Nexframe Stereotactic System Kits, Model Number(s):… recallZ-0849-2008 · Medtronic Image Guided Neurologics, Inc. · initiated Oct 4, 2007
Terminated All recalls posted in February 2008
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Image Guided Neurologics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Sterilty (package integrity) Compromised: Some failures were for damage to the outer pouch, while another set of failures were for the seals on this pouch. The seal between the inner tray and lid has not been compromised and the contents remain sterile, however, the sterility of the outer surfaces of the inner tray and lid cannot be assured.
FDA root cause category: Package design/selection
What the company is doing
On 10/4/2007, consignees were notified by Medtronic field personnel, either by personal visits or telephone. All wholesalers of this product are OUS. Medtronic OUS Regulatory personnel communicated this issue directly to the wholesalers, who provided the letters to their customers. Medtronic field personnel are responsible for completing and returning reply cards confirming all notifications (100%).
Recall details
- Recall number
- Z-0849-2008
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 4, 2007
- Posted by the FDA
- Feb 7, 2008
- Recalling firm
- Medtronic Image Guided Neurologics, Inc.
- Quantity
- 385
- Product code
- HAW
- FDA submissions
- K012366, K022100, K040334, K043091
- Distribution
- Worldwide, USA, Taiwan, Thailand, Japan, Spain, Sweden and The Netherlands.
Product description
Nexframe Stereotactic System Kits, Model Number(s): DB-1021-MR, DB-1031, DB-1040-BL, DB-1040-ST, DB-1041, DB-1041-BL, DB-1041-ST, DB-1042, DB-1042-BL, DB-1042-ST, DB-1043, DB-1043-BL, DB-1043-ST, DB-2031, DB-2040-BL, DB-2040-ST, DB-2041, DB-2041-BL, DB-2041-ST, DB-2042, DB-2042-BL, DB-2042-ST, DB-2043, DB-2043-BL, DB-2043-ST,, Medtronic, Inc, Minneapolis, Minnesota
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Lot #s: 070380, 061248, 070175, 070375, 070621, 070825, 070634, 6510106A, 046510106A, 46515606, 046520506A, 046521606A, 046523406A, 61262, 70160, 070326, 070171, 60215, 070719, 070378, 070169, 070320, 70543, 070172, 610127, 610133, 70169, 61208, 70313, 51067, 60174, 60215 60578, 046510106C, 70153, 60175, 046510106F, 46534905, 046520506B, 070164, 070633, 070376, 070375, 070538, 070724, 070376, 61264, 61212, 070855, 070854, 070830, 070863, 070726, 070628, 60542, 60573, 60574, 60659, 61262, 70312, 070379, 61216, 70312, 070628, 070726, 61264, 070723 070166, 070629, 070164, 070379, 070856, 60583, 61237, 61238, 606110, 70312, 070159, 70174, 70144, 070160, 70375, 070631, 070825, 070726, 070628, 070863, 61208, 70145, 610134, 70321, 61247, 70315, 70334, 070724 070720, 070602, 070326, 070602, 070631, 070325, 070719, 70123, 70160, 046502607B, 070165, 070826, 046505707B, 60434, 60433, 046510106E
Other devices under the same FDA submission
8 devices cleared under K012366, K022100, K040334, K043091 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Nexframe® DB-2040Medtronic, Inc. · Neurological stereotaxic Instrument1 recall5 reports
- NEXDRIVE™ MI-2000Medtronic, Inc. · Neurological stereotaxic Instrument1 recall1 report
- Nexprobe™ NP-1000Medtronic, Inc. · Neurological stereotaxic Instrument1 recall0 reports
- NEXDRIVE™ MI-1000Medtronic, Inc. · Neurological stereotaxic Instrument1 recall0 reports
- Nexdrive™ MI-1000Medtronic, Inc. · Neurological stereotaxic Instrument3 reports
- Nexdrive™ MI-2000Medtronic, Inc. · Neurological stereotaxic Instrument1 report
- Nexframe® DB-2040Medtronic, Inc. · Neurological stereotaxic Instrument1 report
- Nexframe DB-2040-00780Fhc, Inc. · Neurological stereotaxic Instrument1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page