructions to cease further distribution, complete and return a tracking form, and product return instructions.
Recall details
- Recall number
- Z-1512-06
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jul 21, 2006
- Posted by the FDA
- Sep 23, 2006
- Recalling firm
- Boston Scientific
- Quantity
- 1,000
- Product code
- LOX
- FDA submissions
- P950020
- Distribution
- Class II Recall - Nationwide distribution ---- including states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV, and WY.
Product description
Boston Scientific Cutting Balloon Ultra2 Monorail Device, Boston Scientific, One Boston Scientific Place, Natick, MA 01760.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Material Number (UPN) / Catalog Number / Lot/Batch #: H749CBM200060 / CBM20006 / EG0438 & EG0499; H749CBM200100 / CBM20010 / EG0427; H749CBM225060 / CBM22506 / EG0448 & EG0471; H749CBM225100 / CBM22510 / EG0444 & EG0502; H749CBM250060 / CBM25006 / EG0439 & EG0472; H749CBM250100 / CBM25010 / EGO425, EG0440, EG0474 & EG0475; H749CBM250150 / CBM25015 / EG0466; H749CBM275060 / CBM27506 / EG0420 & EG0487; H749CBM275100 / CBM27510 / EG0419 & EG0456; H749CBM275150 / CBM27515 / EG0424; H749CBM300060 / CBM30006 / EG0423, EG0473 & EG0497; H749CBM300100 / CBM30010 / EG0426, EG0441, EG0465, EG0485; H749CBM300150 / CBM30015 /EG0443 & EG0418; H749CBM325060 / CBM32506 / EG0500; H749CBM325100 / CBM32510 / EG0422 & EG0501; H749CBM325150 / CBM32515 / EG0467; H749CBM350060 / CBM35006 / EG0437; H749CBM350100 / CBM35010 / EG0349, EG0428, EG0468; H749CBM350150 / CBM35015 / EG0421 & EG0486; H749CBM375060 / CBM37506 / EG0503; H749CBM375100 / CBM37510 / EG0457; H749CBM375150 / CBM37515 / EG0455; H749CBM400060 / CBM40006 / EG0504; H749CBM400100 / CBM40010 / EG0442 & EG0498; H749CBM400150 / CBM40015 / EG0476.
Other devices under the same FDA submission
37 devices cleared under P950020 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Wolverine Coronary Cutting Balloon H74939401103000Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring1 death36 reports
- Wolverine Coronary Cutting Balloon H74939401103500Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring21 reports
- Wolverine Coronary Cutting Balloon H74939401104000Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring19 reports
- Wolverine Coronary Cutting Balloon H74939401153000Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring19 reports
- Wolverine Coronary Cutting Balloon H74939401153500Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring19 reports
- Wolverine Coronary Cutting Balloon H74939401102500Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring16 reports
- Wolverine Coronary Cutting Balloon H74939401154000Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring15 reports
- Wolverine Coronary Cutting Balloon H74939401063000Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring9 reports
- Wolverine Coronary Cutting Balloon H74939401062500Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring8 reports
- Flextome® Cutting Balloon® H749CBM3400150Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring7 reports
- Wolverine Coronary Cutting Balloon H74939401102000Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring7 reports
- Wolverine Coronary Cutting Balloon H74939401152500Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring7 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page