Infinity Docking Station (IDS) Gamma/Gamma XL and Infinity… recallZ-0302-2007 · Draeger Medical, Inc. · initiated Nov 17, 2006

Terminated All recalls posted in December 2006

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Draeger Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Top cover of assembly may separate. The situation can result in the monitor falling, which could result in an injury to the clinician or patient.

FDA root cause category: Other

What the company is doing

The recaling firm issued a recall letter to the US customers 11/17/2006 via certified mail. An authorized service representative will inspect each potentially affected unit. If a unit is found to have an affected docking station, the unit will be repaired or replaced free or charge. Notification to international customers will occur through the international subsidiaries.

Recall details

Recall number
Z-0302-2007
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Nov 17, 2006
Posted by the FDA
Dec 19, 2006
Recalling firm
Draeger Medical, Inc.
Quantity
23,345 units - 4336 US
Product code
MHX
FDA submissions
K053484
Distribution
Nationwide

Product description

Infinity Docking Station (IDS) Gamma/Gamma XL and Infinity Vista Series patient Monitors. Model numbers 4715319, 5206110, 5732388, and 7489375.

Code information (models, lots, serial numbers)

Model numbers 4715319, 5206110, 5732388, and 7489375.

Other devices under the same FDA submission

1 device cleared under K053484 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.