Blood glucose meter, marketed with multiple brand names… recallZ-0112-2008 · Abbott Diabetes Care, Inc. · initiated Aug 2, 2007
Terminated All recalls posted in February 2008
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Diabetes Care, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Damage could lead to no display. Meters manufactured after January 31, 2007 could exhibit meter display damage if dropped on a hard surface. These meters could exhibit unreadable lot number fields and date/time fields, in addition to complete blanking of the numerical reading portion of the display.
FDA root cause category: Component design/selection
What the company is doing
Abbott sent Urgent Medical Device Correction letters, dated 31Aug07, to registered users/customers via FedEx or mail. The letter instructs users to verify that their display is correct, and if not, to discontinue use and call Abbott Customer Care at 1-877-844-4404 for a new meter. Warranty kits will have fliers added to them to alert users. Email and telephone contacts may be made as needed. U.S. consignees and registered users will be contacted by letter, telephone call or visit starting 23Aug07. Worldwide, customer lists will be created and communications tracked according to local requirements. Abbott issued a Press Release August 31, 2007 also.
Recall details
- Recall number
- Z-0112-2008
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Aug 2, 2007
- Posted by the FDA
- Feb 14, 2008
- Recalling firm
- Abbott Diabetes Care, Inc.
- Quantity
- 882,601 total meters distributed, 316,439 in the U.S. and 566,162 outside of the U.S. including Mexico and Canada
- Product code
- NBW
- FDA submissions
- K033845, K040814
- Distribution
- Worldwide Distribution - USA, Canada, Mexico, Australia, Malaysia, Philippines, Germany, New Zealand, China, India, Singapore, Hong Kong, Taiwan, Argentina, Brazil, Chile, Colombia, Costa Rica, Ecuador, Mexico, Peru, Puerto Rico, Uruguay, Venezuela, and Japan.
Product description
Blood glucose meter, marketed with multiple brand names including: Precision Xtra New and Improved meter, Precision Xceed meter, Optium Xceed meter, Optium meter, ReliOn Ultima meter, Rite Aid meter, Kroger meter, Omron HEA-214 meter, Optium Xido meter, Boots meter.
Code information (models, lots, serial numbers)
Show the full code information (137k characters)
Model Number Model Name Kit Lot Number Meter Lot Number 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K094D XCB523 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K094D XCB559 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K094D XCB562 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K094D XCB571 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K116D XCB589 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K116D XCB590 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K116D XCB603 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K131D XCB676 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K131D XCB678 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K131D XCB709 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K131D XCB712 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K134D XCB590 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K134D XCB603 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K134D XCB641 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K169D XCB812 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K169D XCB813 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K169D XCB821 70055-50 ReliOn Ultima Blood Glucose Monitoring System 01K204D XCB850 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K046D XCB335 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K046D XCB336 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K046D XCB337 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K046D XCB358 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K047D XCB335 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K047D XCB336 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K047D XCB337 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K047D XCB358 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K068D XCB469 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K088D XCB518 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K088D XCB519 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K093D XCB519 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K093D XCB520 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K094D XCB560 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K094D XCB561 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K094D XCB562 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K116D XCB588 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K116D XCB590 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K130D XCB676 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K130D XCB678 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K130D XCB709 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K130D XCB712 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K131D XCB675 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K131D XCB677 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K193D XCB826 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K193D XCB847 70055-50 ReliOn Ultima Blood Glucose Monitoring System 02K193D XCB848 70055-50 ReliOn Ultima Blood Glucose Monitoring System 03K078D XCB502 70055-50 ReliOn Ultima Blood Glucose Monitoring System 03K078D XCB503 70055-50 ReliOn Ultima Blood Glucose Monitoring System 03K087D XCB518 70055-50 ReliOn Ultima Blood Glucose Monitoring System 03K087D XCB519 70055-50 ReliOn Ultima Blood Glucose Monitoring System 03K093D XCB523 70055-50 ReliOn Ultima Blood Glucose Monitoring System 03K093D XCB559 70055-50 ReliOn Ultima Blood Glucose Monitoring System 03K093D XCB562 70055-50 ReliOn Ultima Blood Glucose Monitoring System 03K093D XCB571 70055-50 ReliOn Ultima Blood Glucose Monitoring System 03K094D XCB561 70055-50 ReliOn Ultima Blood Glucose Monitoring System 03K134D XCB675 70055-50 ReliOn Ultima Blood Glucose Monitoring System 03K169D XCB812 70055-50 ReliOn Ultima Blood Glucose Monitoring System 03K170D XCB814 70055-50 ReliOn Ultima Blood Glucose Monitoring System 03K170D XCB821 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K071D XCB469 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K093D XCB560 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K093D XCB561 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K093D XCB562 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K094D XCB520 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K094D XCB523 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K099D XCB561 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K099D XCB572 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K099D XCB581 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K099D XCB583 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K131D XCB677 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K131D XCB710 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K131D XCB712 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K157D XCB754 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K187D XCB823 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K187D XCB824 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K187D XCB825 70055-50 ReliOn Ultima Blood Glucose Monitoring System 04K187D XCB826 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K075D XCB502 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K075D XCB503 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K080D XCB470 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K080D XCB516 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K081D XCB470 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K081D XCB516 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K088D XCB518 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K089D XCB518 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K092D XCB560 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K092D XCB561 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K092D XCB562 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K115D XCB589 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K115D XCB590 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K115D XCB603 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K130D XCB677 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K130D XCB710 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K130D XCB712 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K131D XCB590 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K131D XCB603 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K131D XCB641 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K170D XCB812 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K170D XCB813 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K170D XCB821 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K186D XCB823 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K186D XCB824 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K186D XCB825 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K186D XCB826 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K190D XCB823 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K190D XCB824 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K190D XCB825 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K190D XCB826 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K192D XCB826 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K192D XCB847 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K192D XCB848 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K199D XCB824 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K199D XCB848 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K199D XCB850 70055-50 ReliOn Ultima Blood Glucose Monitoring System 05K199D XCB852 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K057D XCB413 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K075D XCB470 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K093D XCB520 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K093D XCB523 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K096D XCB561 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K096D XCB572 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K096D XCB581 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K096D XCB583 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K131D XCB641 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K131D XCB709 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K133D XCB641 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K133D XCB709 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K156D XCB754 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K180D XCB811 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K180D XCB814 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K184D XCB823 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K184D XCB824 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K184D XCB825 70055-50 ReliOn Ultima Blood Glucose Monitoring System 06K184D XCB826 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K078D XCB501 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K078D XCB502 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K078D XCB507 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K092D XCB519 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K092D XCB520 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K109D XCB583 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K109D XCB588 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K109D XCB600 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K110D XCB583 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K110D XCB588 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K110D XCB600 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K115D XCB588 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K115D XCB590 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K129D XCB676 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K129D XCB678 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K129D XCB709 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K129D XCB712 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K130D XCB675 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K130D XCB677 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K186D XCB824 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K191D XCB826 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K191D XCB847 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K191D XCB848 70055-50 ReliOn Ultima Blood Glucose Monitoring System 07K204D XCB850 70055-50 ReliOn Ultima Blood Glucose Monitoring System 08K075D XCB501 70055-50 ReliOn Ultima Blood Glucose Monitoring System 08K075D XCB502 70055-50 ReliOn Ultima Blood Glucose Monitoring System 08K075D XCB507 70055-50 ReliOn Ultima Blood Glucose Monitoring System 08K093D XCB561 70055-50 ReliOn Ultima Blood Glucose Monitoring System 08K096D XCB600 70055-50 ReliOn Ultima Blood Glucose Monitoring System 08K131D XCB675 70055-50 ReliOn Ultima Blood Glucose Monitoring System 08K165D XCB812 70055-50 ReliOn Ultima Blood Glucose Monitoring System 08K169D XCB814 70055-50 ReliOn Ultima Blood Glucose Monitoring System 08K169D XCB821 70055-50 ReliOn Ultima Blood Glucose Monitoring System 09K045D XCB335 70055-50 ReliOn Ultima Blood Glucose Monitoring System 09K045D XCB336 70055-50 ReliOn Ultima Blood Glucose Monitoring System 09K045D XCB337 70055-50 ReliOn Ultima Blood Glucose Monitoring System 09K045D XCB358 70055-50 ReliOn Ultima Blood Glucose Monitoring System 09K074D XCB470 70055-50 ReliOn Ultima Blood Glucose Monitoring System 09K130D XCB590 70055-50 ReliOn Ultima Blood Glucose Monitoring System 09K130D XCB603 70055-50 ReliOn Ultima Blood Glucose Monitoring System 09K130D XCB641 70055-50 ReliOn Ultima Blood Glucose Monitoring System 09K166D XCB812 70055-50 ReliOn Ultima Blood Glucose Monitoring System 09K180D XCB811 70055-50 ReliOn Ultima Blood Glucose Monitoring System 09K180D XCB824 70055-50 ReliOn Ultima Blood Glucose Monitoring System 09K190D XCB826 70055-50 ReliOn Ultima Blood Glucose Monitoring System 09K190D XCB847 70055-50 ReliOn Ultima Blood Glucose Monitoring System 09K190D XCB848 70055-50 ReliOn Ultima Blood Glucose Monitoring System 09K201D XCB850 70055-50 ReliOn Ultima Blood Glucose Monitoring System 10K095D XCB561 70055-50 ReliOn Ultima Blood Glucose Monitoring System 10K095D XCB572 70055-50 ReliOn Ultima Blood Glucose Monitoring System 10K095D XCB581 70055-50 ReliOn Ultima Blood Glucose Monitoring System 10K095D XCB583 70055-50 ReliOn Ultima Blood Glucose Monitoring System 10K166D XCB812 70055-50 ReliOn Ultima Blood Glucose Monitoring System 10K166D XCB813 70055-50 ReliOn Ultima Blood Glucose Monitoring System 10K166D XCB821 70055-50 ReliOn Ultima Blood Glucose Monitoring System 10K172D XCB812 70055-50 ReliOn Ultima Blood Glucose Monitoring System 11K198D XCB824 70055-50 ReliOn Ultima Blood Glucose Monitoring System 11K198D XCB848 70055-50 ReliOn Ultima Blood Glucose Monitoring System 11K198D XCB850 70055-50 ReliOn Ultima Blood Glucose Monitoring System 11K198D XCB852 70079-35 Precision Xceed Diabetes Management System 01K180C XCB555 70079-35 Precision Xceed Diabetes Management System 01K180C XCB565 70079-35 Precision Xceed Diabetes Management System 02K149C XCB491 70079-35 Precision Xceed Diabetes Management System 02K149C XCB511 70079-35 Precision Xceed Diabetes Management System 03K149C XCB491 70079-35 Precision Xceed Diabetes Management System 04K145C XCB491 70079-35 Precision Xceed Diabetes Management System 04K177C XCB564 70079-35 Precision Xceed Diabetes Management System 04K177C XCB565 70079-35 Precision Xceed Diabetes Management System 05K072C XCB464 70079-35 Precision Xceed Diabetes Management System 08K145C XCB491 70079-35 Precision Xceed Diabetes Management System 08K145C XCB511 70079-35 Precision Xceed Diabetes Management System 09K178C XCB563 70079-35 Precision Xceed Diabetes Management System 09K180C XCB563 70079-35 Precision Xceed Diabetes Management System 09K180C XCB565 70079-35 Precision Xceed Diabetes Management System 10K071C XCB464 70079-35 Precision Xceed Diabetes Management System 10K071C XCB465 70079-35 Precision Xceed Diabetes Management System 10K179C XCB563 70079-35 Precision Xceed Diabetes Management System 10K179C XCB610 70079-35 Precision Xceed Diabetes Management System 11K178C XCB564 70079-35 Precision Xceed Diabetes Management System 11K178C XCB565 70079-35 Precision Xceed Diabetes Management System 12K179C XCB563 70079-35 Precision Xceed Diabetes Management System 13K177C XCB564 70079-35 Precision Xceed Diabetes Management System 13K177C XCB574 70079-35 Precision Xceed Diabetes Management System 13K177C XCB580 70079-35 Precision Xceed Diabetes Management System 13K180C XCB563 70079-35 Precision Xceed Diabetes Management System 13K180C XCB610 