Skip to content

RapidLab 1200 Systems, Model 1265 - Blood gases… recallZ-1465-06 · Siemens Healthcare Diagnostics · initiated Apr 4, 2006

Terminated All recalls posted in August 2006

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Invalid CO-oximeter values; Bayer HealthCare determined that under very specific conditions, the RapidLab 1245 or RapidLab 1265 systems may report invalid CO-oximeter values to an LIS, Rapidlink, or Rapidcomm™ data management system.

FDA root cause category: Software design

What the company is doing

On 4/04/06, Bayer HealthCare LLC (Diagnostics Division) sent Support Bulletins to all affected Bayer HealthCare LLC Branches via e-mail in order to inform them of the issue, and to provide them with instructions for the interim work-around until the new software version is installed. Customer Bulletins were prepared and sent to the Branches both in the United States and globally for communication with the affected customers.

Recall details

Recall number
Z-1465-06
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 4, 2006
Posted by the FDA
Aug 29, 2006
Recalling firm
Siemens Healthcare Diagnostics
Quantity
Domestic - 124 units, Foreign-364 units (all systems)
Product code
CHL
FDA submissions
K031560
Distribution
Worldwide

Product description

RapidLab 1200 Systems, Model 1265 - Blood gases, electrolyte and blood pH test system, Bayer HealthCare LLC, Diagnostics Division

Code information (models, lots, serial numbers)

Part No. 05063769; all units using Version 1.0, 1.01, or 1.02 software connected to LIS, Rapidlink¿, or Rapidcomm™ data management systems.

Other devices under the same FDA submission

2 devices cleared under K031560 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 35944).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page