Amplatzer Cardiac Septal Occluder Delivery Systems labeled… recallZ-0255-2007 · AGA Medical Corporation · initiated Oct 2, 2006
Terminated All recalls posted in December 2006
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from AGA Medical Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
AGA Medical has found that microscopic tears can occur in the delivery system sterile packaging under accelerated stress testing with routine shipping configurations. These microscopic tears are a potential breach of the sterile barrier. AGA has no confirmed complaints or adverse events related to this failure mode.
FDA root cause category: Other
What the company is doing
An Urgent Recall Notice, dated 09/29/06, was sent to customers via facsimile or email and via Federal Express. The letter describes the issue and what product is affected. Unexpired product that is returned to AGA will be replaced. A response form is asked to be returned to AGA.
Recall details
- Recall number
- Z-0255-2007
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 2, 2006
- Posted by the FDA
- Dec 19, 2006
- Recalling firm
- AGA Medical Corporation
- Quantity
- 90,516
- Product code
- GBK
- FDA submissions
- P020024
- Distribution
- Worldwide, including USA, Puerto Rico, Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Cuba, Denmark, El Salvador, Finland, France, Germany, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Israel, Italy, Japan, Jordan, Korea, Kuwait, Malaysia, Moroc, Mexico, Norway, Pakistan, Peru, Phillipines, Poland, Portugal, Saudi Arabia, Singapore, Slovak Republic, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, The Netherlands, Tunisia, Turkey, United Kingdom, Uruguay, Venezuela, and Vietnam.
Product description
Amplatzer Cardiac Septal Occluder Delivery Systems labeled as follows: a) AMPLATZER Delivery System, Order no. 9-DEL-5F-180/60, 9-DEL-6F-45/60, 9-DEL-6F-180/80, 9-DEL-7F-45/60, 9-DEL-7F-45/80, 9-DEL-7F-180/80, 9-DEL-8F-45/60, 9-DEL-8F-45/80, 9-DEL-8F-180/80, 9-DEL-9F-45/80, 9-DEL-9F-180/80, 9-DEL-10F-45/80, 9-DEL-12F-45/80. [Breakdown of order no. ''9-DEL-5F-180/60'' is as follows: 9-DEL, French size (5F) - degree angle curve (180) / usable length in cm (60).] b) the following devices are not approved for US distribution: AMPLATZER TorqVue Delivery System, order no. 9-ITV06F45/60, 9-ITV07F45/60, 9-ITV07F45/80, 9-ITV08F45/60, 9-ITV08F45/80, 9-ITV09F45/80, 9-ITV10F45/80, 9-ITV12F45/80, 9-ITV05F180/60, 9-ITV06F180/80, 9-ITV07F180/80, 9-ITV08F180/80, 9-ITV09F180/80. Manufactured in USA, Sterile EO. Headquarters AGA Medical Corporation, 682 Mendelssohn Avenue, Golden Valley, MN 55427 U.S.A. The Delivery System is comprised of a delivery sheath, delivery cable, dilator, loading device and pin vise.