70079-35 Precision Xceed Diabetes Management System 14K071C XCB464 70079-35 Precision Xceed Diabetes Management System 14K179C XCB563 70079-35 Precision Xceed Diabetes Management System 14K179C XCB565 70079-35 Precision Xceed Diabetes Management System 15K178C XCB564 70079-35 Precision Xceed Diabetes Management System 15K178C XCB574 70079-35 Precision Xceed Diabetes Management System 15K178C XCB580 70079-35 Precision Xceed Diabetes Management System 15K179C XCB555 70079-35 Precision Xceed Diabetes Management System 15K179C XCB565 70079-35 Precision Xceed Diabetes Management System 16K176C XCB564 70079-35 Precision Xceed Diabetes Management System 16K176C XCB574 70079-35 Precision Xceed Diabetes Management System 16K176C XCB580 70082-35 Precision Xceed Diabetes Management System 02K060C XC1618 70082-35 Precision Xceed Diabetes Management System 02K193C XC1628 70082-35 Precision Xceed Diabetes Management System 05K192C XC1627 70082-35 Precision Xceed Diabetes Management System 05K192C XC1628 70082-35 Precision Xceed Diabetes Management System 06K171C XC1624 70082-35 Precision Xceed Diabetes Management System 07K193C XC1619 70082-35 Precision Xceed Diabetes Management System 08K176C XC1618 70082-35 Precision Xceed Diabetes Management System 09K176C XC1624 70082-35 Precision Xceed Diabetes Management System 10K079C XC1618 70082-35 Precision Xceed Diabetes Management System 12K159C XC1623 70082-35 Precision Xceed Diabetes Management System 12K159C XC1625 70082-35 Precision Xceed Diabetes Management System 13K191C XC1627 70082-35 Precision Xceed Diabetes Management System 13K191C XC1628 70082-35 Precision Xceed Diabetes Management System 15K159C XC1624 70082-35 Precision Xceed Diabetes Management System 15K173C XC1618 70082-35 Precision Xceed Diabetes Management System 15K177C XC1624 70082-35 Precision Xceed Diabetes Management System 15K192C XC1628 70087-35 Optium Xceed Diabetes Monitoring System 01K141C XCB532 70087-35 Optium Xceed Diabetes Monitoring System 01K157C XCB608 70087-35 Optium Xceed Diabetes Monitoring System 01K157C XCB609 70087-35 Optium Xceed Diabetes Monitoring System 01K159C XCB609 70087-35 Optium Xceed Diabetes Monitoring System 01K165C XCB758 70087-35 Optium Xceed Diabetes Monitoring System 01K169C XCB758 70087-35 Optium Xceed Diabetes Monitoring System 01K169C XCB775 70087-35 Optium Xceed Diabetes Monitoring System 01K171C XCB761 70087-35 Optium Xceed Diabetes Monitoring System 01K173C XCB397 70087-35 Optium Xceed Diabetes Monitoring System 01K173C XCB398 70087-35 Optium Xceed Diabetes Monitoring System 01K192C XCB777 70087-35 Optium Xceed Diabetes Monitoring System 01K192C XCB781 70087-35 Optium Xceed Diabetes Monitoring System 01K193C XCB776 70087-35 Optium Xceed Diabetes Monitoring System 01K193C XCB777 70087-35 Optium Xceed Diabetes Monitoring System 02K138C XCB527 70087-35 Optium Xceed Diabetes Monitoring System 02K138C XCB531 70087-35 Optium Xceed Diabetes Monitoring System 02K141C XCB521 70087-35 Optium Xceed Diabetes Monitoring System 02K141C XCB522 70087-35 Optium Xceed Diabetes Monitoring System 02K159C XCB608 70087-35 Optium Xceed Diabetes Monitoring System 02K197C XCB783 70087-35 Optium Xceed Diabetes Monitoring System 03K131C XCB399 70087-35 Optium Xceed Diabetes Monitoring System 03K131C XCB553 70087-35 Optium Xceed Diabetes Monitoring System 03K138C XCB522 70087-35 Optium Xceed Diabetes Monitoring System 03K142C XCB531 70087-35 Optium Xceed Diabetes Monitoring System 03K165C XCB527 70087-35 Optium Xceed Diabetes Monitoring System 03K165C XCB608 70087-35 Optium Xceed Diabetes Monitoring System 03K165C XCB645 70087-35 Optium Xceed Diabetes Monitoring System 03K165C XCB761 70087-35 Optium Xceed Diabetes Monitoring System 04K130C XCB354 70087-35 Optium Xceed Diabetes Monitoring System 04K130C XCB356 70087-35 Optium Xceed Diabetes Monitoring System 04K131C XCB553 70087-35 Optium Xceed Diabetes Monitoring System 04K138C XCB531 70087-35 Optium Xceed Diabetes Monitoring System 04K141C XCB531 70087-35 Optium Xceed Diabetes Monitoring System 04K141C XCB532 70087-35 Optium Xceed Diabetes Monitoring System 04K173C XCB398 70087-35 Optium Xceed Diabetes Monitoring System 04K173C XCB451 70087-35 Optium Xceed Diabetes Monitoring System 04K194C XCB779 70087-35 Optium Xceed Diabetes Monitoring System 04K194C XCB782 70087-35 Optium Xceed Diabetes Monitoring System 05K131C XCB356 70087-35 Optium Xceed Diabetes Monitoring System 05K131C XCB399 70087-35 Optium Xceed Diabetes Monitoring System 05K138C XCB521 70087-35 Optium Xceed Diabetes Monitoring System 05K138C XCB645 70087-35 Optium Xceed Diabetes Monitoring System 05K141C XCB522 70087-35 Optium Xceed Diabetes Monitoring System 05K141C XCB532 70087-35 Optium Xceed Diabetes Monitoring System 05K165C XCB759 70087-35 Optium Xceed Diabetes Monitoring System 05K165C XCB760 70087-35 Optium Xceed Diabetes Monitoring System 05K166C XCB758 70087-35 Optium Xceed Diabetes Monitoring System 05K176C XCB448 70087-35 Optium Xceed Diabetes Monitoring System 05K176C XCB455 70087-35 Optium Xceed Diabetes Monitoring System 05K177C XCB448 70087-35 Optium Xceed Diabetes Monitoring System 05K177C XCB451 70087-35 Optium Xceed Diabetes Monitoring System 06K138C XCB531 70087-35 Optium Xceed Diabetes Monitoring System 06K141C XCB531 70087-35 Optium Xceed Diabetes Monitoring System 06K143C XCB531 70087-35 Optium Xceed Diabetes Monitoring System 06K158C XCB608 70087-35 Optium Xceed Diabetes Monitoring System 06K158C XCB609 70087-35 Optium Xceed Diabetes Monitoring System 06K159C XCB608 70087-35 Optium Xceed Diabetes Monitoring System 06K159C XCB609 70087-35 Optium Xceed Diabetes Monitoring System 06K176C XCB451 70087-35 Optium Xceed Diabetes Monitoring System 06K180C XCB397 70087-35 Optium Xceed Diabetes Monitoring System 06K193C XCB777 70087-35 Optium Xceed Diabetes Monitoring System 06K193C XCB781 70087-35 Optium Xceed Diabetes Monitoring System 07K158C XCB551 70087-35 Optium Xceed Diabetes Monitoring System 07K158C XCB609 70087-35 Optium Xceed Diabetes Monitoring System 07K164C XCB098 70087-35 Optium Xceed Diabetes Monitoring System 07K164C XCB758 70087-35 Optium Xceed Diabetes Monitoring System 07K164C XCB760 70087-35 Optium Xceed Diabetes Monitoring System 07K171C XCB759 70087-35 Optium Xceed Diabetes Monitoring System 07K185C XCB057 70087-35 Optium Xceed Diabetes Monitoring System 07K185C XCB450 70087-35 Optium Xceed Diabetes Monitoring System 07K185C XCB454 70087-35 Optium Xceed Diabetes Monitoring System 08K138C XCB522 70087-35 Optium Xceed Diabetes Monitoring System 08K138C XCB532 70087-35 Optium Xceed Diabetes Monitoring System 08K158C XCB609 70087-35 Optium Xceed Diabetes Monitoring System 08K192C XCB777 70087-35 Optium Xceed Diabetes Monitoring System 08K193C XCB782 70087-35 Optium Xceed Diabetes Monitoring System 08K193C XCB783 70087-35 Optium Xceed Diabetes Monitoring System 09K158C XCB608 70087-35 Optium Xceed Diabetes Monitoring System 09K172C XCB759 70087-35 Optium Xceed Diabetes Monitoring System 09K173C XCB455 70087-35 Optium Xceed Diabetes Monitoring System 09K193C XCB776 70087-35 Optium Xceed Diabetes Monitoring System 10K113C XCB135 70087-35 Optium Xceed Diabetes Monitoring System 10K113C XCB356 70087-35 Optium Xceed Diabetes Monitoring System 10K130C XCB553 70087-35 Optium Xceed Diabetes Monitoring System 10K134C XCB356 70087-35 Optium Xceed Diabetes Monitoring System 10K134C XCB399 70087-35 Optium Xceed Diabetes Monitoring System 11K138C XCB531 70087-35 Optium Xceed Diabetes Monitoring System 11K138C XCB532 70087-35 Optium Xceed Diabetes Monitoring System 11K164C XCB759 70087-35 Optium Xceed Diabetes Monitoring System 11K165C XCB551 70087-35 Optium Xceed Diabetes Monitoring System 11K193C XCB779 70087-35 Optium Xceed Diabetes Monitoring System 11K193C XCB782 70087-35 Optium Xceed Diabetes Monitoring System 12K130C XCB399 70087-35 Optium Xceed Diabetes Monitoring System 12K130C XCB553 70087-35 Optium Xceed Diabetes Monitoring System 12K138C XCB521 70087-35 Optium Xceed Diabetes Monitoring System 12K138C XCB522 70087-35 Optium Xceed Diabetes Monitoring System 12K165C XCB396 70087-35 Optium Xceed Diabetes Monitoring System 12K166C XCB396 70087-35 Optium Xceed Diabetes Monitoring System 12K173C XCB448 70087-35 Optium Xceed Diabetes Monitoring System 12K178C XCB454 70087-35 Optium Xceed Diabetes Monitoring System 12K178C XCB455 70087-35 Optium Xceed Diabetes Monitoring System 12K192C XCB776 70087-35 Optium Xceed Diabetes Monitoring System 12K192C XCB777 70087-35 Optium Xceed Diabetes Monitoring System 13K130C XCB356 70087-35 Optium Xceed Diabetes Monitoring System 13K130C XCB399 70087-35 Optium Xceed Diabetes Monitoring System 13K164C XCB527 70087-35 Optium Xceed Diabetes Monitoring System 13K164C XCB608 70087-35 Optium Xceed Diabetes Monitoring System 13K164C XCB645 70087-35 Optium Xceed Diabetes Monitoring System 13K164C XCB761 70087-35 Optium Xceed Diabetes Monitoring System 13K169C XCB397 70087-35 Optium Xceed Diabetes Monitoring System 13K173C XCB451 70087-35 Optium Xceed Diabetes Monitoring System 13K178C XCB448 70087-35 Optium Xceed Diabetes Monitoring System 13K179C XCB448 70087-35 Optium Xceed Diabetes Monitoring System 13K193C XCB777 70087-35 Optium Xceed Diabetes Monitoring System 13K194C XCB783 70087-35 Optium Xceed Diabetes Monitoring System 13K197C XCB776 70087-35 Optium Xceed Diabetes Monitoring System 14K137C XCB522 70087-35 Optium Xceed Diabetes Monitoring System 14K138C XCB532 70087-35 Optium Xceed Diabetes Monitoring System 14K164C XCB759 70087-35 Optium Xceed Diabetes Monitoring System 14K164C XCB760 70087-35 Optium Xceed Diabetes Monitoring System 14K165C XCB758 70087-35 Optium Xceed Diabetes Monitoring System 14K172C XCB761 70087-35 Optium Xceed Diabetes Monitoring System 15K137C XCB531 70087-35 Optium Xceed Diabetes Monitoring System 15K164C XCB758 70087-35 Optium Xceed Diabetes Monitoring System 15K170C XCB396 70087-35 Optium Xceed Diabetes Monitoring System 15K170C XCB397 70087-35 Optium Xceed Diabetes Monitoring System 16K137C XCB521 70087-35 Optium Xceed Diabetes Monitoring System 16K137C XCB645 70087-35 Optium Xceed Diabetes Monitoring System 16K173C XCB448 70087-35 Optium Xceed Diabetes Monitoring System 16K173C XCB455 70087-35 Optium Xceed Diabetes Monitoring System 17K176C XCB448 70087-35 Optium Xceed Diabetes Monitoring System 17K176C XCB451 70087-35 Optium Xceed Diabetes Monitoring System 18K172C XCB398 70087-35 Optium Xceed Diabetes Monitoring System 18K172C XCB451 70087-35 Optium Xceed Diabetes Monitoring System 18K191C XCB777 70087-35 Optium Xceed Diabetes Monitoring System 18K191C XCB781 70087-35 Optium Xceed Diabetes Monitoring System 19K165C XCB759 70087-35 Optium Xceed Diabetes Monitoring System 19K170C XCB761 70087-35 Optium Xceed Diabetes Monitoring System 20K170C XCB397 70089-01 Optium Xceed Diabetes Monitoring System 01K130C XCB354 70089-01 Optium Xceed Diabetes Monitoring System 01K130C XCB357 70089-01 Optium Xceed Diabetes Monitoring System 01K152C XCB535 70089-01 Optium Xceed Diabetes Monitoring System 01K156C XCB551 70089-01 Optium Xceed Diabetes Monitoring System 01K176C XCB454 70089-01 Optium Xceed Diabetes Monitoring System 01K186C XCB449 70089-01 Optium Xceed Diabetes Monitoring System 01K186C XCB453 70089-01 Optium Xceed Diabetes Monitoring System 02K152C XCB526 70089-01 Optium Xceed Diabetes Monitoring System 02K152C XCB535 70089-01 Optium Xceed Diabetes Monitoring System 02K155C XCB537 70089-01 Optium Xceed Diabetes Monitoring System 02K155C XCB551 70089-01 Optium Xceed Diabetes Monitoring System 02K191C XCB775 70089-01 Optium Xceed Diabetes Monitoring System 02K191C XCB778 70089-01 Optium Xceed Diabetes Monitoring System 03K128C XCB315 70089-01 Optium Xceed Diabetes Monitoring System 03K128C XCB355 70089-01 Optium Xceed Diabetes Monitoring System 03K130C XCB354 70089-01 Optium Xceed Diabetes Monitoring System 03K152C XCB527 70089-01 Optium Xceed Diabetes Monitoring System 03K152C XCB645 70089-01 Optium Xceed Diabetes Monitoring System 03K155C XCB537 70089-01 Optium Xceed Diabetes Monitoring System 03K191C XCB778 70089-01 Optium Xceed Diabetes Monitoring System 04K071C XCB302 70089-01 Optium Xceed Diabetes Monitoring System 04K071C XCB399 70089-01 Optium Xceed Diabetes Monitoring System 04K129C XCB135 70089-01 Optium Xceed Diabetes Monitoring System 04K129C XCB317 70089-01 Optium Xceed Diabetes Monitoring System 04K129C XCB645 70089-01 Optium Xceed Diabetes Monitoring System 04K191C XCB778 70089-01 Optium Xceed Diabetes Monitoring System 04K191C XCB780 70089-01 Optium Xceed Diabetes Monitoring System 04K192C XCB774 70089-01 Optium Xceed Diabetes Monitoring System 04K192C XCB781 70089-01 Optium Xceed Diabetes Monitoring System 05K128C XCB355 70089-01 Optium Xceed Diabetes Monitoring System 05K156C XCB551 70089-01 Optium Xceed Diabetes Monitoring System 05K186C XCB450 70089-01 Optium Xceed Diabetes Monitoring System 05K186C XCB454 70089-01 Optium Xceed Diabetes Monitoring System 05K191C XCB780 70089-01 Optium Xceed Diabetes Monitoring System 06K128C XCB355 70089-01 Optium Xceed Diabetes Monitoring System 06K128C XCB357 70089-01 Optium Xceed Diabetes Monitoring System 06K151C XCB526 70089-01 Optium Xceed Diabetes Monitoring System 06K151C XCB535 70089-01 Optium Xceed Diabetes Monitoring System 06K179C XCB454 70089-01 Optium Xceed Diabetes Monitoring System 06K187C XCB449 70089-01 Optium Xceed Diabetes Monitoring System 06K187C XCB453 70089-01 Optium Xceed Diabetes Monitoring System 07K151C XCB526 70089-01 Optium Xceed Diabetes Monitoring System 07K155C XCB527 70089-01 Optium Xceed Diabetes Monitoring System 07K155C XCB535 70089-01 Optium Xceed Diabetes Monitoring System 07K157C XCB309 70089-01 Optium Xceed Diabetes Monitoring System 07K157C XCB551 70089-01 Optium Xceed Diabetes Monitoring System 07K186C XCB449 70089-01 Optium Xceed Diabetes Monitoring System 07K191C XCB781 70089-01 Optium Xceed Diabetes Monitoring System 08K128C XCB357 70089-01 Optium Xceed Diabetes Monitoring System 08K129C XCB354 70089-01 Optium Xceed Diabetes Monitoring System 08K152C XCB526 70089-01 Optium Xceed Diabetes Monitoring System 08K152C XCB537 70089-01 Optium Xceed Diabetes Monitoring System 08K156C XCB527 70089-01 Optium Xceed Diabetes Monitoring System 08K156C XCB535 70089-01 Optium Xceed Diabetes Monitoring System 08K157C XCB527 70089-01 Optium Xceed Diabetes Monitoring System 08K157C XCB535 70089-01 Optium Xceed Diabetes Monitoring System 08K177C XCB397 70089-01 Optium Xceed Diabetes Monitoring System 08K191C XCB774 70089-01 Optium Xceed Diabetes Monitoring System 08K191C XCB780 70089-01 Optium Xceed Diabetes Monitoring System 09K152C XCB527 70089-01 Optium Xceed Diabetes Monitoring System 09K152C XCB535 70089-01 Optium Xceed Diabetes Monitoring System 09K191C XCB774 70089-01 Optium Xceed Diabetes Monitoring System 09K192C XCB774 70089-01 Optium Xceed Diabetes Monitoring System 09K192C XCB780 70089-01 Optium Xceed Diabetes Monitoring System 10K128C XCB135 70089-01 Optium Xceed Diabetes Monitoring System 10K128C XCB317 70089-01 Optium Xceed Diabetes Monitoring System 10K128C XCB645 70089-01 Optium Xceed Diabetes Monitoring System 10K151C XCB527 70089-01 Optium Xceed Diabetes Monitoring System 10K151C XCB645 70089-01 Optium Xceed Diabetes Monitoring System 10K152C XCB537 70089-01 Optium Xceed Diabetes Monitoring System 10K155C XCB551 70089-01 Optium Xceed Diabetes Monitoring System 10K192C XCB774 70089-01 Optium Xceed Diabetes Monitoring System 10K198C XCB806 70089-01 Optium Xceed Diabetes Monitoring System 10K198C XCB809 70089-01 Optium Xceed Diabetes Monitoring System 11K152C XCB537 70089-01 Optium Xceed Diabetes Monitoring System 11K152C XCB551 70089-01 Optium Xceed Diabetes Monitoring System 11K176C XCB397 70089-01 Optium Xceed Diabetes Monitoring System 11K185C XCB449 70089-01 Optium Xceed Diabetes Monitoring System 11K185C XCB450 70089-01 Optium Xceed Diabetes Monitoring System 12K185C XCB449 70089-01 Optium Xceed Diabetes Monitoring System 12K187C XCB449 70089-01 Optium Xceed Diabetes Monitoring System 12K187C XCB450 70089-01 Optium Xceed Diabetes Monitoring System 13K186C XCB449 70089-01 Optium Xceed Diabetes Monitoring System 13K186C XCB450 70089-01 Optium Xceed Diabetes Monitoring System 13K187C XCB449 70089-01 Optium Xceed Diabetes Monitoring System 14K185C XCB449 70089-01 Optium Xceed Diabetes Monitoring System 14K185C XCB453 70089-01 Optium Xceed Diabetes Monitoring System 15K190C XCB780 70089-01 Optium Xceed Diabetes Monitoring System 16K190C XCB778 70089-01 Optium Xceed Diabetes Monitoring System 16K190C XCB780 70089-01 Optium Xceed Diabetes Monitoring System 17K190C XCB775 70089-01 Optium Xceed Diabetes Monitoring System 17K190C XCB778 70089-01 Optium Xceed Diabetes Monitoring System 17K191C XCB774 70089-01 Optium Xceed Diabetes Monitoring System 17K191C XCB781 70089-01 Optium Xceed Diabetes Monitoring System 18K190C XCB778 70095-35 Optium Xceed Diabetes Monitoring System 01K177C XCB733 70095-35 Optium Xceed Diabetes Monitoring System 01K198C XCB794 70095-35 Optium Xceed Diabetes Monitoring System 02K170C XCB373 70095-35 Optium Xceed Diabetes Monitoring System 02K170C XCB380 70095-35 Optium Xceed