Medtronic Intersept Custom Tubing Pack. Do Not Reuse. The… recallZ-0340-2007 · Medtronic Perfusion Systems · initiated Sep 13, 2006
Terminated All recalls posted in January 2007
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Perfusion Systems or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An out of specification condition exists with the tapered tubing found in some Custom Tubing Packs. The wall thickness is undersized and does not meet specifications.
FDA root cause category: Other
What the company is doing
An Important Customer Notification, dated 09/06/06, was sent to affected customers. The letter identifies the affected model and lot numbers. Medtronic requests that affected product be returned. A suplemental pack can be built and shipped to the customer upon request. Medtronic will replace all returned product. A Recall Certificate is asked to be completed by customer and faxed to Medtronic.
Recall details
- Recall number
- Z-0340-2007
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Sep 13, 2006
- Posted by the FDA
- Jan 25, 2007
- Recalling firm
- Medtronic Perfusion Systems
- Quantity
- 146
- Product code
- DWE
- FDA submissions
- K800178
- Distribution
- Distributed in 6 states: CA, FL, IL, IN, LA and TN
Product description
Medtronic Intersept Custom Tubing Pack. Do Not Reuse. The design and components of this product were specified by the user from an array of components and designs offered by the manufacturer. Any change in the cardiopulmonary bypass procedure that compromises the function of this Custom Tubing Pack is the responsibility of the user. The tubing is made of polyvinyl chloride (PVC). Portions of the tapered tubing are used in the vent, suction and arterial roller heads of the heart lung machine. Manufacturer: Medtronic, Inc., Minneapolis, MN 55432-5604 USA. Tube supplier: Natvar (Tekni-Plex Co.), 19555 E. Arenth Ave., City of Industry, CA 91748.
Code information (models, lots, serial numbers)
The following Model / Lot Number are affected: CB4W39R1 / 4876667; 0A84R2 / 4531036; 1708R18 / 4627100; 0V68R4 / 4212182; 1P07R6 / 4433014; 1P07R6 / 4678134; 2Y91R / 4628170; 2Y91R / 4800224; 2Y91R / 4902036; CB0Q49R14 / 4800849; 4D87R2 / 4184927.
Other devices under the same FDA submission
113 devices cleared under K800178 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA BB9L72R1Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS3 recalls0 reports
- NA BB8M16R3Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls1 report
- NA BB7P48R4Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA BB7P46R4Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA BB8L58R3Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA BB8B70R7Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA BB8D40R3Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA BB8T77R2Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA BB7V77RMedtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS1 recall1 report
- NA BB7V76RMedtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS1 recall1 report
- NA HY7S51R10Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS1 recall1 report
- NA BB8D56R9Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS1 recall1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.