Code information (models, lots, serial numbers)
Show the full code information (10k characters)
Order no. 9-DEL-5F-180/60, Lot Numbers: M03J19-05, M03L18-49, M04B02-02, M04C01-19, M04E18-13, M04J13-15, M04L02-04, M04M28-10, M05C15-31, M05C25-27, M05D29-51, M05G20-44, M05G28-19, M05H30-45, M05J23-32. Order no. 9-DEL-6F-45/60, Lot Numbers: M03L11-45, M03J23-04, M04B09-16, M04E27-19, M04L22-12, M05A03-10, M05B04-20, M05B08-04, M05C08-23, M05C09-05, M05E24-48, M05H01-51, M05J12-16, M06C20-25, M06D03-02, M06D19-71, M06E22-46. Order no. 9-DEL-6F-180/80, Lot Numbers: M03K16-22, M03K28-33, M03L11-47, M03M02-09, M03M03-09, M03M12-18, M04A19-31, M04B03-10, M04C01-17, M04C23-12, M04D06-10, M04D13-16, M04E24-16, M04F23-21, M04F28-40, M04G01-22, M04G27-27, M04H03-17, M04H12-11, M04H24-23, M04L08-02, M04M16-21, M04M30-04, M05A03-09, M05A05-27, M05A12-07, M05A17-07, M05A18-04, M05A20-09, M05A27-27, M05A31-06, M05B02-26, M05B21-13, M05B23-09, M05C08-04, M05C25-02, M05D07-08, M05D13-09, M05D14-13, M05G07-20, M05G11-03, M05G12-06, M05G15-24, M05G15-31, M05G27-41, M05H08-19, M05H04-10, M05H05-29, M05H09-09, M05K14-06, M05K20-10, M05K28-02, M05M28-23, M06A13-17, M06D19-53, M06D19-57, M06D19-55, M06D19-59, M06E16-68, M06E22-65, M06E22-67, M06F07-90, M06F13-86. Order no. 9-DEL-7F-45/60, Lot Numbers: . Order no. , Lot Numbers: M03L12-09, M04C23-09, M04E24-17, M04G01-24, M04M10-18, M05B04-18, M05B07-06, M05B16-02, M05D08-09, M05F13-11, M05G15-16, M05G19-24, M05J08-03. Order no. 9-DEL-7F-45/80, Lot Numbers: M03J19-07, M03K14-06, M03K28-35, M03L11-48, M04B03-09, M04B03-14, M04C08-12, M04D06-12, M04E11-28, M04E24-20, M04F08-23, M04H03-23, M04J17-07, M04K07-02, M04K28-06, M04L30-04, M05A11-12, M05A11-10, M05A12-05, M05B04-09, M05B16-04, M05B16-23, M05B23-07, M05C10-07, M05C14-07, M05C15-29, M05D04-15, M05E03-02, M05E12-16, M05E12-03, M05E16-21, M05E17-45, M05F01-02, M05F30-25, M05G07-02, M05G21-06, M05G29-23, M05H01-11, M05H17-09, M05J16-13, M05J16-68, M05J28-08, M05K04-04, M05M05-11, M06A13-19, M06C20-23, M06C28-42, M06D04-77, M06D12-20, M06E16-80, M06F01-43, M06F07-88, M06F13-76, M06G25-96. Order no. 9-DEL-7F-180/80, Lot Numbers: M03K24-14, M03L12-10, M03M12-19, M03G15-34, M04B03-08, M04B09-17, M04C01-18, M04D01-05, M04D06-11, M04D20-07, M04E04-05, M04F08-21, M04H03-18, M04H24-25, M04K22-13, M04L30-12, M04M13-05, M05A18-26, M05A20-11, M05B08-06, M05B14-22, M05B18-10, M05B18-12, M05C16-33, M05C16-18, M05D05-39, M05E11-30, M05E18-04, M05E26-23, M05G08-07, M05G15-29, M05G26-16, M05H02-22, M05H03-11, M05J13-14, M05J19-66, M05J23-07, M05J26-19, M05K19-04, M05L30-22, M05M22-20, M06D19-67, M05D19-69, M06E16-76, M06E22-71, M06E22-69. Order no. 9-DEL-8F-45/60, Lot Numbers: M03K14-07, M03L18-50, M03M03-11, M04C23-10, M04E24-18, M04G01-20, M04L30-02, M04M15-20, M05B07-08, M05B14-04, M05B15-07, M05B24-14, M05B25-06, M05H19-13. Order no. 9-DEL-8F-45/80, Lot Numbers: M03J19-08, M03J25-27, M03K09-23, M03K28-36, M03L11-46, M03M03-12, M03M29-07, M04A19-28, M04A28-16, M04B03-13, M04B09-19, M04B17-33, M04B24-08, M04C23-11, M04D06-13, M04D26-06, M04D26-10, M04F08-22, M04F27-47, M04H24-22, M04J07-09, M04J10-22, M04J17-08, M04J17-09, M04K06-02, M04K11-08, M04K15-10, M04K25-25, M04L02-02, M04L08-12, M04L29-18, M04M01-02, M04M10-08, M04M20-16, M04M22-02, M04M30-05, M05A11-08, M05A18-28, M05A21-20, M05A21-24, M05A24-34, M05A27-13, M05A28-13, M05A31-08, M05B17-04, M05B17-06, M05C21-24, M05C24-04, M05C25-18, M05C30-24, M05C31-28, M05D06-07, M05D12-03, M05D14-09, M05D20-06, M05D25-19, M05D27-20, M05D28-13, M05E24-17, M05E25-21, M05E27-46, M05F06-46, M05F21-43, M05F21-41, M05G20-10, M05G27-39, M05H12-16, M05H17-13, M05H24-19, M05J09-10, M05J09-12, M05K11-03, M05K18-10, M05K20-08, M05K21-05, M05L17-19, M05L21-04, M05L28-04, M05L30-20, M06A09-15, M06A12-10, M06A12-12, M06A20-25, M06A20-27, M06A24-18, M06A20-29, M06C17-36, M06C17-38, M06D04-73, M06D11-05, M06D12-12, M06D12-14, M06D12-16, M06D12-18, M06E22-49, M06E22-51, M06F01-47, M06F07-82, M06F07-84, M06F07-86, M06F13-72, M06F13-74, M06F22-19, M06F29-12, M06G24-46. Order no. 9-DEL-8F-180/80, Lot Numbers: M03L18-52, M04B09-18, M04C23-13, M04E27-20, M04F08-24, M04H03-20, M04K26-18, M04L04-17. M05B11-08, M05C01-06, M05C01-08, M05C11-02, M05E02-21, M05F28-12, M05J01-38, M05J22-28, M05K18-04. Order no. 9-DEL-9F-45/80, Lot Numbers: M03K14-08, M03K24-15, M03K28-37, M03M02-12, M03M03-08, M03M29-03, M03M29-06, M03M30-13, M04A19-29, M04A28-18, M04B03-12, M04B17-38, M04C08-13, M04C15-04, M04C15-06, M04D06-09, M04D13-17, M04D20-08, M04D26-07, M04D26-09, M04E24-19, M04E27-24, M04F21-43, M04F23-23, M04G27-25, M04H12-12, M04H24-20, M04J22-13, M04K08-02, M04L05-08, M04L19-11, M04L29-02, M04L29-17, M04L29-23, M04L30-10, M04M14-02, M04M14-04, M04M14-21, M04M16-23, M04N22-23, M05A05-25, M05A03-12, M05A11-06, M05A13-19, M05A18-24, M05A19-04, M05A24-09, M05A24-32, M05A24-36, M05A25-19, M05A28-11, M05B01-05, M05B02-24, M05B04-22, M05B10-05, M05B11-06, M05B22-03, M05B23-05, M05B24-20, M05C10-25, M05C15-27, M05C24-02, M05D04-13, M05D06-05, M05D07-04, M05D08-11, M05D12-20, M05D26-26, M05E11-14, M05E11-16, M05E13-09, M05E31-04, M05E31-10, M05E31-12, M05F06-44, M05F03-12, M05F09-06, M05F10-07, M05F14-26, M05F16-17, M05F27-06, M05F29-06, M05G06-07, M05G18-30, M05G19-15, M05G27-07, M05G27-43, M05H01-09, M05H12-14, M05H18-34, M05H19-08, M0530-36, M05J07-10, M05J08-05, M05J16-57, M05L01-02, M05L18-02, M05L30-06, M05M02-21, M05M06-13, M05M08-16, M05M22-12, M05M22-16, M06A09-12, M06A16-23, M06C15-52, M06C20-27, M06C24-22, M06D04-69, M06D04-71, M06D19-61, M06D19-63, M06D21-03, M06D21-05, M06D21-07, M06E22-53, M06E22-55, M06F13-78, M06F29-10, M06G25-93. Order