Diabetes Monitoring System 03K166C XCB374 70095-35 Optium Xceed Diabetes Monitoring System 03K176C XCB373 70095-35 Optium Xceed Diabetes Monitoring System 03K176C XCB374 70095-35 Optium Xceed Diabetes Monitoring System 03K177C XCB733 70095-35 Optium Xceed Diabetes Monitoring System 03K197C XCB793 70095-35 Optium Xceed Diabetes Monitoring System 03K197C XCB796 70095-35 Optium Xceed Diabetes Monitoring System 04K151C XCB372 70095-35 Optium Xceed Diabetes Monitoring System 04K151C XCB378 70095-35 Optium Xceed Diabetes Monitoring System 04K171C XCB411 70095-35 Optium Xceed Diabetes Monitoring System 05K170C XCB374 70095-35 Optium Xceed Diabetes Monitoring System 05K180C XCB412 70095-35 Optium Xceed Diabetes Monitoring System 05K180C XCB414 70095-35 Optium Xceed Diabetes Monitoring System 05K197C XCB796 70095-35 Optium Xceed Diabetes Monitoring System 06K194C XCB794 70095-35 Optium Xceed Diabetes Monitoring System 06K194C XCB796 70095-35 Optium Xceed Diabetes Monitoring System 07K141C XCB339 70095-35 Optium Xceed Diabetes Monitoring System 07K141C XCB372 70095-35 Optium Xceed Diabetes Monitoring System 07K194C XCB796 70095-35 Optium Xceed Diabetes Monitoring System 08K180C XCB374 70095-35 Optium Xceed Diabetes Monitoring System 10K176C XCB412 70095-35 Optium Xceed Diabetes Monitoring System 10K176C XCB733 70095-35 Optium Xceed Diabetes Monitoring System 10K197C XCB794 70095-35 Optium Xceed Diabetes Monitoring System 10K197C XCB796 70095-35 Optium Xceed Diabetes Monitoring System 11K170C XCB380 70095-35 Optium Xceed Diabetes Monitoring System 11K170C XCB411 70095-35 Optium Xceed Diabetes Monitoring System 11K177C XCB412 70095-35 Optium Xceed Diabetes Monitoring System 11K177C XCB733 70095-35 Optium Xceed Diabetes Monitoring System 11K179C XCB412 70095-35 Optium Xceed Diabetes Monitoring System 11K179C XCB414 70095-35 Optium Xceed Diabetes Monitoring System 12K150C XCB372 70095-35 Optium Xceed Diabetes Monitoring System 12K150C XCB378 70095-35 Optium Xceed Diabetes Monitoring System 12K171C XCB373 70095-35 Optium Xceed Diabetes Monitoring System 12K176C XCB414 70095-35 Optium Xceed Diabetes Monitoring System 12K177C XCB414 70095-35 Optium Xceed Diabetes Monitoring System 12K194C XCB793 70095-35 Optium Xceed Diabetes Monitoring System 12K194C XCB796 70095-35 Optium Xceed Diabetes Monitoring System 12K197C XCB794 70095-35 Optium Xceed Diabetes Monitoring System 12K197C XCB795 70095-35 Optium Xceed Diabetes Monitoring System 13K150C XCB338 70095-35 Optium Xceed Diabetes Monitoring System 13K150C XCB339 70095-35 Optium Xceed Diabetes Monitoring System 13K150C XCB372 70095-35 Optium Xceed Diabetes Monitoring System 14K150C XCB372 70095-35 Optium Xceed Diabetes Monitoring System 14K170C XCB411 70095-35 Optium Xceed Diabetes Monitoring System 14K177C XCB374 70095-35 Optium Xceed Diabetes Monitoring System 14K177C XCB414 70095-35 Optium Xceed Diabetes Monitoring System 14K177C XCB415 70095-35 Optium Xceed Diabetes Monitoring System 15K150C XCB372 70095-35 Optium Xceed Diabetes Monitoring System 15K197C XCB794 70095-35 Optium Xceed Diabetes Monitoring System 18K169C XCB373 70095-35 Optium Xceed Diabetes Monitoring System 18K169C XCB380 70095-35 Optium Xceed Diabetes Monitoring System 18K176C XCB733 70095-35 Optium Xceed Diabetes Monitoring System 20K165C XCB374 70096-01 Optium Xceed Diabetes Monitoring System 03K156C XCB378 70096-01 Optium Xceed Diabetes Monitoring System 05K155C XCB378 70096-01 Optium Xceed Diabetes Monitoring System 08K155C XCB378 70096-01 Optium Xceed Diabetes Monitoring System 08K190C XCB412 70096-01 Optium Xceed Diabetes Monitoring System 08K190C XCB415 70096-01 Optium Xceed Diabetes Monitoring System 08K190C XCB733 70096-01 Optium Xceed Diabetes Monitoring System 08K190C XCB734 70096-01 Optium Xceed Diabetes Monitoring System 14K186C XCB734 70096-01 Optium Xceed Diabetes Monitoring System 18K187C XCB734 70098-01 Optium Xceed Diabetes Monitoring System 01K102D XCB341 70098-01 Optium Xceed Diabetes Monitoring System 01K102D XCB344 70098-01 Optium Xceed Diabetes Monitoring System 01K102D XCB362 70098-01 Optium Xceed Diabetes Monitoring System 01K103D XCB362 70098-01 Optium Xceed Diabetes Monitoring System 01K103D XCB410 70098-01 Optium Xceed Diabetes Monitoring System 01K103D XCB462 70098-01 Optium Xceed Diabetes Monitoring System 01K103D XCB463 70098-01 Optium Xceed Diabetes Monitoring System 01K106D XCB458 70098-01 Optium Xceed Diabetes Monitoring System 01K106D XCB459 70098-01 Optium Xceed Diabetes Monitoring System 01K106D XCB462 70098-01 Optium Xceed Diabetes Monitoring System 01K124D XCB481 70098-01 Optium Xceed Diabetes Monitoring System 01K124D XCB488 70098-01 Optium Xceed Diabetes Monitoring System 01K127D XCB481 70098-01 Optium Xceed Diabetes Monitoring System 01K127D XCB488 70098-01 Optium Xceed Diabetes Monitoring System 01K180D XCB575 70098-01 Optium Xceed Diabetes Monitoring System 01K180D XCB576 70098-01 Optium Xceed Diabetes Monitoring System 01K181D XCB575 70098-01 Optium Xceed Diabetes Monitoring System 01K181D XCB576 70098-01 Optium Xceed Diabetes Monitoring System 01K202D XCB575 70098-01 Optium Xceed Diabetes Monitoring System 01K202D XCB577 70098-01 Optium Xceed Diabetes Monitoring System 01K202D XCB592 70098-01 Optium Xceed Diabetes Monitoring System 01K202D XCB593 70098-01 Optium Xceed Diabetes Monitoring System 02K101D XCB341 70098-01 Optium Xceed Diabetes Monitoring System 02K101D XCB344 70098-01 Optium Xceed Diabetes Monitoring System 02K101D XCB362 70098-01 Optium Xceed Diabetes Monitoring System 02K102D XCB362 70098-01 Optium Xceed Diabetes Monitoring System 02K102D XCB410 70098-01 Optium Xceed Diabetes Monitoring System 02K102D XCB462 70098-01 Optium Xceed Diabetes Monitoring System 02K102D XCB463 70098-01 Optium Xceed Diabetes Monitoring System 02K107D XCB458 70098-01 Optium Xceed Diabetes Monitoring System 02K107D XCB461 70098-01 Optium Xceed Diabetes Monitoring System 02K127D XCB483 70098-01 Optium Xceed Diabetes Monitoring System 02K127D XCB484 70098-01 Optium Xceed Diabetes Monitoring System 02K127D XCB488 70098-01 Optium Xceed Diabetes Monitoring System 02K201D XCB573 70098-01 Optium Xceed Diabetes Monitoring System 02K201D XCB582 70098-01 Optium Xceed Diabetes Monitoring System 02K201D XCB586 70098-01 Optium Xceed Diabetes Monitoring System 02K201D XCB593 70098-01 Optium Xceed Diabetes Monitoring System 03K102D XCB458 70098-01 Optium Xceed Diabetes Monitoring System 03K102D XCB459 70098-01 Optium Xceed Diabetes Monitoring System 03K102D XCB462 70098-01 Optium Xceed Diabetes Monitoring System 03K103D XCB458 70098-01 Optium Xceed Diabetes Monitoring System 03K103D XCB459 70098-01 Optium Xceed Diabetes Monitoring System 03K103D XCB462 70098-01 Optium Xceed Diabetes Monitoring System 03K123D XCB483 70098-01 Optium Xceed Diabetes Monitoring System 03K123D XCB484 70098-01 Optium Xceed Diabetes Monitoring System 03K123D XCB488 70098-01 Optium Xceed Diabetes Monitoring System 03K124D XCB483 70098-01 Optium Xceed Diabetes Monitoring System 03K124D XCB484 70098-01 Optium Xceed Diabetes Monitoring System 03K124D XCB488 70098-01 Optium Xceed Diabetes Monitoring System 03K178D XCB558 70098-01 Optium Xceed Diabetes Monitoring System 03K178D XCB576 70098-01 Optium Xceed Diabetes Monitoring System 03K178D XCB584 70098-01 Optium Xceed Diabetes Monitoring System 03K180D XCB480 70098-01 Optium Xceed Diabetes Monitoring System 03K180D XCB489 70098-01 Optium Xceed Diabetes Monitoring System 03K180D XCB496 70098-01 Optium Xceed Diabetes Monitoring System 03K201D XCB575 70098-01 Optium Xceed Diabetes Monitoring System 03K201D XCB577 70098-01 Optium Xceed Diabetes Monitoring System 03K201D XCB592 70098-01 Optium Xceed Diabetes Monitoring System 03K201D XCB593 70098-01 Optium Xceed Diabetes Monitoring System 03K204D XCB575 70098-01 Optium Xceed Diabetes Monitoring System 03K204D XCB577 70098-01 Optium Xceed Diabetes Monitoring System 03K204D XCB592 70098-01 Optium Xceed Diabetes Monitoring System 03K204D XCB593 70098-01 Optium Xceed Diabetes Monitoring System 04K072D XCB353 70098-01 Optium Xceed Diabetes Monitoring System 04K072D XCB410 70098-01 Optium Xceed Diabetes Monitoring System 04K106D XCB458 70098-01 Optium Xceed Diabetes Monitoring System 04K106D XCB461 70098-01 Optium Xceed Diabetes Monitoring System 04K178D XCB480 70098-01 Optium Xceed Diabetes Monitoring System 04K178D XCB497 70098-01 Optium Xceed Diabetes Monitoring System 04K178D XCB558 70098-01 Optium Xceed Diabetes Monitoring System 04K179D XCB558 70098-01 Optium Xceed Diabetes Monitoring System 04K179D XCB576 70098-01 Optium Xceed Diabetes Monitoring System 04K179D XCB584 70098-01 Optium Xceed Diabetes Monitoring System 05K124D XCB460 70098-01 Optium Xceed Diabetes Monitoring System 05K124D XCB461 70098-01 Optium Xceed Diabetes Monitoring System 05K124D XCB481 70098-01 Optium Xceed Diabetes Monitoring System 05K177D XCB558 70098-01 Optium Xceed Diabetes Monitoring System 05K177D XCB576 70098-01 Optium Xceed Diabetes Monitoring System 05K177D XCB584 70098-01 Optium Xceed Diabetes Monitoring System 05K204D XCB573 70098-01 Optium Xceed Diabetes Monitoring System 05K204D XCB582 70098-01 Optium Xceed Diabetes Monitoring System 05K204D XCB586 70098-01 Optium Xceed Diabetes Monitoring System 05K204D XCB593 70098-01 Optium Xceed Diabetes Monitoring System 06K071D XCB353 70098-01 Optium Xceed Diabetes Monitoring System 06K071D XCB410 70098-01 Optium Xceed Diabetes Monitoring System 06K101D XCB362 70098-01 Optium Xceed Diabetes Monitoring System 06K101D XCB410 70098-01 Optium Xceed Diabetes Monitoring System 06K101D XCB462 70098-01 Optium Xceed Diabetes Monitoring System 06K101D XCB463 70098-01 Optium Xceed Diabetes Monitoring System 06K179D XCB480 70098-01 Optium Xceed Diabetes Monitoring System 06K179D XCB489 70098-01 Optium Xceed Diabetes Monitoring System 06K179D XCB496 70098-01 Optium Xceed Diabetes Monitoring System 06K200D XCB575 70098-01 Optium Xceed Diabetes Monitoring System 06K200D XCB577 70098-01 Optium Xceed Diabetes Monitoring System 06K200D XCB592 70098-01 Optium Xceed Diabetes Monitoring System 06K200D XCB593 70098-01 Optium Xceed Diabetes Monitoring System 06K205D XCB573 70098-01 Optium Xceed Diabetes Monitoring System 06K205D XCB582 70098-01 Optium Xceed Diabetes Monitoring System 06K205D XCB586 70098-01 Optium Xceed Diabetes Monitoring System 06K205D XCB593 70098-01 Optium Xceed Diabetes Monitoring System 07K123D XCB460 70098-01 Optium Xceed Diabetes Monitoring System 07K123D XCB461 70098-01 Optium Xceed Diabetes Monitoring System 07K123D XCB481 70098-01 Optium Xceed Diabetes Monitoring System 07K178D XCB480 70098-01 Optium Xceed Diabetes Monitoring System 07K178D XCB489 70098-01 Optium Xceed Diabetes Monitoring System 07K178D XCB496 70098-01 Optium Xceed Diabetes Monitoring System 07K199D XCB575 70098-01 Optium Xceed Diabetes Monitoring System 07K199D XCB577 70098-01 Optium Xceed Diabetes Monitoring System 07K199D XCB592 70098-01 Optium Xceed Diabetes Monitoring System 07K199D XCB593 70098-01 Optium Xceed Diabetes Monitoring System 08K199D XCB573 70098-01 Optium Xceed Diabetes Monitoring System 08K199D XCB582 70098-01 Optium Xceed Diabetes Monitoring System 08K199D XCB586 70098-01 Optium Xceed Diabetes Monitoring System 08K199D XCB593 70098-01 Optium Xceed Diabetes Monitoring System 08K200D XCB573 70098-01 Optium Xceed Diabetes Monitoring System 08K200D XCB582 70098-01 Optium Xceed Diabetes Monitoring System 08K200D XCB586 70098-01 Optium Xceed Diabetes Monitoring System 08K200D XCB593 70098-01 Optium Xceed Diabetes Monitoring System 09K123D XCB481 70098-01 Optium Xceed Diabetes Monitoring System 09K123D XCB488 70098-01 Optium Xceed Diabetes Monitoring System 09K179D XCB480 70098-01 Optium Xceed Diabetes Monitoring System 09K179D XCB497 70098-01 Optium Xceed Diabetes Monitoring System 09K179D XCB558 70098-01 Optium Xceed Diabetes Monitoring System 09K198D XCB575 70098-01 Optium Xceed Diabetes Monitoring System 09K198D XCB577 70098-01 Optium Xceed Diabetes Monitoring System 09K198D XCB592 70098-01 Optium Xceed Diabetes Monitoring System 09K198D XCB593 70098-01 Optium Xceed Diabetes Monitoring System 10K100D XCB341 70098-01 Optium Xceed Diabetes Monitoring System 10K100D XCB344 70098-01 Optium Xceed Diabetes Monitoring System 10K100D XCB362 70098-01 Optium Xceed Diabetes Monitoring System 10K177D XCB480 70098-01 Optium Xceed Diabetes Monitoring System 10K177D XCB497 70098-01 Optium Xceed Diabetes Monitoring System 10K177D XCB558 70098-01 Optium Xceed Diabetes Monitoring System 11K179D XCB575 70098-01 Optium Xceed Diabetes Monitoring System 11K179D XCB576 70099-01 Optium Xceed Diabetes Monitoring System 02K163D XCB496 70099-01 Optium Xceed Diabetes Monitoring System 02K204D XCB582 70099-01 Optium Xceed Diabetes Monitoring System 02K204D XCB945 70099-01 Optium Xceed Diabetes Monitoring System 08K158D XCB437 70099-01 Optium Xceed Diabetes Monitoring System 08K158D XCB481 70099-01 Optium Xceed Diabetes Monitoring System 08K158D XCB484 70099-01 Optium Xceed Diabetes Monitoring System 08K158D XCB488 70099-01 Optium Xceed Diabetes Monitoring System 08K158D XCB496 70099-01 Optium Xceed Diabetes Monitoring System 08K211D XCB573 70099-01 Optium Xceed Diabetes Monitoring System 08K211D XCB976 70100-01 Optium Xceed Diabetes Monitoring System 01K135E XCB253 70100-01 Optium Xceed Diabetes Monitoring System 01K135E XCB363 70100-01 Optium Xceed Diabetes Monitoring System 01K137E XCB363 70100-01 Optium Xceed Diabetes Monitoring System 01K137E XCB366 70100-01 Optium Xceed Diabetes Monitoring System 01K138E XCB364 70100-01 Optium Xceed Diabetes Monitoring System 01K138E XCB366 70100-01 Optium Xceed Diabetes Monitoring System 01K158E XCB364 70100-01 Optium Xceed Diabetes Monitoring System 01K159E XCB387 70100-01 Optium Xceed Diabetes Monitoring System 01K162E XCB387 70100-01 Optium Xceed Diabetes Monitoring System 01K197E XCB387 70100-01 Optium Xceed Diabetes Monitoring System 01K197E XCB389 70100-01 Optium Xceed Diabetes Monitoring System 01K198E XCB387 70100-01 Optium Xceed Diabetes Monitoring System 01K198E XCB389 70100-01 Optium Xceed Diabetes Monitoring System 01K207E XCB382 70100-01 Optium Xceed Diabetes Monitoring System 01K207E XCB383 70100-01 Optium Xceed Diabetes Monitoring System 02K136E XCB253 70100-01 Optium Xceed Diabetes Monitoring System 02K136E XCB363 70100-01 Optium Xceed Diabetes Monitoring System 02K137E XCB364 70100-01 Optium Xceed Diabetes Monitoring System 02K137E XCB366 70100-01 Optium Xceed Diabetes Monitoring System 02K158E XCB364 70100-01 Optium Xceed Diabetes Monitoring System 02K158E XCB387 70100-01 Optium Xceed Diabetes Monitoring System 02K194E XCB387 70100-01 Optium Xceed Diabetes Monitoring System 02K194E XCB389 70100-01 Optium Xceed Diabetes Monitoring System 02K199E XCB389 70100-01 Optium Xceed Diabetes Monitoring System 02K206E XCB382 70100-01 Optium Xceed Diabetes Monitoring System 02K206E XCB389 70100-01 Optium Xceed Diabetes Monitoring System 02K208E XCB382 70100-01 Optium Xceed Diabetes Monitoring System 02K208E XCB383 70100-01 Optium Xceed Diabetes Monitoring System 02K208E XCB384 70100-01 Optium Xceed Diabetes Monitoring System 03K157E XCB364 70100-01 Optium Xceed Diabetes Monitoring System 03K194E XCB387 70100-01 Optium Xceed Diabetes Monitoring System 03K194E XCB389 70100-01 Optium Xceed Diabetes Monitoring System 03K198E XCB389 70100-01 Optium Xceed Diabetes Monitoring System 03K205E XCB382 70100-01 Optium Xceed Diabetes Monitoring System 03K205E XCB389 70100-01 Optium Xceed Diabetes Monitoring System 03K207E XCB382 70100-01 Optium Xceed Diabetes Monitoring System 03K207E XCB383 70100-01 Optium Xceed Diabetes Monitoring System 03K207E XCB384 70100-01 Optium Xceed Diabetes Monitoring System 04K136E XCB363 70100-01 Optium Xceed Diabetes Monitoring System 04K136E XCB366 70100-01 Optium Xceed Diabetes Monitoring System 04K206E XCB382 70100-01 Optium Xceed Diabetes Monitoring System 04K206E XCB383 70100-01 Optium Xceed Diabetes Monitoring System 05K134E XCB253 70100-01 Optium Xceed Diabetes Monitoring System 05K134E XCB363 70100-01 Optium Xceed Diabetes Monitoring System 08K067D XCB410 70101-01 Optium Xceed Diabetes Monitoring System 01K106E XCB370 70101-01 Optium Xceed Diabetes Monitoring System 01K106E XCB404 70101-01 Optium Xceed Diabetes Monitoring System 01K122E XCB404 70101-01 Optium Xceed Diabetes Monitoring System 01K123E XCB404 70101-01 Optium Xceed Diabetes Monitoring System 01K136E XCA908 70101-01 Optium Xceed Diabetes Monitoring System 01K136E XCB547 70101-01 Optium Xceed Diabetes Monitoring System 01K166E XCB547 70101-01 Optium Xceed Diabetes Monitoring System 01K166E XCB818 70101-01 Optium Xceed Diabetes Monitoring System 01K169E XCB818 70101-01 Optium Xceed Diabetes Monitoring System 01K170E XCB818 70101-01 Optium Xceed Diabetes Monitoring System 01K199E XCB819 70101-01 Optium Xceed Diabetes Monitoring System 01K199E XCB822 70101-01 Optium Xceed Diabetes Monitoring System 02K093E XCB370 70101-01 Optium Xceed Diabetes Monitoring System 02K197E XCB819 70101-01 Optium Xceed Diabetes Monitoring System 02K198E XCB819 70101-01 Optium Xceed Diabetes Monitoring System 03K082E XCB170 