ConMed Laparoscopic Instrument: Babcock Disposable Grasper… recallZ-0096-2007 · Conmed Corporation · initiated Oct 12, 2006
Terminated All recalls posted in October 2006
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Conmed Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
Recall details
- Recall number
- Z-0096-2007
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 12, 2006
- Posted by the FDA
- Oct 27, 2006
- Recalling firm
- Conmed Corporation
- Quantity
- 32 units
- Product code
- GEJ
- FDA submissions
- K943327
- Distribution
- Worldwide
Product description
ConMed Laparoscopic Instrument: Babcock Disposable Grasper, 5mm x 43cm length, REF/Product Code 60-6047-431. --- The product is distributed sterile in heat-sealed tray and labeled with an expiration date that pertains only to the sterility.
Code information (models, lots, serial numbers)
REF/Product Code 60-6047-431, all lot codes manufactured between 08/25/2003 and 03/24/2005. [Lot codes on unopened boxes and packaging contain a lot code in the following form: for example 0409271 - 04 represents the year 2004, 09 represents the month of the year, 27 represents the day of the month, and the last digit 1 represents the manufacturing shift code. Lot codes are located on the device in this form: for example 40927 - 4 represents the year 2004, 09 represents the month of the year, and 27 represents the day of the month.]
Other devices under the same FDA submission
1 device cleared under K943327 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA 60-6045-003Conmed Corporation · Laparoscope, general & plastic surgery1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 36515).
- Z-0083-2007 · TerminatedConMed DetachaTip Laparoscopic Instrument: Allis Multiple Use Grasper, 5mm x 33cm length, REF/Product Code 1-1019. --- The product is distributed sterile in heat-sealed tray and la…
- Z-0084-2007 · TerminatedConMed DetachaTip Laparoscopic Instrument: Allis Multiple Use Grasper, 5mm x 43cm length, REF/Product Code 1-4319. --- The product is distributed sterile in heat-sealed tray and la…
- Z-0085-2007 · TerminatedConMed DetachaTip Laparoscopic Instrument: Babcock Multiple Use Grasper, 5mm x 33cm length, REF/Product Code 1-1005. --- The product is distributed sterile in heat-sealed tray and …
- Z-0086-2007 · TerminatedConMed DetachaTip Laparoscopic Instrument: Babcock Multiple Use Grasper, 5mm x 43cm length, REF/Product Code 1-4305. --- The product is distributed sterile in heat-sealed tray and …
- Z-0087-2007 · TerminatedConMed DetachaTip Laparoscopic Instrument: Curved ''Maryland'' Multiple Use Grasper, 5mm x 33cm length, REF/Product Code 1-1009. --- The product is distributed sterile in heat-seal…
- Z-0088-2007 · TerminatedConMed DetachaTip Laparoscopic Instrument: Curved ''Maryland'' Multiple Use Grasper, 5mm x 43cm length, REF/Product Code 1-4308. --- The product is distributed sterile in heat-seal…
- Z-0089-2007 · TerminatedConMed DetachaTip Laparoscopic Instrument: Endoweave Multiple Use Grasper, 5mm x 43cm length, REF/Product Code 1-4328. --- The product is distributed sterile in heat-sealed tray an…
- Z-0090-2007 · TerminatedConMed DetachaTip Laparoscopic Instrument: Fenestrated (Duckbill) Multiple Use Grasper, 5mm x 43cm, REF/Product Code 1-4307. --- The product is distributed sterile in heat-sealed t…
- Z-0091-2007 · TerminatedConMed DetachaTip Laparoscopic Instrument: Meeker (Right Angle) Multiple Use Grasper, 5mm x 33cm length, REF/Product Code 1-1017. --- The product is distributed sterile in heat-sea…
- Z-0092-2007 · TerminatedConMed DetachaTip Laparoscopic Instrument: Meeker (Right Angle) Multiple Use Grasper, 5mm x 43cm length, REF/Product Code 1-4317. --- The product is distributed sterile in heat-sea…
- Z-0093-2007 · TerminatedConMed Laparoscopic Instrument: Allis Disposable Grasper, 5mm x 33cm length, REF/Product Code 60-6045-006. --- The product is distributed sterile in heat-sealed tray and labeled w…
