Skip to content

bioMerieux BacT/ALERT SN Anaerobic Culture Bottle-REF… recallZ-0243-2007 · bioMerieux, Inc. · initiated Aug 17, 2006

Terminated All recalls posted in December 2006

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from bioMerieux, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Recovery Compromised-during manufacturing ambient air (including oxygen) was inadvertently introduced into the culture bottles, changing their reduction/oxidation potential thus leading to a compromised recovery of some obligate anaerobes and longer detection times.

FDA root cause category: Other

What the company is doing

Affected Customers were notified of this recall on August 18, 2006 and August 21, 2006 via phone and instructed to destroy product remaining in stock.

Recall details

Recall number
Z-0243-2007
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 17, 2006
Posted by the FDA
Dec 5, 2006
Recalling firm
bioMerieux, Inc.
Quantity
34,440
Product code
MDB
FDA submissions
K021123
Distribution
Nationwide Distribution ---- including states of AL, CA, CT, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MS, MT, NC, NE, NJ, NY, OH, PA, RI, TX, VA, WA, WI, and WV.

Product description

bioMerieux BacT/ALERT SN Anaerobic Culture Bottle-REF 259790

Code information (models, lots, serial numbers)

Lot number 1013177 (Product No. 259790)

Other devices under the same FDA submission

1 device cleared under K021123 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates