Skip to content

Anspach BlackMax Pneumatic Motor System recallZ-1039-2008 · The Anspach Effort, Inc. · initiated Dec 18, 2006

Terminated All recalls posted in February 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from The Anspach Effort, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Incorrect Sterilization Process --The Operating Manual for the product contains an incorrect statement regarding the sterilization of the Autolube foot control. The manual states that the product can withstand ETO sterilization, but there is no data to support this process.

FDA root cause category: Labeling False and Misleading

What the company is doing

Anspach issued an Urgent Device Correction letter on 11/27/06, to all manual holders advising them of the error and requested that they update the manual with the corrected instructions. A reply receipt confirming notification was included.

Recall details

Recall number
Z-1039-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Dec 18, 2006
Posted by the FDA
Feb 13, 2008
Recalling firm
The Anspach Effort, Inc.
Quantity
7,399 total units
Product code
ERL
FDA submissions
K831756, K965080
Distribution
Worldwide Distribution

Product description

Anspach BlackMax Pneumatic Motor System; ENT Surgical Drill, including handpiece and Accessories Attached. All Models; Black Motors; Anspach Effort, Inc., Palm Beach FL 33410.

Code information (models, lots, serial numbers)

All Serial Numbers.

Other devices under the same FDA submission

2 devices cleared under K831756, K965080 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page