no. 9-DEL-9F-180/80, Lot Numbers: M03K28-34, M03L18-51, M04L22-14, M05C28-12, M05D07-10, M05E03-13, M05F22-09, M05J26-21. Order no. 9-DEL-10F-45/80, Lot Numbers: M03J25-26, M03K01-02, M03K09-22, M03K14-05, M03K16-16, M03K16-17, M03K24-09, M03L18-56, M03M30-11, M03M30-14, M03M30-15, M04A19-30, M04B03-11, M04B17-34, M04B24-09, M04C30-08, M04F08-20, M04F23-20, M04F28-46, M04G01-19, M04G01-21, M04H23-20, M04H23-21, M04J07-04, M04L29-19, M04L29-20, M04M22-25, M04M17-19, M04M29-09, M04M29-10, M05A03-11, M05A10-07, M05A10-09, M05A14-08, M05B07-24, M05B07-25, M05B07-26, M05B18-06, M05B21-04, M05B24-08, M05C03-08, M05C07-06, M05C10-05, M05C10-27, M05C15-25, M05C23-07, M05C29-13, M05D29-05, M05E04-17, M05E05-24, M05E09-19, M05E09-21, M05E16-19, M05E17-43, M05E24-13, M05E27-38, M05E26-33, M05F09-08, M05F28-08, M05F28-10, M05H03-37, M05H22-06, M05H25-15, M05H25-13, M05J01-13, M05J22-23, M05K10-07, M05K13-04, M05K19-02, M05K21-03, M05L18-04, M05L29-12, M05M02-23, M05M01-05, M05M22-18, M06A13-15, M06C15-55, M06C24-20, M06D04-65, M06D04-67, M06E08-10, M06E08-12, M06F07-80, M06F13-82, M06G25-94. Order no. 9-DEL-12F-45/80, Lot Numbers: M03J25-28, M03K16-21, M03K24-13, M03L18-55, M03M03-13, M03M30-12, M03M30-16, M04A28-19, M04C09-03, M04C09-04, M04C15-05, M04D01-04, M04D06-08, M04D26-08, M04F21-42, M04F23-22, M04G27-26, M04H03-22, M04H24-21, M04J30-02, M04K05-02, M04K25-23, M04L04-09, M04M10-05, M04M23-02, M05A13-03, M05A14-11, M05A19-06, M05B04-24, M05B21-28, M05C03-04, M05C02-06, M05C07-08, M05C08-02, M05C21-25, M05D07-06, M05D15-28, M05D19-02, M05D18-45, M05D25-04, M05D28-23, M05D29-19, M05E02-08, M05E10-29, M05E13-26, M05E19-07, M05E20-17, M05E24-15, M05F07-12, M05F15-35, M05F15-43, M05F20-05, M05F21-54, M05G13-16, M05G25-15, M05G27-09, M05H05-27, M05H26-08, M05H29-16, M05J09-27, M05J12-09, M05J12-07, M05K06-10, M05K07-25, M05K31-28, M06C15-57, M06D04-75, M06D12-10, M06E16-78, M06F01-45, M06F13-84. the following devices are not approved for US distribution: AMPLATZER TorqVue Delivery System, order no. 9-ITV06F45/60, Lot numbers: M05K24-25, M05L03-06, M06C03-17, M06C27-05, M06E22-57. Order no. 9-ITV07F45/60, Lot numbers: M05K24-27, M05L04-13, M06B10-21, M06E01-40, M06F13-96. Order no. 9-ITV07F45/80, Lot numbers: M05K24-04, M05L10-07, M05L14-13, M05L15-34, M06B01-31, M06C02-14, M06F01-33, M06F07-68, M06F07-70. Order no. 9-ITV08F45/60, Lot numbers: M05K24-29, M05L03-08, M06B10-25. Order no. 9-ITV08F45/80, Lot numbers: M05K24-31, M05L11-15, M05M06-17, M05M07-10, M05M12-16, M06A05-25, M06B01-33, M06B10-27, M06B10-23, M06E03-10, M06E03-08, M06E16-84, M06F01-41, M06F13-90, M06G06-18, M06G25-79. Order