HomeChoice PRO Automated Peritoneal Dialysis Systems recallZ-0947-2008 · Baxter Healthcare Renal Div · initiated Nov 20, 2007
Terminated All recalls posted in March 2008
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Renal Div or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Unrecorded Infusion; If the system is powered down or a power failure occurs during a fill cycle, the HomeChoice system may not record the last 1-4 pump strokes infused into the patient when the power is restored. This situation may create a potential unrecorded delivery of approximately 15-60 mL of fluid being infused into the patient.
FDA root cause category: Device Design
What the company is doing
Urgent Product Recall letters dated 11/20/07 were sent via first class mail to all of the affected customers, informing them of the potential for overfill if the system is powered down or a power failure occurs during a fill cycle. A Baxter representative will contact the customer to schedule the software update which corects this anomaly. The accounts were requested to complete and return the enclosed Customer Reply Form, indicating that they have disseminated the information to their staff and home patients. Any questions were directed to Baxter Global Technical Services at 1-800-553-6898, prompt 1. Baxter sent a follow-up letter dated 3/17/08 to the consignees, accompanied by a copy of the 11/20/07 letter. The letter reiterated the information in the 11/20/07 letter, emphasizing the serious consequences of an overfill for neonates and young pediatric patients, and providing information describing the symptoms of excess fluid in the abdomen and steps to be taken if the user suspects overfill of a neonate or young pediatric patient. Baxter requested that Baxter be contacted if an overfill should occur so Baxter can perform a complete investigation into the event. The accounts were requested to complete and return the enclosed Customer Reply Form, indicating that they have disseminated the information to their staff and home patients. Any questions were directed to Baxter Global Technical Services at 1-800-553-6898, prompt 1.
Recall details
- Recall number
- Z-0947-2008
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Nov 20, 2007
- Posted by the FDA
- Mar 7, 2008
- Recalling firm
- Baxter Healthcare Renal Div
- Quantity
- 32,229 units
- Product code
- FKX
- FDA submissions
- K012988
- Distribution
- Worldwide; USA, Australia, New Zealand, China, Hong Kong, the Philippines, Thailand, Asia, Canada, Algeria, Austria, Bahrain, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Iraq, Ireland, Israel, Italy, Jordan, Kuwait, Latvia, Lebanon, Lithuania, Malta, Morocco, The Netherlands, Norway, Oman, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Senegal, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, United Arab Emirates, United Kingdom, Ukraine, Latin America and Japan.
Product description
HomeChoice PRO Automated Peritoneal Dialysis Systems; Catalog Numbers; 5C8310, 5C8310R, R5C8320, R5C8320R, T5C8300, and T5C8300R, Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A
Code information (models, lots, serial numbers)
All serial numbers released prior to 08/23/07
Other devices under the same FDA submission
1 device cleared under K012988 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- LOW RECIRCULATION VOLUME APD SET 5C8302CBaxter International Inc. · SYSTEM, PERITONEAL, AUTOMATIC DELIVERY9 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 45861).
- Z-0946-2008 · TerminatedBaxter HomeChoice Automated Peritoneal Dialysis Systems; Catalog Numbers: 5C4471, 5C4471R, 5C4474, 5C4474R, T5C4441, and T5C4441R, Manufactured for Baxter Healthcare Corporation, D…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page