70101-01 Optium Xceed Diabetes Monitoring System 03K082E XCB370 70101-01 Optium Xceed Diabetes Monitoring System 03K103E XCB370 70101-01 Optium Xceed Diabetes Monitoring System 03K103E XCB404 70101-01 Optium Xceed Diabetes Monitoring System 03K135E XCA908 70101-01 Optium Xceed Diabetes Monitoring System 03K135E XCB547 70101-01 Optium Xceed Diabetes Monitoring System 03K136E XCB547 70101-01 Optium Xceed Diabetes Monitoring System 03K197E XCB819 70101-01 Optium Xceed Diabetes Monitoring System 04K198E XCB819 70101-01 Optium Xceed Diabetes Monitoring System 04K198E XCB822 70102-01 Optium Xceed Diabetes Monitoring System 01K095E XCB370 70102-01 Optium Xceed Diabetes Monitoring System 02K124E XCB404 70102-01 Optium Xceed Diabetes Monitoring System 02K124E XCB476 70102-01 Optium Xceed Diabetes Monitoring System 02K124E XCB547 70102-01 Optium Xceed Diabetes Monitoring System 02K170E XCB818 70102-01 Optium Xceed Diabetes Monitoring System 02K170E XCB819 70102-01 Optium Xceed Diabetes Monitoring System 03K094E XCB370 70105-35 Precision Xceed Diabetes Management System 02K171C XCB419 70105-35 Precision Xceed Diabetes Management System 02K171C XCB456 70105-35 Precision Xceed Diabetes Management System 04K170C XCB456 70105-35 Precision Xceed Diabetes Management System 07K130C XCB444 70105-35 Precision Xceed Diabetes Management System 12K135C XCB444 70105-35 Precision Xceed Diabetes Management System 16K170C XCB419 70105-35 Precision Xceed Diabetes Management System 16K170C XCB456 70106-01 Precision Xceed Diabetes Management System 11K094C XCB444 70107-35 Precision Xceed Diabetes Management System 03K172C XCB456 70107-35 Precision Xceed Diabetes Management System 08K121C XCB444 70107-35 Precision Xceed Diabetes Management System 08K122C XCB444 70107-35 Precision Xceed Diabetes Management System 10K122C XCB444 70107-35 Precision Xceed Diabetes Management System 16K191C XCB871 70107-35 Precision Xceed Diabetes Management System 17K171C XCB456 70107-35 Precision Xceed Diabetes Management System 18K121C XCB444 70109-01 Optium Xceed Diabetes Monitoring System 01K072E XCB334 70109-01 Optium Xceed Diabetes Monitoring System 01K072E XCB365 70109-01 Optium Xceed Diabetes Monitoring System 01K073E XCB369 70109-01 Optium Xceed Diabetes Monitoring System 01K073E XCB467 70109-01 Optium Xceed Diabetes Monitoring System 01K074E XCB467 70109-01 Optium Xceed Diabetes Monitoring System 01K074E XCB468 70109-01 Optium Xceed Diabetes Monitoring System 01K074E XCB499 70109-01 Optium Xceed Diabetes Monitoring System 01K075E XCB468 70109-01 Optium Xceed Diabetes Monitoring System 01K075E XCB499 70109-01 Optium Xceed Diabetes Monitoring System 01K078E XCB499 70109-01 Optium Xceed Diabetes Monitoring System 01K079E XCB499 70109-01 Optium Xceed Diabetes Monitoring System 01K085E XCB499 70109-01 Optium Xceed Diabetes Monitoring System 01K085E XCB500 70109-01 Optium Xceed Diabetes Monitoring System 01K088E XCB500 70109-01 Optium Xceed Diabetes Monitoring System 01K093E XCB500 70109-01 Optium Xceed Diabetes Monitoring System 01K093E XCB517 70109-01 Optium Xceed Diabetes Monitoring System 01K094E XCB517 70109-01 Optium Xceed Diabetes Monitoring System 01K101E XCB517 70109-01 Optium Xceed Diabetes Monitoring System 01K101E XCB549 70109-01 Optium Xceed Diabetes Monitoring System 01K101E XCB550 70109-01 Optium Xceed Diabetes Monitoring System 01K124E XCB613 70109-01 Optium Xceed Diabetes Monitoring System 01K124E XCB614 70109-01 Optium Xceed Diabetes Monitoring System 01K127E XCB613 70109-01 Optium Xceed Diabetes Monitoring System 01K127E XCB614 70109-01 Optium Xceed Diabetes Monitoring System 01K127E XCB624 70109-01 Optium Xceed Diabetes Monitoring System 02K068E XCB409 70109-01 Optium Xceed Diabetes Monitoring System 02K073E XCB365 70109-01 Optium Xceed Diabetes Monitoring System 02K073E XCB369 70109-01 Optium Xceed Diabetes Monitoring System 02K074E XCB467 70109-01 Optium Xceed Diabetes Monitoring System 02K074E XCB499 70109-01 Optium Xceed Diabetes Monitoring System 02K075E XCB468 70109-01 Optium Xceed Diabetes Monitoring System 02K075E XCB499 70109-01 Optium Xceed Diabetes Monitoring System 02K078E XCB499 70109-01 Optium Xceed Diabetes Monitoring System 02K087E XCB500 70109-01 Optium Xceed Diabetes Monitoring System 02K100E XCB517 70109-01 Optium Xceed Diabetes Monitoring System 02K100E XCB550 70109-01 Optium Xceed Diabetes Monitoring System 02K101E XCB517 70109-01 Optium Xceed Diabetes Monitoring System 02K101E XCB550 70109-01 Optium Xceed Diabetes Monitoring System 02K103E XCB549 70109-01 Optium Xceed Diabetes Monitoring System 02K103E XCB550 70109-01 Optium Xceed Diabetes Monitoring System 02K103E XCB552 70109-01 Optium Xceed Diabetes Monitoring System 02K106E XCB550 70109-01 Optium Xceed Diabetes Monitoring System 02K106E XCB552 70109-01 Optium Xceed Diabetes Monitoring System 02K107E XCB550 70109-01 Optium Xceed Diabetes Monitoring System 02K107E XCB552 70109-01 Optium Xceed Diabetes Monitoring System 02K122E XCB552 70109-01 Optium Xceed Diabetes Monitoring System 02K122E XCB614 70109-01 Optium Xceed Diabetes Monitoring System 02K122E XCB615 70109-01 Optium Xceed Diabetes Monitoring System 02K123E XCB614 70109-01 Optium Xceed Diabetes Monitoring System 02K123E XCB615 70109-01 Optium Xceed Diabetes Monitoring System 02K127E XCB613 70109-01 Optium Xceed Diabetes Monitoring System 02K127E XCB624 70109-01 Optium Xceed Diabetes Monitoring System 02K128E XCB613 70109-01 Optium Xceed Diabetes Monitoring System 02K128E XCB624 70109-01 Optium Xceed Diabetes Monitoring System 02K128E XCB628 70109-01 Optium Xceed Diabetes Monitoring System 02K128E XCB642 70109-01 Optium Xceed Diabetes Monitoring System 02K131E XCB642 70109-01 Optium Xceed Diabetes Monitoring System 03K067E XCB409 70109-01 Optium Xceed Diabetes Monitoring System 03K071E XCB334 70109-01 Optium Xceed Diabetes Monitoring System 03K071E XCB365 70109-01 Optium Xceed Diabetes Monitoring System 03K072E XCB369 70109-01 Optium Xceed Diabetes Monitoring System 03K072E XCB467 70109-01 Optium Xceed Diabetes Monitoring System 03K073E XCB467 70109-01 Optium Xceed Diabetes Monitoring System 03K073E XCB468 70109-01 Optium Xceed Diabetes Monitoring System 03K073E XCB499 70109-01 Optium Xceed Diabetes Monitoring System 03K074E XCB468 70109-01 Optium Xceed Diabetes Monitoring System 03K074E XCB499 70109-01 Optium Xceed Diabetes Monitoring System 03K075E XCB499 70109-01 Optium Xceed Diabetes Monitoring System 03K100E XCB517 70109-01 Optium Xceed Diabetes Monitoring System 03K100E XCB549 70109-01 Optium Xceed Diabetes Monitoring System 03K100E XCB550 70109-01 Optium Xceed Diabetes Monitoring System 03K101E XCB549 70109-01 Optium Xceed Diabetes Monitoring System 03K102E XCB549 70109-01 Optium Xceed Diabetes Monitoring System 03K102E XCB550 70109-01 Optium Xceed Diabetes Monitoring System 03K102E XCB552 70109-01 Optium Xceed Diabetes Monitoring System 03K107E XCB550 70109-01 Optium Xceed Diabetes Monitoring System 03K107E XCB552 70109-01 Optium Xceed Diabetes Monitoring System 03K123E XCB614 70109-01 Optium Xceed Diabetes Monitoring System 04K068D XCB475 70109-01 Optium Xceed Diabetes Monitoring System 04K072E XCB365 70109-01 Optium Xceed Diabetes Monitoring System 04K072E XCB369 70109-01 Optium Xceed Diabetes Monitoring System 04K073E XCB467 70109-01 Optium Xceed Diabetes Monitoring System 04K073E XCB499 70109-01 Optium Xceed Diabetes Monitoring System 04K074E XCB468 70109-01 Optium Xceed Diabetes Monitoring System 04K074E XCB499 70109-01 Optium Xceed Diabetes Monitoring System 04K075E XCB499 70109-01 Optium Xceed Diabetes Monitoring System 04K103E XCB550 70109-01 Optium Xceed Diabetes Monitoring System 04K103E XCB552 70109-01 Optium Xceed Diabetes Monitoring System 04K106E XCB550 70109-01 Optium Xceed Diabetes Monitoring System 04K106E XCB552 70109-01 Optium Xceed Diabetes Monitoring System 05K082E XCB499 70109-01 Optium Xceed Diabetes Monitoring System 05K082E XCB500 70109-01 Optium Xceed Diabetes Monitoring System 07K087D XCB174 70109-01 Optium Xceed Diabetes Monitoring System 07K087D XCB475 70109-01 Optium Xceed Diabetes Monitoring System 11K086D XCB174 70109-01 Optium Xceed Diabetes Monitoring System 11K086D XCB475 70134-35 Precision Xceed Diabetes Management System 02K172C XCB556 70134-35 Precision Xceed Diabetes Management System 02K172C XCB580 70134-35 Precision Xceed Diabetes Management System 04K120C XCB466 70134-35 Precision Xceed Diabetes Management System 04K120C XCB471 70134-35 Precision Xceed Diabetes Management System 04K166C XCB472 70134-35 Precision Xceed Diabetes Management System 04K166C XCB525 70134-35 Precision Xceed Diabetes Management System 04K166C XCB555 70134-35 Precision Xceed Diabetes Management System 04K193C XCB490 70134-35 Precision Xceed Diabetes Management System 04K193C XCB585 70134-35 Precision Xceed Diabetes Management System 04K193C XCB629 70134-35 Precision Xceed Diabetes Management System 05K171C XCB580 70134-35 Precision Xceed Diabetes Management System 05K193C XCB490 70134-35 Precision Xceed Diabetes Management System 05K193C XCB752 70134-35 Precision Xceed Diabetes Management System 06K071C XCB464 70134-35 Precision Xceed Diabetes Management System 06K071C XCB471 70134-35 Precision Xceed Diabetes Management System 07K121C XCB466 70134-35 Precision Xceed Diabetes Management System 08K169C XCB472 70134-35 Precision Xceed Diabetes Management System 08K172C XCB555 70134-35 Precision Xceed Diabetes Management System 08K172C XCB580 70134-35 Precision Xceed Diabetes Management System 12K121C XCB466 70134-35 Precision Xceed Diabetes Management System 12K121C XCB471 70134-35 Precision Xceed Diabetes Management System 12K127C XCB512 70134-35 Precision Xceed Diabetes Management System 13K192C XCB490 70134-35 Precision Xceed Diabetes Management System 13K192C XCB752 70134-35 Precision Xceed Diabetes Management System 15K127C XCB219 70134-35 Precision Xceed Diabetes Management System 15K127C XCB466 70134-35 Precision Xceed Diabetes Management System 15K127C XCB512 70134-35 Precision Xceed Diabetes Management System 15K171C XCB556 70134-35 Precision Xceed Diabetes Management System 15K171C XCB580 70134-35 Precision Xceed Diabetes Management System 16K127C XCB512 70134-35 Precision Xceed Diabetes Management System 17K165C XCB472 70134-35 Precision Xceed Diabetes Management System 17K165C XCB525 70134-35 Precision Xceed Diabetes Management System 17K165C XCB555 70134-35 Precision Xceed Diabetes Management System 18K170C XCB580 70135-01 Precision Xceed Diabetes Management System 01K072C XCB465 70135-01 Precision Xceed Diabetes Management System 01K134C XCB511 70135-01 Precision Xceed Diabetes Management System 01K134C XCB512 70135-01 Precision Xceed Diabetes Management System 01K150C XCB509 70135-01 Precision Xceed Diabetes Management System 01K150C XCB525 70135-01 Precision Xceed Diabetes Management System 01K150C XCB545 70135-01 Precision Xceed Diabetes Management System 02K150C XCB579 70135-01 Precision Xceed Diabetes Management System 03K060C XCB263 70135-01 Precision Xceed Diabetes Management System 03K060C XCB432 70135-01 Precision Xceed Diabetes Management System 04K134C XCB512 70135-01 Precision Xceed Diabetes Management System 04K149C XCB525 70135-01 Precision Xceed Diabetes Management System 04K150C XCB555 70135-01 Precision Xceed Diabetes Management System 04K150C XCB579 70135-01 Precision Xceed Diabetes Management System 05K149C XCB525 70135-01 Precision Xceed Diabetes Management System 06K149C XCB491 70135-01 Precision Xceed Diabetes Management System 06K149C XCB509 70135-01 Precision Xceed Diabetes Management System 07K149C XCB509 70135-01 Precision Xceed Diabetes Management System 07K149C XCB525 70135-01 Precision Xceed Diabetes Management System 07K149C XCB545 70135-01 Precision Xceed Diabetes Management System 08K057C XCB219 70135-01 Precision Xceed Diabetes Management System 08K057C XCB264 70135-01 Precision Xceed Diabetes Management System 08K057C XCB431 70135-01 Precision Xceed Diabetes Management System 08K057C XCB432 70135-01 Precision Xceed Diabetes Management System 08K059C XCB263 70135-01 Precision Xceed Diabetes Management System 08K059C XCB432 70135-01 Precision Xceed Diabetes Management System 08K150C XCB525 70135-01 Precision Xceed Diabetes Management System 08K150C XCB579 70135-01 Precision Xceed Diabetes Management System 09K071C XCB465 70135-01 Precision Xceed Diabetes Management System 10K138C XCB511 70135-01 Precision Xceed Diabetes Management System 10K149C XCB525 70135-01 Precision Xceed Diabetes Management System 10K149C XCB579 70135-01 Precision Xceed Diabetes Management System 10K150C XCB579 70135-01 Precision Xceed Diabetes Management System 11K071C XCB465 70135-01 Precision Xceed Diabetes Management System 11K149C XCB579 70135-01 Precision Xceed Diabetes Management System 12K149C XCB555 70135-01 Precision Xceed Diabetes Management System 12K149C XCB579 70135-01 Precision Xceed Diabetes Management System 13K131C XCB511 70135-01 Precision Xceed Diabetes Management System 13K131C XCB512 70135-01 Precision Xceed Diabetes Management System 13K137C XCB511 70135-01 Precision Xceed Diabetes Management System 14K131C XCB512 70135-01 Precision Xceed Diabetes Management System 14K134C XCB511 70135-01 Precision Xceed Diabetes Management System 17K172C XCB556 70135-01 Precision Xceed Diabetes Management System 17K172C XCB574 70515-01 Precision Xtra Advanced Diabetes Management System 01K057D XCB381 70515-01 Precision Xtra Advanced Diabetes Management System 01K057D XCB391 70515-01 Precision Xtra Advanced Diabetes Management System 01K057D XCB394 70515-01 Precision Xtra Advanced Diabetes Management System 01K057D XCB395 70515-01 Precision Xtra Advanced Diabetes Management System 01K073D XCB416 70515-01 Precision Xtra Advanced Diabetes Management System 01K073D XCB418 70515-01 Precision Xtra Advanced Diabetes Management System 01K073D XCB492 70515-01 Precision Xtra Advanced Diabetes Management System 01K088D XCB536 70515-01 Precision Xtra Advanced Diabetes Management System 01K088D XCB538 70515-01 Precision Xtra Advanced Diabetes Management System 01K128D XCB670 70515-01 Precision Xtra Advanced Diabetes Management System 01K128D XCB673 70515-01 Precision Xtra Advanced Diabetes Management System 01K128D XCB685 70515-01 Precision Xtra Advanced Diabetes Management System 01K129D XCB672 70515-01 Precision Xtra Advanced Diabetes Management System 01K129D XCB674 70515-01 Precision Xtra Advanced Diabetes Management System 01K129D XCB684 70515-01 Precision Xtra Advanced Diabetes Management System 01K155D XCB745 70515-01 Precision Xtra Advanced Diabetes Management System 01K155D XCB746 70515-01 Precision Xtra Advanced Diabetes Management System 01K155D XCB748 70515-01 Precision Xtra Advanced Diabetes Management System 01K155D XCB751 70515-01 Precision Xtra Advanced Diabetes Management System 01K186D XCB830 70515-01 Precision Xtra Advanced Diabetes Management System 01K186D XCB831 70515-01 Precision Xtra Advanced Diabetes Management System 01K201D XCB798 70515-01 Precision Xtra Advanced Diabetes Management System 01K201D XCB864 70515-01 Precision Xtra Advanced Diabetes Management System 01K201D XCB865 70515-01 Precision Xtra Advanced Diabetes Management System 01K201D XCB866 70515-01 Precision Xtra Advanced Diabetes Management System 02K054D XCB381 70515-01 Precision Xtra Advanced Diabetes Management System 02K054D XCB391 70515-01 Precision Xtra Advanced Diabetes Management System 02K054D XCB394 70515-01 Precision Xtra Advanced Diabetes Management System 02K054D XCB395 70515-01 Precision Xtra Advanced Diabetes Management System 02K109D XCB644 70515-01 Precision Xtra Advanced Diabetes Management System 02K128D XCB672 70515-01 Precision Xtra Advanced Diabetes Management System 02K128D XCB674 70515-01 Precision Xtra Advanced Diabetes Management System 02K128D XCB684 70515-01 Precision Xtra Advanced Diabetes Management System 02K129D XCB684 70515-01 Precision Xtra Advanced Diabetes Management System 02K129D XCB708 70515-01 Precision Xtra Advanced Diabetes Management System 02K129D XCB713 70515-01 Precision Xtra Advanced Diabetes Management System 02K152D XCB745 70515-01 Precision Xtra Advanced Diabetes Management System 02K152D XCB746 70515-01 Precision Xtra Advanced Diabetes Management System 02K152D XCB748 70515-01 Precision Xtra Advanced Diabetes Management System 02K152D XCB751 70515-01 Precision Xtra Advanced Diabetes Management System 02K202D XCB798 70515-01 Precision Xtra Advanced Diabetes Management System 02K202D XCB864 70515-01 Precision Xtra Advanced Diabetes Management System 02K202D XCB865 70515-01 Precision Xtra Advanced Diabetes Management System 02K202D XCB866 70515-01 Precision Xtra Advanced Diabetes Management System 03K068D XCB416 70515-01 Precision Xtra Advanced Diabetes Management System 03K068D XCB417 70515-01 Precision Xtra Advanced Diabetes Management System 03K068D XCB418 70515-01 Precision Xtra Advanced Diabetes Management System 03K068D XCB482 70515-01 Precision Xtra Advanced Diabetes Management System 03K088D XCB538 70515-01 Precision Xtra Advanced Diabetes Management System 03K088D XCB539 70515-01 Precision Xtra Advanced Diabetes