Why it was recalled

Plastic bottle samples assigned less sensitive default glass bottle algorithm instead of the plastic bottle algorithm on BacT/ALERT Microbial Detection Systems.

FDA root cause category: Other

What the company is doing

Consignees were notified via A Field Corrective Action Notification on/about 10/18/2006. A second notification was sent on October 31, 2006.

Recall details

Recall number
Z-0773-2007
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 14, 2006
Posted by the FDA
May 3, 2007
Recalling firm
bioMerieux, Inc.
Quantity
10790 units, total for all models
Product code
MDB
FDA submissions
K903505
Distribution
Worldwide, including USA, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dutch Antilles, Egypt, Ecuador, Estonia, Faeroe Islands, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Kenya, Korea, Latvia, Lithuania, Luxembourg, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania , Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Suriname, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.

Product description

BacT/ALERT Classic System Software, Blood Culturing System

Code information (models, lots, serial numbers)

All Models

Other devices under the same FDA submission

5 devices cleared under K903505 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 36826).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page