no. 9-ITV09F45/80, Lot numbers: M05K24-33, M05L10-09, M05M12-18, M05M12-20, M06A12-19, M06B01-37, M06B10-29, M06C03-14, M06C10-18, M06C10-20, M06C13-38, M06C13-40, M06D19-73, M06D26-26, M06D26-28, M06D26-30, M06F01-39, M06F07-76, M06F07-72, M06F13-98, M06F14-03, M06F22-17, M06F14-01, M06G24-28, M06F22-15. Order no. 9-ITV10F45/80, Lot numbers: M05K24-06, M05L02-12, M05M07-12, M05M12-22. M06B01-39, M06C02-16, M06C10-22, M06C10-24, M06C13-42, M06C13-44, M06D12-08, M06D12-06, M06F13-92, M06G06-24, M06G06-22. Order no. 9-ITV12F45/80, Lot numbers: M05K24-35, M05L14-09, M05M05-15, M05M12-24, M06B01-41, M06C02-18, M06C20-39, M06C10-28, M06C10-26, M06D12-01, M06E03-04, M06E09-59, M06G25-89. Order no. 9-ITV05F180/60, Lot numbers: M05K24-02, M05L02-15, M06C27-07. Order no. 9-ITV06F180/60, Lot numbers: Order no. 9-ITV06F180/80, Lot numbers: M05K24-08, M05L03-10, M05L04-15, M05L17-23, M05L28-03, M06C20-29, M06C20-31, M06C27-08, M06C20-33, M06E01-05, M06G06-26. Order no. 9-ITV07F180/80, Lot numbers: M05K24-10, M05L08-06, M05L15-28, M05L17-25, M06B17-34, M06B21-43, M06B21-45, M06B21-47, M06G24-24. Order no. 9-ITV08F180/80, Lot numbers: M05K24-23, M05L11-17, M06B17-28, M06D19-65, M06E08-05. Order no. 9-ITV09F180/80, Lot numbers: M05K24-21, M05L07-06, M06B17-32.
Other devices under the same FDA submission
9 devices cleared under P020024 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Amplatzer Piccolo™ 9-PDAP-05-04-LSt. Jude Medical, Inc. · OCCLUDER, PATENT DUCTUS, ARTERIOSUS1 death45 reports
- Amplatzer Piccolo™ 9-PDAP-05-02-LSt. Jude Medical, Inc. · OCCLUDER, PATENT DUCTUS, ARTERIOSUS4 deaths43 reports
- Amplatzer Piccolo™ 9-PDAP-04-02-LSt. Jude Medical, Inc. · OCCLUDER, PATENT DUCTUS, ARTERIOSUS7 deaths31 reports
- Amplatzer Piccolo™ 9-PDAP-04-04-LSt. Jude Medical, Inc. · OCCLUDER, PATENT DUCTUS, ARTERIOSUS3 deaths24 reports
- Amplatzer Piccolo™ 9-PDAP-05-06-LSt. Jude Medical, Inc. · OCCLUDER, PATENT DUCTUS, ARTERIOSUS12 reports
- Amplatzer Piccolo™ 9-PDAP-03-02-LSt. Jude Medical, Inc. · OCCLUDER, PATENT DUCTUS, ARTERIOSUS2 deaths11 reports
- Amplatzer Piccolo™ 9-PDAP-03-04-LSt. Jude Medical, Inc. · OCCLUDER, PATENT DUCTUS, ARTERIOSUS1 death11 reports
- Amplatzer Piccolo™ 9-PDAP-04-06-LSt. Jude Medical, Inc. · OCCLUDER, PATENT DUCTUS, ARTERIOSUS3 reports
- Amplatzer Piccolo™ 9-PDAP-03-06-LSt. Jude Medical, Inc. · OCCLUDER, PATENT DUCTUS, ARTERIOSUS1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 36440).
- Z-0256-2007 · TerminatedAmplatzer Cardiac Septal Occluder delivery system Exchange Systems labeled as follows: a) AMPLATZER Exchange System, Order no. 9-EXCH-6F-180/80, 9-EXCH-8F-180/80, 9EXCH-9F-45/80,…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page