Management System 03K092D XCB566 70515-01 Precision Xtra Advanced Diabetes Management System 03K099D XCB567 70515-01 Precision Xtra Advanced Diabetes Management System 03K099D XCB596 70515-01 Precision Xtra Advanced Diabetes Management System 03K101D XCB596 70515-01 Precision Xtra Advanced Diabetes Management System 03K101D XCB597 70515-01 Precision Xtra Advanced Diabetes Management System 03K101D XCB599 70515-01 Precision Xtra Advanced Diabetes Management System 03K106D XCB625 70515-01 Precision Xtra Advanced Diabetes Management System 03K106D XCB632 70515-01 Precision Xtra Advanced Diabetes Management System 03K107D XCB625 70515-01 Precision Xtra Advanced Diabetes Management System 03K107D XCB632 70515-01 Precision Xtra Advanced Diabetes Management System 03K113D XCB646 70515-01 Precision Xtra Advanced Diabetes Management System 03K113D XCB650 70515-01 Precision Xtra Advanced Diabetes Management System 03K113D XCB662 70515-01 Precision Xtra Advanced Diabetes Management System 03K114D XCB646 70515-01 Precision Xtra Advanced Diabetes Management System 03K114D XCB650 70515-01 Precision Xtra Advanced Diabetes Management System 03K114D XCB662 70515-01 Precision Xtra Advanced Diabetes Management System 03K128D XCB684 70515-01 Precision Xtra Advanced Diabetes Management System 03K128D XCB708 70515-01 Precision Xtra Advanced Diabetes Management System 03K128D XCB713 70515-01 Precision Xtra Advanced Diabetes Management System 03K129D XCB670 70515-01 Precision Xtra Advanced Diabetes Management System 03K129D XCB672 70515-01 Precision Xtra Advanced Diabetes Management System 03K129D XCB679 70515-01 Precision Xtra Advanced Diabetes Management System 03K130D XCB670 70515-01 Precision Xtra Advanced Diabetes Management System 03K130D XCB673 70515-01 Precision Xtra Advanced Diabetes Management System 03K130D XCB685 70515-01 Precision Xtra Advanced Diabetes Management System 03K151D XCB740 70515-01 Precision Xtra Advanced Diabetes Management System 03K151D XCB746 70515-01 Precision Xtra Advanced Diabetes Management System 03K151D XCB747 70515-01 Precision Xtra Advanced Diabetes Management System 03K155D XCB740 70515-01 Precision Xtra Advanced Diabetes Management System 03K155D XCB746 70515-01 Precision Xtra Advanced Diabetes Management System 03K155D XCB747 70515-01 Precision Xtra Advanced Diabetes Management System 03K164D XCB738 70515-01 Precision Xtra Advanced Diabetes Management System 03K164D XCB762 70515-01 Precision Xtra Advanced Diabetes Management System 03K164D XCB803 70515-01 Precision Xtra Advanced Diabetes Management System 03K184D XCB829 70515-01 Precision Xtra Advanced Diabetes Management System 03K184D XCB831 70515-01 Precision Xtra Advanced Diabetes Management System 03K184D XCB832 70515-01 Precision Xtra Advanced Diabetes Management System 03K186D XCB829 70515-01 Precision Xtra Advanced Diabetes Management System 03K186D XCB831 70515-01 Precision Xtra Advanced Diabetes Management System 03K186D XCB832 70515-01 Precision Xtra Advanced Diabetes Management System 04K058D XCB381 70515-01 Precision Xtra Advanced Diabetes Management System 04K058D XCB391 70515-01 Precision Xtra Advanced Diabetes Management System 04K058D XCB394 70515-01 Precision Xtra Advanced Diabetes Management System 04K058D XCB395 70515-01 Precision Xtra Advanced Diabetes Management System 04K124D XCB670 70515-01 Precision Xtra Advanced Diabetes Management System 04K127D XCB671 70515-01 Precision Xtra Advanced Diabetes Management System 04K127D XCB673 70515-01 Precision Xtra Advanced Diabetes Management System 04K128D XCB665 70515-01 Precision Xtra Advanced Diabetes Management System 04K128D XCB669 70515-01 Precision Xtra Advanced Diabetes Management System 04K128D XCB671 70515-01 Precision Xtra Advanced Diabetes Management System 04K130D XCB670 70515-01 Precision Xtra Advanced Diabetes Management System 04K130D XCB672 70515-01 Precision Xtra Advanced Diabetes Management System 04K130D XCB679 70515-01 Precision Xtra Advanced Diabetes Management System 04K183D XCB829 70515-01 Precision Xtra Advanced Diabetes Management System 04K183D XCB831 70515-01 Precision Xtra Advanced Diabetes Management System 04K183D XCB832 70515-01 Precision Xtra Advanced Diabetes Management System 04K184D XCB830 70515-01 Precision Xtra Advanced Diabetes Management System 04K184D XCB831 70515-01 Precision Xtra Advanced Diabetes Management System 04K201D XCB863 70515-01 Precision Xtra Advanced Diabetes Management System 04K201D XCB864 70515-01 Precision Xtra Advanced Diabetes Management System 04K202D XCB863 70515-01 Precision Xtra Advanced Diabetes Management System 04K202D XCB864 70515-01 Precision Xtra Advanced Diabetes Management System 05K066D XCB416 70515-01 Precision Xtra Advanced Diabetes Management System 05K066D XCB417 70515-01 Precision Xtra Advanced Diabetes Management System 05K066D XCB418 70515-01 Precision Xtra Advanced Diabetes Management System 05K066D XCB482 70515-01 Precision Xtra Advanced Diabetes Management System 05K067D XCB416 70515-01 Precision Xtra Advanced Diabetes Management System 05K067D XCB417 70515-01 Precision Xtra Advanced Diabetes Management System 05K067D XCB418 70515-01 Precision Xtra Advanced Diabetes Management System 05K067D XCB482 70515-01 Precision Xtra Advanced Diabetes Management System 05K072D XCB416 70515-01 Precision Xtra Advanced Diabetes Management System 05K072D XCB418 70515-01 Precision Xtra Advanced Diabetes Management System 05K072D XCB492 70515-01 Precision Xtra Advanced Diabetes Management System 05K087D XCB536 70515-01 Precision Xtra Advanced Diabetes Management System 05K087D XCB538 70515-01 Precision Xtra Advanced Diabetes Management System 05K100D XCB567 70515-01 Precision Xtra Advanced Diabetes Management System 05K100D XCB596 70515-01 Precision Xtra Advanced Diabetes Management System 05K108D XCB644 70515-01 Precision Xtra Advanced Diabetes Management System 05K127D XCB684 70515-01 Precision Xtra Advanced Diabetes Management System 05K127D XCB708 70515-01 Precision Xtra Advanced Diabetes Management System 05K127D XCB713 70515-01 Precision Xtra Advanced Diabetes Management System 05K129D XCB670 70515-01 Precision Xtra Advanced Diabetes Management System 05K129D XCB673 70515-01 Precision Xtra Advanced Diabetes Management System 05K129D XCB685 70515-01 Precision Xtra Advanced Diabetes Management System 05K152D XCB740 70515-01 Precision Xtra Advanced Diabetes Management System 05K152D XCB746 70515-01 Precision Xtra Advanced Diabetes Management System 05K152D XCB747 70515-01 Precision Xtra Advanced Diabetes Management System 05K156D XCB745 70515-01 Precision Xtra Advanced Diabetes Management System 05K156D XCB746 70515-01 Precision Xtra Advanced Diabetes Management System 05K156D XCB748 70515-01 Precision Xtra Advanced Diabetes Management System 05K156D XCB751 70515-01 Precision Xtra Advanced Diabetes Management System 05K183D XCB830 70515-01 Precision Xtra Advanced Diabetes Management System 05K183D XCB831 70515-01 Precision Xtra Advanced Diabetes Management System 05K200D XCB798 70515-01 Precision Xtra Advanced Diabetes Management System 05K200D XCB864 70515-01 Precision Xtra Advanced Diabetes Management System 05K200D XCB865 70515-01 Precision Xtra Advanced Diabetes Management System 05K200D XCB866 70515-01 Precision Xtra Advanced Diabetes Management System 06K087D XCB538 70515-01 Precision Xtra Advanced Diabetes Management System 06K087D XCB539 70515-01 Precision Xtra Advanced Diabetes Management System 06K127D XCB672 70515-01 Precision Xtra Advanced Diabetes Management System 06K127D XCB674 70515-01 Precision Xtra Advanced Diabetes Management System 06K127D XCB684 70515-01 Precision Xtra Advanced Diabetes Management System 06K204D XCB798 70515-01 Precision Xtra Advanced Diabetes Management System 06K204D XCB864 70515-01 Precision Xtra Advanced Diabetes Management System 06K204D XCB865 70515-01 Precision Xtra Advanced Diabetes Management System 06K204D XCB866 70515-01 Precision Xtra Advanced Diabetes Management System 07K053D XCB381 70515-01 Precision Xtra Advanced Diabetes Management System 07K053D XCB391 70515-01 Precision Xtra Advanced Diabetes Management System 07K053D XCB394 70515-01 Precision Xtra Advanced Diabetes Management System 07K053D XCB395 70515-01 Precision Xtra Advanced Diabetes Management System 07K124D XCB671 70515-01 Precision Xtra Advanced Diabetes Management System 07K124D XCB673 70515-01 Precision Xtra Advanced Diabetes Management System 07K127D XCB670 70515-01 Precision Xtra Advanced Diabetes Management System 07K127D XCB673 70515-01 Precision Xtra Advanced Diabetes Management System 07K127D XCB685 70515-01 Precision Xtra Advanced Diabetes Management System 07K200D XCB863 70515-01 Precision Xtra Advanced Diabetes Management System 07K200D XCB864 70515-01 Precision Xtra Advanced Diabetes Management System 08K089D XCB566 70515-01 Precision Xtra Advanced Diabetes Management System 08K127D XCB665 70515-01 Precision Xtra Advanced Diabetes Management System 08K127D XCB669 70515-01 Precision Xtra Advanced Diabetes Management System 08K127D XCB671 70515-01 Precision Xtra Advanced Diabetes Management System 09K100D XCB596 70515-01 Precision Xtra Advanced Diabetes Management System 09K100D XCB597 70515-01 Precision Xtra Advanced Diabetes Management System 09K100D XCB599 70515-01 Precision Xtra Advanced Diabetes Management System 09K107D XCB644 70515-01 Precision Xtra Advanced Diabetes Management System 09K110D XCB646 70515-01 Precision Xtra Advanced Diabetes Management System 09K110D XCB650 70515-01 Precision Xtra Advanced Diabetes Management System 09K110D XCB662 70515-01 Precision Xtra Advanced Diabetes Management System 09K150D XCB740 70515-01 Precision Xtra Advanced Diabetes Management System 09K150D XCB746 70515-01 Precision Xtra Advanced Diabetes Management System 09K150D XCB747 70515-01 Precision Xtra Advanced Diabetes Management System 09K151D XCB745 70515-01 Precision Xtra Advanced Diabetes Management System 09K151D XCB746 70515-01 Precision Xtra Advanced Diabetes Management System 09K151D XCB748 70515-01 Precision Xtra Advanced Diabetes Management System 09K151D XCB751 70515-01 Precision Xtra Advanced Diabetes Management System 09K163D XCB738 70515-01 Precision Xtra Advanced Diabetes Management System 09K163D XCB762 70515-01 Precision Xtra Advanced Diabetes Management System 09K163D XCB803 70515-01 Precision Xtra Advanced Diabetes Management System 10K110D XCB662 70515-01 Precision Xtra Advanced Diabetes Management System 10K123D XCB670 70517-50 Medline Optium Blood Glucose Monitoring System 01K092D XCB578 70517-50 Medline Optium Blood Glucose Monitoring System 01K123D XCB649 70517-50 Medline Optium Blood Glucose Monitoring System 01K158D XCB773 70517-50 Medline Optium Blood Glucose Monitoring System 01K158D XCB790 70517-50 Medline Optium Blood Glucose Monitoring System 03K108D XCB637 70517-50 Medline Optium Blood Glucose Monitoring System 03K115D XCB637 70517-50 Medline Optium Blood Glucose Monitoring System 03K115D XCB647 70517-50 Medline Optium Blood Glucose Monitoring System 03K115D XCB649 70517-50 Medline Optium Blood Glucose Monitoring System 03K122D XCB649 70517-50 Medline Optium Blood Glucose Monitoring System 04K158D XCB773 70517-50 Medline Optium Blood Glucose Monitoring System 05K094D XCB578 70517-50 Medline Optium Blood Glucose Monitoring System 07K114D XCB637 70517-50 Medline Optium Blood Glucose Monitoring System 07K114D XCB647 70517-50 Medline Optium Blood Glucose Monitoring System 07K114D XCB649 70517-50 Medline Optium Blood Glucose Monitoring System 08K107D XCB637 70517-50 Medline Optium Blood Glucose Monitoring System 08K173D XCB827 70517-50 Medline Optium Blood Glucose Monitoring System 09K157D XCB773 70517-50 Medline Optium Blood Glucose Monitoring System 09K157D XCB790 70517-50 Medline Optium Blood Glucose Monitoring System 15K172D XCB827 70708-01 OMRON HEA-214 01K181E XCB767 70708-01 OMRON HEA-214 01K183E XCB767 70708-01 OMRON HEA-214 01K183E XCB769 70708-01 OMRON HEA-214 01K184E XCB769 70708-01 OMRON HEA-214 01K184E XCB770 70708-01 OMRON HEA-214 01K185E XCB770 70708-01 OMRON HEA-214 01K186E XCB771 70708-01 OMRON HEA-214 01K187E XCB772 70708-01 OMRON HEA-214 01K187E XCB899 70708-01 OMRON HEA-214 01K188E XCB899 70708-01 OMRON HEA-214 01K190E XCB893 70708-01 OMRON HEA-214 01K190E XCB897 70708-01 OMRON HEA-214 01K191E XCB897 70708-01 OMRON HEA-214 01K192E XCB896 70708-01 OMRON HEA-214 01K192E XCB902 70708-01 OMRON HEA-214 02K178E XCB766 70708-01 OMRON HEA-214 02K179E XCB766 70708-01 OMRON HEA-214 02K180E XCB767 70708-01 OMRON HEA-214 02K185E XCB768 70708-01 OMRON HEA-214 02K185E XCB770 70708-01 OMRON HEA-214 02K185E XCB771 70708-01 OMRON HEA-214 02K185E XCB772 70708-01 OMRON HEA-214 02K186E XCB771 70708-01 OMRON HEA-214 02K186E XCB772 70708-01 OMRON HEA-214 02K187E XCB772 70708-01 OMRON HEA-214 02K188E XCB893 70708-01 OMRON HEA-214 02K188E XCB899 70708-01 OMRON HEA-214 02K190E XCB893 70708-01 OMRON HEA-214 02K192E XCB902 70708-01 OMRON HEA-214 03K177E XCB766 70708-01 OMRON HEA-214 03K184E XCB769 70708-01 OMRON HEA-214 03K191E XCB896 70708-01 OMRON HEA-214 03K191E XCB897 70708-01 OMRON HEA-214 04K178E XCB766 70708-01 OMRON HEA-214 04K179E XCB766 70708-01 OMRON HEA-214 04K179E XCB767 70708-01 OMRON HEA-214 04K183E XCB769 70708-01 OMRON HEA-214 05K180E XCB766 70708-01 OMRON HEA-214 06K179E XCB767 70722-01 Optium Xido Glucose Monitoring System 03K120C XCB376 70722-01 Optium Xido Glucose Monitoring System 05K120C XCB375 70722-01 Optium Xido Glucose Monitoring System 05K120C XCB377 70722-01 Optium Xido Glucose Monitoring System 06K120C XCB376 70722-01 Optium Xido Glucose Monitoring System 06K120C XCB377 70722-01 Optium Xido Glucose Monitoring System 07K120C XCB375 70722-01 Optium Xido Glucose Monitoring System 08K179C XCB420 70722-01 Optium Xido Glucose Monitoring System 12K120C XCB375 70722-01 Optium Xido Glucose Monitoring System 13K120C XCB377 70732-35 Optium Xceed Diabetes Monitoring System 01K197C XCB792 70732-35 Optium Xceed Diabetes Monitoring System 03K180C XCB374 70732-35 Optium Xceed Diabetes Monitoring System 03K180C XCB379 70732-35 Optium Xceed Diabetes Monitoring System 03K192C XCB734 70732-35 Optium Xceed Diabetes Monitoring System 03K192C XCB792 70732-35 Optium Xceed Diabetes Monitoring System 03K193C XCB792 70732-35 Optium Xceed Diabetes Monitoring System 03K193C XCB793 70732-35 Optium Xceed Diabetes Monitoring System 10K173C XCB374 70732-35 Optium Xceed Diabetes Monitoring System 10K173C XCB415 70732-35 Optium Xceed Diabetes Monitoring System 10K193C XCB793 70732-35 Optium Xceed Diabetes Monitoring System 11K192C XCB792 70732-35 Optium Xceed Diabetes Monitoring System 12K180C XCB411 70732-35 Optium Xceed Diabetes Monitoring System 12K180C XCB415 70732-35 Optium Xceed Diabetes Monitoring System 13K206C XCB795 70732-35 Optium Xceed Diabetes Monitoring System 14K176C XCB411 70732-35 Optium Xceed Diabetes Monitoring System 14K192C XCB792 70732-35 Optium Xceed Diabetes Monitoring System 14K192C XCB793 70732-35 Optium Xceed Diabetes Monitoring System 14K194C XCB792 70732-35 Optium Xceed Diabetes Monitoring System 15K172C XCB374 70732-35 Optium Xceed Diabetes Monitoring System 15K172C XCB415 70732-35 Optium Xceed Diabetes Monitoring System 15K191C XCB734 70732-35 Optium Xceed Diabetes Monitoring System 15K191C XCB792 70732-35 Optium Xceed Diabetes Monitoring System 16K177C XCB415 70732-35 Optium Xceed Diabetes Monitoring System 18K193C XCB792 70733-35 Optium Xceed Diabetes Monitoring System 01K136C XCB540 70733-35 Optium Xceed Diabetes Monitoring System 01K136C XCB553 70733-35 Optium Xceed Diabetes Monitoring System 01K136C XCB645 70733-35 Optium Xceed Diabetes Monitoring System 01K190C XCB452 70733-35 Optium Xceed Diabetes Monitoring System 01K191C XCB452 70733-35 Optium Xceed Diabetes Monitoring System 01K191C XCB453 70733-35 Optium Xceed Diabetes Monitoring System 02K194C XCB779 70733-35 Optium Xceed Diabetes Monitoring System 02K211C XCB805 70733-35 Optium Xceed Diabetes Monitoring System 03K199C XCB805 70733-35 Optium Xceed Diabetes Monitoring System 03K199C XCB809 70733-35 Optium Xceed Diabetes Monitoring System 05K135C XCB521 70733-35 Optium Xceed Diabetes Monitoring System 05K135C XCB540 70733-35 Optium Xceed Diabetes Monitoring System 05K200C XCB786 70733-35 Optium Xceed Diabetes Monitoring System 05K200C XCB805 70733-35 Optium Xceed Diabetes Monitoring System 06K190C XCB452 70733-35 Optium Xceed Diabetes Monitoring System 06K190C XCB453 70733-35 Optium Xceed Diabetes Monitoring System 07K190C XCB452 70733-35 Optium Xceed Diabetes Monitoring System 07K190C XCB540 70733-35 Optium Xceed Diabetes Monitoring System 07K190C XCB608 70733-35 Optium Xceed Diabetes Monitoring System 07K190C XCB758 70733-35 Optium Xceed Diabetes Monitoring System 07K190C XCB775 70733-35 Optium Xceed Diabetes Monitoring System 07K200C XCB809 70733-35 Optium Xceed Diabetes Monitoring System 07K204C XCB786 70733-35 Optium Xceed Diabetes Monitoring System 07K204C XCB805 70733-35 Optium Xceed Diabetes Monitoring System 08K137C XCB540 70733-35 Optium Xceed Diabetes Monitoring System 08K137C XCB553 70733-35 Optium Xceed Diabetes Monitoring System 08K137C XCB645 70733-35 Optium Xceed