Skip to content

PADPRO ADULT Radiotranslucent Pad w/Zoll connector.… recallZ-0756-2007 · Conmed Corporation · initiated Mar 8, 2007

Terminated All recalls posted in April 2007

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Conmed Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Firm was notified of three incidents over the past 15 months in which customers opened the pouch & found an incorrect connector. This error makes the product unusable. The Zoll and Medtronic connectors are not interchangeable and are unique to their specific device.

FDA root cause category: Other

What the company is doing

On 3/08/07, "URGENT DEVICE RECALL" letters (dated 3/05/07) will be sent via UPS to the US distributors and direct user accounts. The letters inform them of the problem and of the recall. The letters instruct them to immediately discontinue using the affected lots, inspect their stocks, and to return them as soon as possible. The letters instruct the consignees to notify their customers of the recall of the devices to the user level if they have been distributed, and request the return of the products. Additionally, when requested, ConMed will notify distributor's customers (sub-accounts) by the same means. Returned devices will be automatically replaced by ConMed Corporation. A separate letter is being sent to the foreign accounts today instructing them to return the devices back to ConMed Corporation for final disposition. The letters also instruct the consignees to complete the accompanying reply form and fax it to Nancy Crisino at (315) 624-3089. ConMed has contracted Stericycle in Indianapolis, IN, to handle the recall and returns from the domestic accounts.

Recall details

Recall number
Z-0756-2007
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Mar 8, 2007
Posted by the FDA
Apr 24, 2007
Recalling firm
Conmed Corporation
Quantity
9712 pairs
Product code
MLN
FDA submissions
K020203
Distribution
Worldwide, including USA, Canada, Romania, Chile, and Belgium

Product description

PADPRO ADULT Radiotranslucent Pad w/Zoll connector. Defibrillation/Pacing/Cardioversion/Monitoring Electrodes, REF/Catalog # 2516Z. RX ONLY. Made in USA for: CONMED CORPORATION Utica, New York, 13502.

Code information (models, lots, serial numbers)

Lots Y040605-1, Y040605-2, Y052605-4, Y120805-2, Y111605-3, Y092805-6, Y010506-10, Y010506-9.

Other devices under the same FDA submission

3 devices cleared under K020203 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 37495).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page