- Z-0094-2007 · TerminatedConMed Laparoscopic Instrument: Allis Disposable Grasper, 5mm x 43cm length, REF/Product Code 60-6045-436. --- The product is distributed sterile in heat-sealed tray and labeled wi…
- Z-0095-2007 · TerminatedConMed Laparoscopic Instrument: Babcock Disposable Grasper, 5mm x 33cm length, REF/Product Code 60-6047-001. --- The product is distributed sterile in heat-sealed tray and labeled …
- Z-0097-2007 · TerminatedConMed Laparoscopic Instrument: Curved ''Maryland'' Disposable Grasper, 5mm x 33cm length, REF/Product Code 60-6045-002. --- The product is distributed sterile in heat-sealed tray …
- Z-0098-2007 · TerminatedConMed Laparoscopic Instrument: Curved ''Maryland'' Disposable Grasper, 5mm x 43cm length, REF/Product Code 60-6045-432. --- The product is distributed sterile in heat-sealed tray …
- Z-0099-2007 · TerminatedConMed Laparoscopic Instrument: Endoweave Disposable Grasper, 5mm x 33cm length, REF/Product Code 60-6045-005. --- The product is distributed sterile in heat-sealed tray and label…
- Z-0100-2007 · TerminatedConMed Laparoscopic Instrument: Endoweave Disposable Grasper, 5mm x 43cm length, REF/Product Code 60-6045-435. --- The product is distributed sterile in heat-sealed tray and label…
- Z-0101-2007 · TerminatedConMed Laparoscopic Instrument: Fenestrated (Duckbill) Disposable Grasper, 5mm x 33cm length, REF/Product Code 60-6045-003. --- The product is distributed sterile in heat-sealed t…
- Z-0102-2007 · TerminatedConMed Laparoscopic Instrument: Fenestrated (Duckbill) Disposable Grasper, 5mm x 43cm length, REF/Product Code 60-6045-433. --- The product is distributed sterile in heat-sealed tr…
- Z-0103-2007 · TerminatedConMed Laparoscopic Instrument: Meeker (Right Angle) Disposable Grasper, 5mm x 33cm length, REF/Product Code 60-6045-004. --- The product is distributed sterile in heat-sealed tray…
- Z-0104-2007 · TerminatedConMed Laparoscopic Instrument: Meeker (Right Angle) Disposable Grasper, 5mm x 43cm length, REF/Product Code 60-6045-434. --- The product is distributed sterile in heat-sealed tray…
- Z-0105-2007 · TerminatedConMed Laparoscopic Instrument: Allis Disposable Dissector, 5mm x 33cm length, REF/Product Code 60-6040-006. --- The product is distributed sterile in heat-sealed tray and labeled …
- Z-0106-2007 · TerminatedConMed Laparoscopic Instrument: Allis Disposable Dissector, 5mm x 43cm length, REF/Product Code 60-6040-436. --- The product is distributed sterile in heat-sealed tray and labeled …
- Z-0107-2007 · TerminatedConMed Laparoscopic Instrument: Curved ''Maryland'' Disposable Dissector, 5mm x 33cm length, REF/Product Code 60-6040-002. --- The product is distributed sterile in heat-sealed t…
- Z-0108-2007 · TerminatedConMed Laparoscopic Instrument: Curved ''Maryland'' Disposable Dissector, 5mm x 43cm length, REF/Product Code 60-6040-432. --- The product is distributed sterile in heat-sealed tr…
- Z-0109-2007 · TerminatedConMed Laparoscopic Instrument: Endoweave Disposable Dissector, 5mm x 33cm length, REF/Product Code 60-6040-005. --- The product is distributed sterile in heat-sealed tray and lab…
- Z-0110-2007 · TerminatedConMed Laparoscopic Instrument: Endoweave Disposable Dissector, 5mm x 43cm length, REF/Product Code 60-6040-435. --- The product is distributed sterile in heat-sealed tray and labe…
- Z-0111-2007 · TerminatedConMed Laparoscopic Instrument: Fenestrated (Duckbill) Disposable Dissector, 5mm x 33cm length, REF/Product Code 60-6040-003. --- The product is distributed sterile in heat-sealed…
- Z-0112-2007 · TerminatedConMed Laparoscopic Instrument: Fenestrated (Duckbill) Disposable Dissector, 5mm x 43cm length, REF/Product Code 60-6040-433. --- The product is distributed sterile in heat-sealed…
- Z-0113-2007 · TerminatedConMed Laparoscopic Instrument: Meeker (Right Angle) Disposable Dissector, 5mm x 33cm length, REF/Product Code 60-6040-004. --- The product is distributed sterile in heat-sealed tr…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page