Diabetes Monitoring System 08K194C XCB776 70733-35 Optium Xceed Diabetes Monitoring System 08K194C XCB779 70733-35 Optium Xceed Diabetes Monitoring System 09K135C XCB521 70733-35 Optium Xceed Diabetes Monitoring System 09K197C XCB776 70733-35 Optium Xceed Diabetes Monitoring System 09K197C XCB779 70733-35 Optium Xceed Diabetes Monitoring System 10K191C XCB452 70733-35 Optium Xceed Diabetes Monitoring System 10K191C XCB540 70733-35 Optium Xceed Diabetes Monitoring System 10K191C XCB608 70733-35 Optium Xceed Diabetes Monitoring System 10K191C XCB758 70733-35 Optium Xceed Diabetes Monitoring System 10K191C XCB775 70733-35 Optium Xceed Diabetes Monitoring System 12K190C XCB453 70733-35 Optium Xceed Diabetes Monitoring System 12K204C XCB805 70733-35 Optium Xceed Diabetes Monitoring System 12K204C XCB809 70733-35 Optium Xceed Diabetes Monitoring System 13K199C XCB809 70733-35 Optium Xceed Diabetes Monitoring System 16K187C XCB452 70733-35 Optium Xceed Diabetes Monitoring System 16K193C XCB779 70733-35 Optium Xceed Diabetes Monitoring System 16K198C XCB805 70733-35 Optium Xceed Diabetes Monitoring System 16K198C XCB809 70733-35 Optium Xceed Diabetes Monitoring System 17K199C XCB786 70733-35 Optium Xceed Diabetes Monitoring System 17K199C XCB805 70733-35 Optium Xceed Diabetes Monitoring System 18K208C XCB805 70733-35 Optium Xceed Diabetes Monitoring System 19K135C XCB540 70733-35 Optium Xceed Diabetes Monitoring System 19K135C XCB553 70733-35 Optium Xceed Diabetes Monitoring System 19K135C XCB645 70737-35 Optium Xceed Diabetes Monitoring System 03K194C XCB786 70737-35 Optium Xceed Diabetes Monitoring System 13K170C XCB397 70737-35 Optium Xceed Diabetes Monitoring System 13K170C XCB452 70737-35 Optium Xceed Diabetes Monitoring System 15K193C XCB783 70737-35 Optium Xceed Diabetes Monitoring System 15K193C XCB786 70737-35 Optium Xceed Diabetes Monitoring System 17K193C XCB786 70755-35 Optium Xceed Diabetes Monitoring System 06K170C XCB379 70755-35 Optium Xceed Diabetes Monitoring System 10K177C XCB379 70755-35 Optium Xceed Diabetes Monitoring System 13K176C XCB379 70755-35 Optium Xceed Diabetes Monitoring System 18K171C XCB379 70756-35 Optium Xceed Diabetes Monitoring System 15K169C XCB374 70899-01 Optium Blood Glucose Monitoring System 01K093D XCB510 70899-01 Optium Blood Glucose Monitoring System 01K093D XCB514 70899-01 Optium Blood Glucose Monitoring System 01K115D XCB548 70899-01 Optium Blood Glucose Monitoring System 01K115D XCB554 70899-01 Optium Blood Glucose Monitoring System 01K173D XCB718 70899-01 Optium Blood Glucose Monitoring System 01K173D XCB719 70899-01 Optium Blood Glucose Monitoring System 01K173D XCB722 70899-01 Optium Blood Glucose Monitoring System 02K174D XCB718 70899-01 Optium Blood Glucose Monitoring System 02K174D XCB719 70899-01 Optium Blood Glucose Monitoring System 02K174D XCB722 70899-01 Optium Blood Glucose Monitoring System 03K071D XCB438 70899-01 Optium Blood Glucose Monitoring System 04K086D XCB438 70899-01 Optium Blood Glucose Monitoring System 04K086D XCB440 70899-01 Optium Blood Glucose Monitoring System 04K086D XCB479 70899-01 Optium Blood Glucose Monitoring System 04K174D XCB719 70899-01 Optium Blood Glucose Monitoring System 04K174D XCB721 70899-01 Optium Blood Glucose Monitoring System 04K176D XCB718 70899-01 Optium Blood Glucose Monitoring System 04K176D XCB719 70899-01 Optium Blood Glucose Monitoring System 04K176D XCB722 70899-01 Optium Blood Glucose Monitoring System 05K074D XCB438 70899-01 Optium Blood Glucose Monitoring System 05K093D XCB510 70899-01 Optium Blood Glucose Monitoring System 05K093D XCB513 70899-01 Optium Blood Glucose Monitoring System 05K113D XCB548 70899-01 Optium Blood Glucose Monitoring System 06K074D XCB405 70899-01 Optium Blood Glucose Monitoring System 06K074D XCB438 70899-01 Optium Blood Glucose Monitoring System 06K173D XCB719 70899-01 Optium Blood Glucose Monitoring System 06K173D XCB721 70899-01 Optium Blood Glucose Monitoring System 06K176D XCB719 70899-01 Optium Blood Glucose Monitoring System 06K176D XCB721 70899-01 Optium Blood Glucose Monitoring System 08K085D XCB438 70899-01 Optium Blood Glucose Monitoring System 08K085D XCB440 70899-01 Optium Blood Glucose Monitoring System 08K085D XCB479 70899-01 Optium Blood Glucose Monitoring System 10K074D XCB438 70899-01 Optium Blood Glucose Monitoring System 10K074D XCB440 70899-01 Optium Blood Glucose Monitoring System 11K114D XCB548 70899-01 Optium Blood Glucose Monitoring System 11K114D XCB554 70899-01 Optium Blood Glucose Monitoring System 13K172D XCB719 70899-01 Optium Blood Glucose Monitoring System 13K172D XCB721 70899-01 Optium Blood Glucose Monitoring System 14K172D XCB718 70899-01 Optium Blood Glucose Monitoring System 14K172D XCB719 70899-01 Optium Blood Glucose Monitoring System 14K172D XCB722 80146-50 Rite-Aid Blood Glucose Monitoring System 01K130D XCB658 80146-50 Rite-Aid Blood Glucose Monitoring System 01K130D XCB659 80146-50 Rite-Aid Blood Glucose Monitoring System 01K130D XCB661 80146-50 Rite-Aid Blood Glucose Monitoring System 01K130D XCB663 80146-50 Rite-Aid Blood Glucose Monitoring System 02K110D XCB659 80146-50 Rite-Aid Blood Glucose Monitoring System 02K110D XCB661 80146-50 Rite-Aid Blood Glucose Monitoring System 06K088D XCB529 80146-50 Rite-Aid Blood Glucose Monitoring System 06K095D XCB528 80146-50 Rite-Aid Blood Glucose Monitoring System 06K095D XCB529 80146-50 Rite-Aid Blood Glucose Monitoring System 06K129D XCB658 80146-50 Rite-Aid Blood Glucose Monitoring System 06K129D XCB659 80146-50 Rite-Aid Blood Glucose Monitoring System 06K129D XCB661 80146-50 Rite-Aid Blood Glucose Monitoring System 06K129D XCB663 80146-50 Rite-Aid Blood Glucose Monitoring System 07K100D XCB528 80146-50 Rite-Aid Blood Glucose Monitoring System 08K109D XCB659 80146-50 Rite-Aid Blood Glucose Monitoring System 08K109D XCB661 80347-20 Precision Xtra Advanced Diabetes Management System 01K082D XCB506 80347-20 Precision Xtra Advanced Diabetes Management System 01K082D XCB508 80347-20 Precision Xtra Advanced Diabetes Management System 01K089D XCB539 80347-20 Precision Xtra Advanced Diabetes Management System 01K100D XCB606 80347-20 Precision Xtra Advanced Diabetes Management System 01K100D XCB612 80347-20 Precision Xtra Advanced Diabetes Management System 01K108D XCB631 80347-20 Precision Xtra Advanced Diabetes Management System 01K108D XCB638 80347-20 Precision Xtra Advanced Diabetes Management System 01K109D XCB625 80347-20 Precision Xtra Advanced Diabetes Management System 01K109D XCB631 80347-20 Precision Xtra Advanced Diabetes Management System 01K109D XCB640 80347-20 Precision Xtra Advanced Diabetes Management System 01K135D XCB726 80347-20 Precision Xtra Advanced Diabetes Management System 01K135D XCB728 80347-20 Precision Xtra Advanced Diabetes Management System 01K145D XCB732 80347-20 Precision Xtra Advanced Diabetes Management System 01K145D XCB736 80347-20 Precision Xtra Advanced Diabetes Management System 01K150D XCB738 80347-20 Precision Xtra Advanced Diabetes Management System 01K150D XCB742 80347-20 Precision Xtra Advanced Diabetes Management System 01K150D XCB743 80347-20 Precision Xtra Advanced Diabetes Management System 01K150D XCB744 80347-20 Precision Xtra Advanced Diabetes Management System 01K192D XCB798 80347-20 Precision Xtra Advanced Diabetes Management System 01K211D XCB860 80347-20 Precision Xtra Advanced Diabetes Management System 01K211D XCB861 80347-20 Precision Xtra Advanced Diabetes Management System 02K074D XCB446 80347-20 Precision Xtra Advanced Diabetes Management System 02K074D XCB447 80347-20 Precision Xtra Advanced Diabetes Management System 02K079D XCB504 80347-20 Precision Xtra Advanced Diabetes Management System 02K079D XCB505 80347-20 Precision Xtra Advanced Diabetes Management System 02K080D XCB495 80347-20 Precision Xtra Advanced Diabetes Management System 02K080D XCB506 80347-20 Precision Xtra Advanced Diabetes Management System 02K080D XCB508 80347-20 Precision Xtra Advanced Diabetes Management System 02K081D XCB506 80347-20 Precision Xtra Advanced Diabetes Management System 02K081D XCB508 80347-20 Precision Xtra Advanced Diabetes Management System 02K106D XCB627 80347-20 Precision Xtra Advanced Diabetes Management System 02K135D XCB716 80347-20 Precision Xtra Advanced Diabetes Management System 02K145D XCB739 80347-20 Precision Xtra Advanced Diabetes Management System 02K145D XCB741 80347-20 Precision Xtra Advanced Diabetes Management System 02K145D XCB742 80347-20 Precision Xtra Advanced Diabetes Management System 02K151D XCB749 80347-20 Precision Xtra Advanced Diabetes Management System 02K151D XCB765 80347-20 Precision Xtra Advanced Diabetes Management System 02K156D XCB735 80347-20 Precision Xtra Advanced Diabetes Management System 02K156D XCB736 80347-20 Precision Xtra Advanced Diabetes Management System 02K156D XCB738 80347-20 Precision Xtra Advanced Diabetes Management System 02K157D XCB735 80347-20 Precision Xtra Advanced Diabetes Management System 02K157D XCB736 80347-20 Precision Xtra Advanced Diabetes Management System 02K157D XCB738 80347-20 Precision Xtra Advanced Diabetes Management System 02K162D XCB741 80347-20 Precision Xtra Advanced Diabetes Management System 02K162D XCB763 80347-20 Precision Xtra Advanced Diabetes Management System 02K186D XCB830 80347-20 Precision Xtra Advanced Diabetes Management System 02K186D XCB833 80347-20 Precision Xtra Advanced Diabetes Management System 02K186D XCB834 80347-20 Precision Xtra Advanced Diabetes Management System 02K186D XCB843 80347-20 Precision Xtra Advanced Diabetes Management System 02K186D XCB845 80347-20 Precision Xtra Advanced Diabetes Management System 02K187D XCB830 80347-20 Precision Xtra Advanced Diabetes Management System 02K187D XCB833 80347-20 Precision Xtra Advanced Diabetes Management System 02K187D XCB834 80347-20 Precision Xtra Advanced Diabetes Management System 02K187D XCB843 80347-20 Precision Xtra Advanced Diabetes Management System 02K187D XCB845 80347-20 Precision Xtra Advanced Diabetes Management System 03K080D XCB504 80347-20 Precision Xtra Advanced Diabetes Management System 03K080D XCB505 80347-20 Precision Xtra Advanced Diabetes Management System 03K109D XCB631 80347-20 Precision Xtra Advanced Diabetes Management System 03K109D XCB638 80347-20 Precision Xtra Advanced Diabetes Management System 03K110D XCB627 80347-20 Precision Xtra Advanced Diabetes Management System 03K110D XCB653 80347-20 Precision Xtra Advanced Diabetes Management System 03K110D XCB662 80347-20 Precision Xtra Advanced Diabetes Management System 03K167D XCB800 80347-20 Precision Xtra Advanced Diabetes Management System 03K167D XCB803 80347-20 Precision Xtra Advanced Diabetes Management System 03K183D XCB830 80347-20 Precision Xtra Advanced Diabetes Management System 03K183D XCB833 80347-20 Precision Xtra Advanced Diabetes Management System 03K183D XCB834 80347-20 Precision Xtra Advanced Diabetes Management System 03K183D XCB843 80347-20 Precision Xtra Advanced Diabetes Management System 03K183D XCB845 80347-20 Precision Xtra Advanced Diabetes Management System 03K202D XCB860 80347-20 Precision Xtra Advanced Diabetes Management System 03K202D XCB863 80347-20 Precision Xtra Advanced Diabetes Management System 03K202D XCB870 80347-20 Precision Xtra Advanced Diabetes Management System 04K079D XCB495 80347-20 Precision Xtra Advanced Diabetes Management System 04K079D XCB504 80347-20 Precision Xtra Advanced Diabetes Management System 04K080D XCB506 80347-20 Precision Xtra Advanced Diabetes Management System 04K080D XCB508 80347-20 Precision Xtra Advanced Diabetes Management System 04K089D XCB504 80347-20 Precision Xtra Advanced Diabetes Management System 04K089D XCB508 80347-20 Precision Xtra Advanced Diabetes Management System 04K089D XCB536 80347-20 Precision Xtra Advanced Diabetes Management System 04K109D XCB627 80347-20 Precision Xtra Advanced Diabetes Management System 04K109D XCB653 80347-20 Precision Xtra Advanced Diabetes Management System 04K109D XCB662 80347-20 Precision Xtra Advanced Diabetes Management System 04K110D XCB625 80347-20 Precision Xtra Advanced Diabetes Management System 04K110D XCB631 80347-20 Precision Xtra Advanced Diabetes Management System 04K110D XCB640 80347-20 Precision Xtra Advanced Diabetes Management System 04K134D XCB728 80347-20 Precision Xtra Advanced Diabetes Management System 04K134D XCB729 80347-20 Precision Xtra Advanced Diabetes Management System 04K145D XCB738 80347-20 Precision Xtra Advanced Diabetes Management System 04K145D XCB742 80347-20 Precision Xtra Advanced Diabetes Management System 04K145D XCB743 80347-20 Precision Xtra Advanced Diabetes Management System 04K145D XCB744 80347-20 Precision Xtra Advanced Diabetes Management System 04K151D XCB740 80347-20 Precision Xtra Advanced Diabetes Management System 04K151D XCB741 80347-20 Precision Xtra Advanced Diabetes Management System 04K190D XCB830 80347-20 Precision Xtra Advanced Diabetes Management System 04K190D XCB833 80347-20 Precision Xtra Advanced Diabetes Management System 04K190D XCB834 80347-20 Precision Xtra Advanced Diabetes Management System 04K190D XCB843 80347-20 Precision Xtra Advanced Diabetes Management System 04K190D XCB845 80347-20 Precision Xtra Advanced Diabetes Management System 04K191D XCB830 80347-20 Precision Xtra Advanced Diabetes Management System 04K191D XCB833 80347-20 Precision Xtra Advanced Diabetes Management System 04K191D XCB834 80347-20 Precision Xtra Advanced Diabetes Management System 04K191D XCB843 80347-20 Precision Xtra Advanced Diabetes Management System 04K191D XCB845 80347-20 Precision Xtra Advanced Diabetes Management System 04K204D XCB860 80347-20 Precision Xtra Advanced Diabetes Management System 04K204D XCB863 80347-20 Precision Xtra Advanced Diabetes Management System 04K204D XCB870 80347-20 Precision Xtra Advanced Diabetes Management System 04K208D XCB860 80347-20 Precision Xtra Advanced Diabetes Management System 04K208D XCB861 80347-20 Precision Xtra Advanced Diabetes Management System 05K073D XCB418 80347-20 Precision Xtra Advanced Diabetes Management System 05K073D XCB446 80347-20 Precision Xtra Advanced Diabetes Management System 05K079D XCB495 80347-20 Precision Xtra Advanced Diabetes Management System 05K079D XCB506 80347-20 Precision Xtra Advanced Diabetes Management System 05K079D XCB508 80347-20 Precision Xtra Advanced Diabetes Management System 05K114D XCB654 80347-20 Precision Xtra Advanced Diabetes Management System 05K114D XCB662 80347-20 Precision Xtra Advanced Diabetes Management System 05K134D XCB716 80347-20 Precision Xtra Advanced Diabetes Management System 05K134D XCB726 80347-20 Precision Xtra Advanced Diabetes Management System 05K144D XCB732 80347-20 Precision Xtra Advanced Diabetes Management System 05K144D XCB737 80347-20 Precision Xtra Advanced Diabetes Management System 05K145D XCB732 80347-20 Precision Xtra Advanced Diabetes Management System 05K145D XCB737 80347-20 Precision Xtra Advanced Diabetes Management System 05K149D XCB738 80347-20 Precision Xtra Advanced Diabetes Management System 05K149D XCB742 80347-20 Precision Xtra Advanced Diabetes Management System 05K149D XCB743 80347-20 Precision Xtra Advanced Diabetes Management System 05K149D XCB744 80347-20 Precision Xtra Advanced Diabetes Management System 05K150D XCB740 80347-20 Precision Xtra Advanced Diabetes Management System 05K150D XCB741 80347-20 Precision Xtra Advanced Diabetes Management System 05K151D XCB739 80347-20 Precision Xtra Advanced Diabetes Management System 05K151D XCB741 80347-20 Precision Xtra Advanced Diabetes Management System 05K151D XCB742 80347-20 Precision Xtra Advanced Diabetes Management System 05K171D XCB800 80347-20 Precision Xtra Advanced Diabetes Management System 05K184D XCB830 80347-20 Precision Xtra Advanced Diabetes Management System 05K184D XCB833 80347-20 Precision Xtra Advanced Diabetes Management System 05K184D XCB834 80347-20 Precision Xtra Advanced Diabetes Management System 05K184D XCB843 80347-20 Precision Xtra Advanced Diabetes Management System 05K184D XCB845 80347-20 Precision Xtra Advanced Diabetes Management System 06K144D XCB732 80347-20 Precision Xtra Advanced Diabetes Management System 06K144D XCB736 80347-20 Precision Xtra Advanced Diabetes Management System 06K145D XCB740 80347-20 Precision Xtra Advanced Diabetes Management System 06K145D XCB741 80347-20 Precision Xtra Advanced Diabetes Management System 06K150D XCB739 80347-20 Precision Xtra Advanced Diabetes Management System 06K150D XCB741 80347-20 Precision Xtra Advanced Diabetes Management System 06K150D XCB742 80347-20 Precision Xtra Advanced Diabetes Management System 06K151D XCB749 80347-20 Precision Xtra Advanced Diabetes Management System 06K151D XCB750 80347-20 Precision Xtra Advanced Diabetes Management System 06K159D XCB741 80347-20 Precision Xtra Advanced Diabetes Management System 06K159D XCB763 80347-20 Precision Xtra Advanced Diabetes Management System 06K166D XCB800 80347-20 Precision Xtra Advanced Diabetes Management System 06K166D XCB803 80347-20 Precision Xtra Advanced Diabetes Management System 06K191D XCB798 80347-20 Precision Xtra Advanced Diabetes Management System 06K207D XCB860 80347-20 Precision Xtra Advanced Diabetes Management System 06K207D XCB861 80347-20 Precision Xtra Advanced Diabetes Management System 07K071D XCB492 80347-20 Precision Xtra Advanced Diabetes Management System 07K074D XCB418 80347-20 Precision Xtra Advanced Diabetes Management System 07K074D XCB446 80347-20 Precision Xtra Advanced Diabetes Management System 07K099D XCB606 80347-20 Precision Xtra Advanced Diabetes Management System 07K099D XCB612 80347-20 Precision Xtra Advanced Diabetes Management System 07K107D XCB644 80347-20 Precision Xtra Advanced Diabetes Management System 07K134D XCB716 80347-20 Precision Xtra Advanced Diabetes Management System 07K150D XCB764 80347-20 Precision Xtra Advanced Diabetes Management System 07K150D XCB765 80347-20 Precision Xtra Advanced Diabetes Management System 07K155D XCB735 80347-20 Precision Xtra Advanced Diabetes Management System 07K155D XCB736 80347-20 Precision Xtra Advanced Diabetes Management System 07K155D XCB738 80347-20 Precision Xtra Advanced Diabetes Management System 07K157D XCB741 80347-20 Precision Xtra Advanced Diabetes Management System 07K201D XCB860 80347-20 Precision Xtra Advanced Diabetes Management System 07K201D XCB863 80347-20 Precision Xtra Advanced Diabetes Management System 07K201D XCB870 80347-20 Precision Xtra Advanced Diabetes Management System 08K073D XCB446 80347-20 Precision Xtra Advanced Diabetes Management System 08K073D XCB447 80347-20 Precision Xtra Advanced Diabetes Management System 08K088D XCB504 80347-20 Precision Xtra Advanced Diabetes Management System 08K088D XCB508 80347-20 Precision Xtra Advanced Diabetes Management System 08K088D XCB536 80347-20 Precision Xtra Advanced Diabetes Management System 08K108D XCB625 80347-20 Precision Xtra Advanced Diabetes Management System 08K134D XCB726 80347-20 Precision Xtra Advanced Diabetes Management System 08K134D XCB728 80347-20 Precision Xtra Advanced Diabetes Management System 08K144D XCB738 80347-20 Precision Xtra Advanced Diabetes Management System 08K144D XCB742 80347-20 Precision Xtra Advanced Diabetes Management System 08K144D XCB743 80347-20 Precision Xtra Advanced Diabetes Management System 08K144D XCB744 80347-20 Precision Xtra Advanced Diabetes Management System 08K150D XCB749 80347-20 Precision Xtra Advanced Diabetes Management System 08K150D XCB765 80347-20 Precision Xtra Advanced Diabetes Management System 09K088D XCB539 80347-20 Precision Xtra Advanced Diabetes Management System 09K108D XCB625 80347-20 Precision Xtra Advanced Diabetes Management System 09K108D XCB631 80347-20 Precision Xtra Advanced Diabetes Management System 09K108D XCB640 80347-20 Precision Xtra Advanced Diabetes Management System 10K078D XCB495 80347-20 Precision Xtra Advanced Diabetes Management System 10K078D XCB504 80347-20 Precision Xtra Advanced Diabetes Management System 10K150D XCB749 80347-20 Precision Xtra Advanced Diabetes Management System 10K150D XCB750 80347-20 Precision Xtra Advanced Diabetes Management System 11K170D XCB800 80374-50 Boots Blood Glucose Monitoring System 01K102C XCB605 80374-50 Boots Blood Glucose Monitoring System 01K102C XCB623 80374-50 Boots Blood Glucose Monitoring System 01K106C XCB595 80374-50 Boots Blood Glucose Monitoring System 01K106C XCB626 80374-50 Boots Blood Glucose Monitoring System 01K108C XCB591 80374-50 Boots Blood Glucose Monitoring System 02K106C XCB626 80374-50 Boots Blood Glucose Monitoring System 02K169C XCB784 80374-50 Boots Blood Glucose Monitoring System 03K108C XCB591 80374-50 Boots Blood Glucose Monitoring System 04K100C XCB569 80374-50 Boots Blood Glucose Monitoring System 04K102C XCB595 80374-50 Boots Blood Glucose Monitoring System 04K103C XCB605 80374-50 Boots Blood Glucose Monitoring System 05K108C XCB591 80374-50 Boots Blood Glucose Monitoring System 06K166C XCB784 80374-50 Boots Blood Glucose Monitoring System 07K101C XCB569 80374-50 Boots Blood Glucose Monitoring System 07K102C XCB623 80374-50 Boots Blood Glucose Monitoring System 07K107C XCB591 80374-50 Boots Blood Glucose Monitoring System 07K113C XCB651 80374-50 Boots Blood Glucose Monitoring System 07K113C XCB652 80374-50 Boots Blood Glucose Monitoring System 07K169C XCB784 80374-50 Boots Blood Glucose Monitoring System 07K169C XCB785 80374-50 Boots Blood Glucose Monitoring System 08K107C XCB591 80374-50 Boots Blood Glucose Monitoring System 09K106C XCB569 80374-50 Boots Blood Glucose Monitoring System 09K106C XCB626 80374-50 Boots Blood Glucose Monitoring System 09K107C XCB591 80374-50 Boots Blood Glucose Monitoring System 09K128C XCB591 80374-50 Boots Blood Glucose Monitoring System 09K128C XCB717 80374-50 Boots Blood Glucose Monitoring System 10K101C XCB605 80374-50 Boots Blood Glucose Monitoring System 10K101C XCB623 80374-50 Boots Blood Glucose Monitoring System 10K166C XCB784 80374-50 Boots Blood Glucose Monitoring System 10K166C XCB785 80374-50 Boots Blood Glucose Monitoring System 11K101C XCB595 80374-50 Boots Blood Glucose Monitoring System 11K101C XCB623 80374-50 Boots Blood Glucose Monitoring System 12K103C XCB595 80374-50 Boots Blood Glucose Monitoring System 12K103C XCB626 80374-50 Boots Blood Glucose Monitoring System 12K170C XCB784 80374-50 Boots Blood Glucose Monitoring System 12K170C XCB785 80374-50 Boots Blood Glucose Monitoring System 13K102C XCB605 80374-50 Boots Blood Glucose Monitoring System 13K114C XCB651 80374-50 Boots Blood Glucose Monitoring System 13K114C XCB652 80374-50 Boots Blood Glucose Monitoring System 17K127C XCB591 80374-50 Boots Blood Glucose Monitoring System 17K127C XCB717 80374-50 Boots Blood Glucose Monitoring System 18K165C XCB784 80374-50 Boots Blood Glucose Monitoring System 18K180C XCB784 98814-20 Precision Xtra Advanced Diabetes Management System 01K198D XCB802 98814-20 Precision Xtra Advanced Diabetes Management System 02K072D XCB493 98814-20 Precision Xtra Advanced Diabetes Management System 02K072D XCB494 98814-20 Precision Xtra Advanced Diabetes Management System 02K100D XCB597 98814-20 Precision Xtra Advanced Diabetes Management System 02K113D XCB650 98814-20 Precision Xtra Advanced Diabetes Management System 02K113D XCB654 98814-20 Precision Xtra Advanced Diabetes Management System 03K158D XCB742 98814-20 Precision Xtra Advanced Diabetes Management System 03K158D XCB762 98814-20 Precision Xtra Advanced Diabetes Management System 04K114D XCB650 98814-20 Precision Xtra Advanced Diabetes Management System 04K114D XCB654 98814-20 Precision Xtra Advanced Diabetes Management System 04K170D XCB799 98814-20 Precision Xtra Advanced Diabetes Management System 04K170D XCB801 98814-20 Precision Xtra Advanced Diabetes Management System 05K110D XCB662 98814-20 Precision Xtra Advanced Diabetes Management System 06K134D XCB671 98814-20 Precision Xtra Advanced Diabetes Management System 06K134D XCB713 98814-20 Precision Xtra Advanced Diabetes Management System 06K157D XCB742 98814-20 Precision Xtra Advanced Diabetes Management System 06K157D XCB762 98814-20 Precision Xtra Advanced Diabetes Management System 06K169D XCB799 98814-20 Precision Xtra Advanced Diabetes Management System 06K169D XCB801 98814-20 Precision Xtra Advanced Diabetes Management System 07K144D XCB723 98814-20 Precision Xtra Advanced Diabetes Management System 07K144D XCB737 98814-20 Precision Xtra Advanced Diabetes Management System 08K071D XCB493 98814-20 Precision Xtra Advanced Diabetes Management System 08K071D XCB494 98814-20 Precision Xtra Advanced Diabetes Management System 08K099D XCB597 98814-20 Precision Xtra Advanced Diabetes Management System 09K197D XCB802 98814-20 Precision Xtra Advanced Diabetes Management System 11K109D XCB662 98814-21 Precision Xtra Advanced Diabetes Management System 02K134D XCB713 98814-21 Precision Xtra Advanced Diabetes Management System 02K183D XCB798 98814-21 Precision Xtra Advanced Diabetes Management System 02K183D XCB832 98814-21 Precision Xtra Advanced Diabetes Management System 03K072D XCB492 98814-21 Precision Xtra Advanced Diabetes Management System 03K072D XCB493 98814-21 Precision Xtra Advanced Diabetes Management System 03K181D XCB650 98814-21 Precision Xtra Advanced Diabetes Management System 03K181D XCB654 98814-21 Precision Xtra Advanced Diabetes Management System 03K181D XCB802 98814-21 Precision Xtra Advanced Diabetes Management System 03K181D XCB804 98814-21 Precision Xtra Advanced Diabetes Management System 04K100D XCB587 98814-21 Precision Xtra Advanced Diabetes Management System 04K100D XCB597 98814-21 Precision Xtra Advanced Diabetes Management System 05K157D XCB762 98814-21 Precision Xtra Advanced Diabetes Management System 05K157D XCB763 98814-21 Precision Xtra Advanced Diabetes Management System 06K110D XCB654 98814-21 Precision Xtra Advanced Diabetes Management System 07K156D XCB762 98814-21 Precision Xtra Advanced Diabetes Management System 07K156D XCB763 98814-21 Precision Xtra Advanced Diabetes Management System 08K180D XCB650 98814-21 Precision Xtra Advanced Diabetes Management System 08K180D XCB654 98814-21 Precision Xtra Advanced Diabetes Management System 08K180D XCB802 98814-21 Precision Xtra Advanced Diabetes Management System 08K180D XCB804 98814-21 Precision Xtra Advanced Diabetes Management System 09K071D XCB492 98814-21 Precision Xtra Advanced Diabetes Management System 09K071D XCB493 98814-21 Precision Xtra Advanced Diabetes Management System 09K099D XCB587 98814-21 Precision Xtra Advanced Diabetes Management System 09K099D XCB597 98814-22 Precision Xtra Advanced Diabetes Management System 01K136D XCB668 98814-22 Precision Xtra Advanced Diabetes Management System 01K136D XCB720 98814-22 Precision Xtra Advanced Diabetes Management System 01K144D XCB714 98814-22 Precision Xtra Advanced Diabetes Management System 01K144D XCB715 98814-22 Precision Xtra Advanced Diabetes Management System 01K144D XCB723 98814-22 Precision Xtra Advanced Diabetes Management System 02K073D XCB493 98814-22 Precision Xtra Advanced Diabetes Management System 02K158D XCB750 98814-22 Precision Xtra Advanced Diabetes Management System 02K158D XCB751 98814-22 Precision Xtra Advanced Diabetes Management System 02K158D XCB753 98814-22 Precision Xtra Advanced Diabetes Management System 03K135D XCB716 98814-22 Precision Xtra Advanced Diabetes Management System 03K135D XCB720 98814-22 Precision Xtra Advanced Diabetes Management System 03K145D XCB714 98814-22 Precision Xtra Advanced Diabetes Management System 03K145D XCB715 98814-22 Precision Xtra Advanced Diabetes Management System 03K145D XCB723 98814-22 Precision Xtra Advanced Diabetes Management System 03K150D XCB714 98814-22 Precision Xtra Advanced Diabetes Management System 03K150D XCB715 98814-22 Precision Xtra Advanced Diabetes Management System 03K150D XCB723 98814-22 Precision Xtra Advanced Diabetes Management System 03K152D XCB750 98814-22 Precision Xtra Advanced Diabetes Management System 03K152D XCB751 98814-22 Precision Xtra Advanced Diabetes Management System 03K152D XCB753 98814-22 Precision Xtra Advanced Diabetes Management System 03K157D XCB750 98814-22 Precision Xtra Advanced Diabetes Management System 03K157D XCB751 98814-22 Precision Xtra Advanced Diabetes Management System 03K157D XCB753 98814-22 Precision Xtra Advanced Diabetes Management System 04K092D XCB557 98814-22 Precision Xtra Advanced Diabetes Management System 04K092D XCB567 98814-22 Precision Xtra Advanced Diabetes Management System 04K108D XCB622 98814-22 Precision Xtra Advanced Diabetes Management System 04K108D XCB635 98814-22 Precision Xtra Advanced Diabetes Management System 04K108D XCB638 98814-22 Precision Xtra Advanced Diabetes Management System 04K108D XCB639 98814-22 Precision Xtra Advanced Diabetes Management System 04K108D XCB644 98814-22 Precision Xtra Advanced Diabetes Management System 04K135D XCB668 98814-22 Precision Xtra Advanced Diabetes Management System 04K135D XCB720 98814-22 Precision Xtra Advanced Diabetes Management System 04K149D XCB714 98814-22 Precision Xtra Advanced Diabetes Management System 04K149D XCB715 98814-22 Precision Xtra Advanced Diabetes Management System 04K149D XCB723 98814-22 Precision Xtra Advanced Diabetes Management System 04K156D XCB750 98814-22 Precision Xtra Advanced Diabetes Management System 04K156D XCB751 98814-22 Precision Xtra Advanced Diabetes Management System 04K156D XCB753 98814-22 Precision Xtra Advanced Diabetes Management System 04K199D XCB798 98814-22 Precision Xtra Advanced Diabetes Management System 04K199D XCB802 98814-22 Precision Xtra Advanced Diabetes Management System 04K199D XCB832 98814-22 Precision Xtra Advanced Diabetes Management System 04K199D XCB844 98814-22 Precision Xtra Advanced Diabetes Management System 04K199D XCB845 98814-22 Precision Xtra Advanced Diabetes Management System 05K099D XCB597 98814-22 Precision Xtra Advanced Diabetes Management System 05K107D XCB622 98814-22 Precision Xtra Advanced Diabetes Management System 05K107D XCB635 98814-22 Precision Xtra Advanced Diabetes Management System 05K107D XCB638 98814-22 Precision Xtra Advanced Diabetes Management System 05K107D XCB639 98814-22 Precision Xtra Advanced Diabetes Management System 05K107D XCB644 98814-22 Precision Xtra Advanced Diabetes Management System 05K109D XCB622 98814-22 Precision Xtra Advanced Diabetes Management System 05K109D XCB635 98814-22 Precision Xtra Advanced Diabetes Management System 05K109D XCB638 98814-22 Precision Xtra Advanced Diabetes Management System 05K109D XCB639 98814-22 Precision Xtra Advanced Diabetes Management System 05K109D XCB644 98814-22 Precision Xtra Advanced Diabetes Management System 05K155D XCB750 98814-22 Precision Xtra Advanced Diabetes Management System 05K155D XCB751 98814-22 Precision Xtra Advanced Diabetes Management System 05K155D XCB753 98814-22 Precision Xtra Advanced Diabetes Management System 06K072D XCB493 98814-22 Precision Xtra Advanced Diabetes Management System 06K198D XCB798 98814-22 Precision Xtra Advanced Diabetes Management System 06K198D XCB802 98814-22 Precision Xtra Advanced Diabetes Management System 06K198D XCB832 98814-22 Precision Xtra Advanced Diabetes Management System 06K198D XCB844 98814-22 Precision Xtra Advanced Diabetes Management System 06K198D XCB845 98814-22 Precision Xtra Advanced Diabetes Management System 07K135D XCB668 98814-22 Precision Xtra Advanced Diabetes Management System 08K106D XCB622 98814-22 Precision Xtra Advanced Diabetes Management System 08K106D XCB635 98814-22 Precision Xtra Advanced Diabetes Management System 08K106D XCB638 98814-22 Precision Xtra Advanced Diabetes Management System 08K106D XCB639 98814-22 Precision Xtra Advanced Diabetes Management System 08K106D XCB644 98814-22 Precision Xtra Advanced Diabetes Management System 08K136D XCB668 99132-50 Optium Blood Glucose Monitoring System 01K068D XCB477 99132-50 Optium Blood Glucose Monitoring System 01K068D XCB478 99132-50 Optium Blood Glucose Monitoring System 01K068D XCB479 99132-50 Optium Blood Glucose Monitoring System 01K071D XCB477 99132-50 Optium Blood Glucose Monitoring System 01K071D XCB478 99132-50 Optium Blood Glucose Monitoring System 01K071D XCB479 99132-50 Optium Blood Glucose Monitoring System 01K095D XCB533 99132-50 Optium Blood Glucose Monitoring System 01K095D XCB534 99132-50 Optium Blood Glucose Monitoring System 01K113D XCB510 99132-50 Optium Blood Glucose Monitoring System 01K113D XCB541 99132-50 Optium Blood Glucose Monitoring System 01K113D XCB543 99132-50 Optium Blood Glucose Monitoring System 01K113D XCB554 99132-50 Optium Blood Glucose Monitoring System 01K114D XCB510 99132-50 Optium Blood Glucose Monitoring System 01K114D XCB541 99132-50 Optium Blood Glucose Monitoring System 01K114D XCB543 99132-50 Optium Blood Glucose Monitoring System 01K114D XCB554 99132-50 Optium Blood Glucose Monitoring System 01K156D XCB666 99132-50 Optium Blood Glucose Monitoring System 01K156D XCB667 99132-50 Optium Blood Glucose Monitoring System 01K156D XCB722 99132-50 Optium Blood Glucose Monitoring System 01K157D XCB666 99132-50 Optium Blood Glucose Monitoring System 01K157D XCB667 99132-50 Optium Blood Glucose Monitoring System 01K157D XCB722 99132-50 Optium Blood Glucose Monitoring System 01K187D XCB721 99132-50 Optium Blood Glucose Monitoring System 01K187D XCB724 99132-50 Optium Blood Glucose Monitoring System 02K065D XCB361 99132-50 Optium Blood Glucose Monitoring System 02K065D XCB435 99132-50 Optium Blood Glucose Monitoring System 02K065D XCB436 99132-50 Optium Blood Glucose Monitoring System 02K065D XCB438 99132-50 Optium Blood Glucose Monitoring System 02K086D XCB510 99132-50 Optium Blood Glucose Monitoring System 02K086D XCB513 99132-50 Optium Blood Glucose Monitoring System 02K096D XCB514 99132-50 Optium Blood Glucose Monitoring System 02K096D XCB533 99132-50 Optium Blood Glucose Monitoring System 02K099D XCB514 99132-50 Optium Blood Glucose Monitoring System 02K099D XCB543 99132-50 Optium Blood Glucose Monitoring System 02K099D XCB544 99132-50 Optium Blood Glucose Monitoring System 02K099D XCB546 99132-50 Optium Blood Glucose Monitoring System 02K124D XCB656 99132-50 Optium Blood Glucose Monitoring System 02K124D XCB660 99132-50 Optium Blood Glucose Monitoring System 02K124D XCB664 99132-50 Optium Blood Glucose Monitoring System 02K124D XCB666 99132-50 Optium Blood Glucose Monitoring System 03K066D XCB361 99132-50 Optium Blood Glucose Monitoring System 03K066D XCB435 99132-50 Optium Blood Glucose Monitoring System 03K066D XCB436 99132-50 Optium Blood Glucose Monitoring System 03K066D XCB438 99132-50 Optium Blood Glucose Monitoring System 03K074D XCB406 99132-50 Optium Blood Glucose Monitoring System 03K074D XCB407 99132-50 Optium Blood Glucose Monitoring System 03K075D XCB405 99132-50 Optium Blood Glucose Monitoring System 03K075D XCB406 99132-50 Optium Blood Glucose Monitoring System 03K075D XCB408 99132-50 Optium Blood Glucose Monitoring System 03K095D XCB534 99132-50 Optium Blood Glucose Monitoring System 03K095D XCB542 99132-50 Optium Blood Glucose Monitoring System 03K095D XCB546 99132-50 Optium Blood Glucose Monitoring System 03K127D XCB548 99132-50 Optium Blood Glucose Monitoring System 03K127D XCB655 99132-50 Optium Blood Glucose Monitoring System 03K127D XCB656 99132-50 Optium Blood Glucose Monitoring System 03K127D XCB657 99132-50 Optium Blood Glucose Monitoring System 04K074D XCB405 99132-50 Optium Blood Glucose Monitoring System 04K074D XCB406 99132-50 Optium Blood Glucose Monitoring System 04K074D XCB408 99132-50 Optium Blood Glucose Monitoring System 04K075D XCB406 99132-50 Optium Blood Glucose Monitoring System 04K075D XCB407 99132-50 Optium Blood Glucose Monitoring System 05K123D XCB656 99132-50 Optium Blood Glucose Monitoring System 05K123D XCB660 99132-50 Optium Blood Glucose Monitoring System 05K123D XCB664 99132-50 Optium Blood Glucose Monitoring System 05K123D XCB666 99132-50 Optium Blood Glucose Monitoring System 06K124D XCB548 99132-50 Optium Blood Glucose Monitoring System 06K124D XCB655 99132-50 Optium Blood Glucose Monitoring System 06K124D XCB656 99132-50 Optium Blood Glucose Monitoring System 06K124D XCB657 99132-50 Optium Blood Glucose Monitoring System 07K067D XCB361 99132-50 Optium Blood Glucose Monitoring System 07K067D XCB435 99132-50 Optium Blood Glucose Monitoring System 07K067D XCB436 99132-50 Optium Blood Glucose Monitoring System 07K067D XCB438 99132-50 Optium Blood Glucose Monitoring System 07K085D XCB479 99132-50 Optium Blood Glucose Monitoring System 07K085D XCB498 99132-50 Optium Blood Glucose Monitoring System 07K085D XCB513 99132-50 Optium Blood Glucose Monitoring System 07K096D XCB514 99132-50 Optium Blood Glucose Monitoring System 07K096D XCB543 99132-50 Optium Blood Glucose Monitoring System 07K096D XCB544 99132-50 Optium Blood Glucose Monitoring System 07K096D XCB546 99132-50 Optium Blood Glucose Monitoring System 08K082D XCB479 99132-50 Optium Blood Glucose Monitoring System 08K082D XCB498 99132-50 Optium Blood Glucose Monitoring System 08K082D XCB513 99132-50 Optium Blood Glucose Monitoring System 08K095D XCB514 99132-50 Optium Blood Glucose Monitoring System 08K095D XCB533 99132-50 Optium Blood Glucose Monitoring System 08K123D XCB548 99132-50 Optium Blood Glucose Monitoring System 08K123D XCB655 99132-50 Optium Blood Glucose Monitoring System 08K123D XCB656 99132-50 Optium Blood Glucose Monitoring System 08K123D XCB657 99132-50 Optium Blood Glucose Monitoring System 09K067D XCB477 99132-50 Optium Blood Glucose Monitoring System 09K067D XCB478 99132-50 Optium Blood Glucose Monitoring System 09K067D XCB479 99132-50 Optium Blood Glucose Monitoring System 10K073D XCB406 99132-50 Optium Blood Glucose Monitoring System 10K073D XCB407 99132-50 Optium Blood Glucose Monitoring System 10K085D XCB510 99132-50 Optium Blood Glucose Monitoring System 10K085D XCB513 99132-50 Optium Blood Glucose Monitoring System 10K094D XCB534 99132-50 Optium Blood Glucose Monitoring System 10K094D XCB542 99132-50 Optium Blood Glucose Monitoring System 10K094D XCB546 99132-50 Optium Blood Glucose Monitoring System 12K094D XCB533 99132-50 Optium Blood Glucose Monitoring System 12K094D XCB534 99682-50 Liberty Optium Blood Glucose Monitoring System 01K066D XCB433 99682-50 Liberty Optium Blood Glucose Monitoring System 01K066D XCB434 99682-50 Liberty Optium Blood Glucose Monitoring System 01K066D XCB485 99682-50 Liberty Optium Blood Glucose Monitoring System 02K108D XCB620 99682-50 Liberty Optium Blood Glucose Monitoring System 02K108D XCB621 99682-50 Liberty Optium Blood Glucose Monitoring System 02K108D XCB633 99682-50 Liberty Optium Blood Glucose Monitoring System 02K108D XCB634 99682-50 Liberty Optium Blood Glucose Monitoring System 02K108D XCB636 99682-50 Liberty Optium Blood Glucose Monitoring System 02K121D XCB680 99682-50 Liberty Optium Blood Glucose Monitoring System 02K121D XCB682 99682-50 Liberty Optium Blood Glucose Monitoring System 02K121D XCB683 99682-50 Liberty Optium Blood Glucose Monitoring System 02K123D XCB682 99682-50 Liberty Optium Blood Glucose Monitoring System 02K123D XCB707 99682-50 Liberty Optium Blood Glucose Monitoring System 03K046D XCB333 99682-50 Liberty Optium Blood Glucose Monitoring System 03K046D XCB351 99682-50 Liberty Optium Blood Glucose Monitoring System 03K046D XCB359 99682-50 Liberty Optium Blood Glucose Monitoring System 03K046D XCB360 99682-50 Liberty Optium Blood Glucose Monitoring System 03K058D XCB352 99682-50 Liberty Optium Blood Glucose Monitoring System 03K058D XCB360 99682-50 Liberty Optium Blood Glucose Monitoring System 03K065D XCB433 99682-50 Liberty Optium Blood Glucose Monitoring System 03K065D XCB434 99682-50 Liberty Optium Blood Glucose Monitoring System 03K065D XCB485 99682-50 Liberty Optium Blood Glucose Monitoring System 03K067D XCB433 99682-50 Liberty Optium Blood Glucose Monitoring System 03K067D XCB434 99682-50 Liberty Optium Blood Glucose Monitoring System 03K067D XCB485 99682-50 Liberty Optium Blood Glucose Monitoring System 04K047D XCB333 99682-50 Liberty Optium Blood Glucose Monitoring System 04K047D XCB351 99682-50 Liberty Optium Blood Glucose Monitoring System 04K047D XCB359 99682-50 Liberty Optium Blood Glucose Monitoring System 04K047D XCB360 99682-50 Liberty Optium Blood Glucose Monitoring System 04K107D XCB620 99682-50 Liberty Optium Blood Glucose Monitoring System 04K107D XCB621 99682-50 Liberty Optium Blood Glucose Monitoring System 04K107D XCB633 99682-50 Liberty Optium Blood Glucose Monitoring System 04K107D XCB634 99682-50 Liberty Optium Blood Glucose Monitoring System 04K107D XCB636 99682-50 Liberty Optium Blood Glucose Monitoring System 04K120D XCB680 99682-50 Liberty Optium Blood Glucose Monitoring System 05K106D XCB620 99682-50 Liberty Optium Blood Glucose Monitoring System 05K106D XCB621 99682-50 Liberty Optium Blood Glucose Monitoring System 05K106D XCB633 99682-50 Liberty Optium Blood Glucose Monitoring System 05K106D XCB634 99682-50 Liberty Optium Blood Glucose Monitoring System 05K106D XCB636 99682-50 Liberty Optium Blood Glucose Monitoring System 06K080D XCB473 99682-50 Liberty Optium Blood Glucose Monitoring System 06K080D XCB485 99682-50 Liberty Optium Blood Glucose Monitoring System 06K082D XCB473 99682-50 Liberty Optium Blood Glucose Monitoring System 06K089D XCB473 99682-50 Liberty Optium Blood Glucose Monitoring System 06K089D XCB474 99682-50 Liberty Optium Blood Glucose Monitoring System 06K100D XCB616 99682-50 Liberty Optium Blood Glucose Monitoring System 06K107D XCB474 99682-50 Liberty Optium Blood Glucose Monitoring System 06K107D XCB630 99682-50 Liberty Optium Blood Glucose Monitoring System 06K107D XCB633 99682-50 Liberty Optium Blood Glucose Monitoring System 06K122D XCB706 99682-50 Liberty Optium Blood Glucose Monitoring System 06K122D XCB707 99682-50 Liberty Optium Blood Glucose Monitoring System 07K088D XCB473 99682-50 Liberty Optium Blood Glucose Monitoring System 07K088D XCB474 99682-50 Liberty Optium Blood Glucose Monitoring System 07K106D XCB474 99682-50 Liberty Optium Blood Glucose Monitoring System 07K106D XCB630 99682-50 Liberty Optium Blood Glucose Monitoring System 07K106D XCB633 99682-50 Liberty Optium Blood Glucose Monitoring System 07K121D XCB682 99682-50 Liberty Optium Blood Glucose Monitoring System 07K121D XCB707 99682-50 Liberty Optium Blood Glucose Monitoring System 07K122D XCB682 99682-50 Liberty Optium Blood Glucose Monitoring System 07K122D XCB707 99682-50 Liberty Optium Blood Glucose Monitoring System 08K045D XCB333 99682-50 Liberty Optium Blood Glucose Monitoring System 08K045D XCB351 99682-50 Liberty Optium Blood Glucose Monitoring System 08K045D XCB359 99682-50 Liberty Optium Blood Glucose Monitoring System 08K045D XCB360 99682-50 Liberty Optium Blood Glucose Monitoring System 08K120D XCB680 99682-50 Liberty Optium Blood Glucose Monitoring System 08K120D XCB681 99682-50 Liberty Optium Blood Glucose Monitoring System 09K120D XCB680 99682-50 Liberty Optium Blood Glucose Monitoring System 09K120D XCB682 99682-50 Liberty Optium Blood Glucose Monitoring System 09K120D XCB683 99682-50 Liberty Optium Blood Glucose Monitoring System 10K057D XCB352 99682-50 Liberty Optium Blood Glucose Monitoring System 10K057D XCB360 99730-20 Precision Xtra Advanced Diabetes Management System 01K099D XCB587 99730-20 Precision Xtra Advanced Diabetes Management System 01K099D XCB606 99730-20 Precision Xtra Advanced Diabetes Management System 01K099D XCB607 99730-20 Precision Xtra Advanced Diabetes Management System 01K120D XCB669 99730-20 Precision Xtra Advanced Diabetes Management System 01K151D XCB732 99730-20 Precision Xtra Advanced Diabetes Management System 01K151D XCB762 99730-20 Precision Xtra Advanced Diabetes Management System 01K151D XCB764 99730-20 Precision Xtra Advanced Diabetes Management System 01K165D XCB800 99730-20 Precision Xtra Advanced Diabetes Management System 01K165D XCB801 99730-20 Precision Xtra Advanced Diabetes Management System 03K176D XCB800 99730-20 Precision Xtra Advanced Diabetes Management System 03K176D XCB804 99730-20 Precision Xtra Advanced Diabetes Management System 04K073D XCB447 99730-20 Precision Xtra Advanced Diabetes Management System 05K085D XCB508 99730-20 Precision Xtra Advanced Diabetes Management System 06K108D XCB644 99730-20 Precision Xtra Advanced Diabetes Management System 06K108D XCB653 99730-20 Precision Xtra Advanced Diabetes Management System 06K123D XCB669 99730-20 Precision Xtra Advanced Diabetes Management System 07K052D XCB244 99730-20 Precision Xtra Advanced Diabetes Management System 07K052D XCB381 99730-20 Precision Xtra Advanced Diabetes Management System 07K117D XCB669 99730-20 Precision Xtra Advanced Diabetes Management System 07K207D XCB861 99730-20 Precision Xtra Advanced Diabetes Management System 07K207D XCB863 99730-20 Precision Xtra Advanced Diabetes Management System 07K207D XCB864 99730-20 Precision Xtra Advanced Diabetes Management System 07K207D XCB868 99730-20 Precision Xtra Advanced Diabetes Management System 07K207D XCB869 99730-20 Precision Xtra Advanced Diabetes Management System 09K164D XCB800 99730-20 Precision Xtra Advanced Diabetes Management System 09K164D XCB801 99825-20 Precision Xtra Advanced Diabetes Management System 01K058D XCB214 99825-20 Precision Xtra Advanced Diabetes Management System 01K058D XCB427 99825-20 Precision Xtra Advanced Diabetes Management System 01K212D XCB486 99825-20 Precision Xtra Advanced Diabetes Management System 01K212D XCB487 99825-20 Precision Xtra Advanced Diabetes Management System 01K212D XCB515 99825-20 Precision Xtra Advanced Diabetes Management System 03K060D XCB427 99825-20 Precision Xtra Advanced Diabetes Management System 03K060D XCB439 99825-20 Precision Xtra Advanced Diabetes Management System 04K211D XCB486 99825-20 Precision Xtra Advanced Diabetes Management System 04K211D XCB487 99825-20 Precision Xtra Advanced Diabetes Management System 04K211D XCB515 99825-20 Precision Xtra Advanced Diabetes Management System 05K059D XCB427 99825-20 Precision Xtra Advanced Diabetes Management System 05K059D XCB439 99825-20 Precision Xtra Advanced Diabetes Management System 06K067D XCB439 99825-20 Precision Xtra Advanced Diabetes Management System 06K067D XCB487 99825-20 Precision Xtra Advanced Diabetes Management System 07K057D XCB214 99825-20 Precision Xtra Advanced Diabetes Management System 07K057D XCB427 99837-20 Precision Xtra Advanced Diabetes Management System 01K080D XCB416 99837-20 Precision Xtra Advanced Diabetes Management System 01K080D XCB447 99837-20 Precision Xtra Advanced Diabetes Management System 01K080D XCB492 99837-20 Precision Xtra Advanced Diabetes Management System 01K080D XCB505 99837-20 Precision Xtra Advanced Diabetes Management System 01K107D XCB622 99837-20 Precision Xtra Advanced Diabetes Management System 01K107D XCB627 99837-20 Precision Xtra Advanced Diabetes Management System 01K107D XCB632 99837-20 Precision Xtra Advanced Diabetes Management System 01K200D XCB844 99837-20 Precision Xtra Advanced Diabetes Management System 01K200D XCB866 99837-20 Precision Xtra Advanced Diabetes Management System 02K089D XCB539 99837-20 Precision Xtra Advanced Diabetes Management System 02K089D XCB566 99837-20 Precision Xtra Advanced Diabetes Management System 03K073D XCB416 99837-20 Precision Xtra Advanced Diabetes Management System 03K073D XCB418 99837-20 Precision Xtra Advanced Diabetes Management System 03K100D XCB606 99837-20 Precision Xtra Advanced Diabetes Management System 03K136D XCB668 99837-20 Precision Xtra Advanced Diabetes Management System 03K136D XCB713 99837-20 Precision Xtra Advanced Diabetes Management System 03K143D XCB714 99837-20 Precision Xtra Advanced Diabetes Management System 03K143D XCB729 99837-20 Precision Xtra Advanced Diabetes Management System 04K081D XCB416 99837-20 Precision Xtra Advanced Diabetes Management System 04K081D XCB447 99837-20 Precision Xtra Advanced Diabetes Management System 04K081D XCB492 99837-20 Precision Xtra Advanced Diabetes Management System 04K081D XCB505 99837-20 Precision Xtra Advanced Diabetes Management System 04K113D XCB650 99837-20 Precision Xtra Advanced Diabetes Management System 04K113D XCB654 99837-20 Precision Xtra Advanced Diabetes Management System 05K135D XCB668 99837-20 Precision Xtra Advanced Diabetes Management System 05K135D XCB713 99837-20 Precision Xtra Advanced Diabetes Management System 05K142D XCB714 99837-20 Precision Xtra Advanced Diabetes Management System 05K142D XCB729 99837-20 Precision Xtra Advanced Diabetes Management System 05K198D XCB844 99837-20 Precision Xtra Advanced Diabetes Management System 05K198D XCB866 99837-20 Precision Xtra Advanced Diabetes Management System 06K099D XCB606 99837-20 Precision Xtra Advanced Diabetes Management System 06K106D XCB622 99837-20 Precision Xtra Advanced Diabetes Management System 06K106D XCB627 99837-20 Precision Xtra Advanced Diabetes Management System 06K106D XCB632 99837-20 Precision Xtra Advanced Diabetes Management System 06K141D XCB714 99837-20 Precision Xtra Advanced Diabetes Management System 06K141D XCB729 99837-20 Precision Xtra Advanced Diabetes Management System 06K177D XCB799 99837-20 Precision Xtra Advanced Diabetes Management System 06K177D XCB804 99837-20 Precision Xtra Advanced Diabetes Management System 06K199D XCB844 99837-20 Precision Xtra Advanced Diabetes Management System 06K199D XCB866 99837-20 Precision Xtra Advanced Diabetes Management System 07K089D XCB557 99837-20 Precision Xtra Advanced Diabetes Management System 07K089D XCB566 99837-20 Precision Xtra Advanced Diabetes Management System 08K110D XCB650 99837-20 Precision Xtra Advanced Diabetes Management System 08K110D XCB654 99837-20 Precision Xtra Advanced Diabetes Management System 09K072D XCB416 99837-20 Precision Xtra Advanced Diabetes Management System 09K072D XCB418 99837-20 Precision Xtra Advanced Diabetes Management System 09K176D XCB799 99837-20 Precision Xtra Advanced Diabetes Management System 09K176D XCB804 99837-20 Precision Xtra Advanced Diabetes Management System 10K088D XCB539 99837-20 Precision Xtra Advanced Diabetes Management System 10K088D XCB566 99837-20 Precision Xtra Advanced Diabetes Management System 10K109D XCB662 99847-50 Kroger Blood Glucose Monitoring System 04K102D XCB611 99847-50 Kroger Blood Glucose Monitoring System 04K102D XCB617 99847-50 Kroger Blood Glucose Monitoring System 05K101D XCB611 99847-50 Kroger Blood Glucose Monitoring System 05K101D XCB617 99847-50 Kroger Blood Glucose Monitoring System 07K176D XCB755 70077-01 Precision Xceed Diabetes Monitoring System GRW78354/124 XCB424 70077-01 Precision Xceed Diabetes Monitoring System 668463/186 XCB886 70077-01 Precision Xceed Diabetes Monitoring System 668463/187 XCB887 70077-01 Precision Xceed Diabetes Monitoring System 668463/188 XCB888 70077-01 Precision Xceed Diabetes Monitoring System 368463/168 XCB568 70077-01 Precision Xceed Diabetes Monitoring System 368463/170 XCB570 70077-01 Precision Xceed Diabetes Monitoring System 268463/123 XCB423 70077-01 Precision Xceed Diabetes Monitoring System 268463/124 XCB424
Other devices under the same FDA submission
1 device cleared under K033845, K040814 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Precision Xtra 71502Abbott Diabetes Care Inc · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER12 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page