Stryker Modular Replacement System Curved Cemented Stem recallZ-0926-2008 · Stryker Howmedica Osteonics Corp. · initiated Nov 15, 2007
Terminated All recalls posted in February 2008
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Howmedica Osteonics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Curvature Discrepancy -- The radius of curvature may be out of specification to varying degrees causing a discrepency in the stem curvature between trials and implants.
FDA root cause category: Process design
What the company is doing
On November 26, 2007, Stryker Sent Urgent Product Recall notification letters to Stryker Agencies/Branches and hospitals (OR Supervisor) by Federal Express. The notification requested that all inventory be examined and returned. Returning instructions were provided in the letter. Questions concerning this recall are being addressed by contacting the Regulatory Compliance Hotline at (201) 831-6633.
Recall details
- Recall number
- Z-0926-2008
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Nov 15, 2007
- Posted by the FDA
- Feb 26, 2008
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Quantity
- 803 total, all sizes
- Product code
- JDI
- FDA submissions
- K952970
- Distribution
- Nationwide Distribution
Product description
Stryker Modular Replacement System Curved Cemented Stem; MRS 15x127 Curved Femoral Stem. Catalog Number: 6485-3-715; Stryker Orthopaedics Mahwah, NJ 07430
Code information (models, lots, serial numbers)
Catalog Number: 6485-3-715; Lot Codes: LCDCF, TEC056A, TEC056B, TEC088B, TEC088B1, TEC088C, TEC088E, TEC088H, TEC141, TEC142, TEC258, TEC338.
Other devices under the same FDA submission
1 device cleared under K952970 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- MRS 6485-3-211Howmedica Osteonics Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 45947).
- Z-0917-2008 · TerminatedStryker Modular Replacement System Curved Cemented Stem; MRS SM Curved Femoral Stem 10x102; Catalog Number: 6485-3-300; Stryker Orthopaedics Mahwah, NJ 07430
- Z-0918-2008 · TerminatedStryker Modular Replacement System Curved Cemented Stem; MRS Small Curved Femoral Stem 8x102. Catalog Number 6485-3-308; Stryker Orthopaedics Mahwah, NJ 07430
- Z-0919-2008 · TerminatedStryker Modular Replacement System Curved Cemented Stem; MRS Small Curved Femoral Stem 9 x 102. Catalog Number: 6485-3-309; Stryker Orthopaedics Mahwah, NJ 07430
- Z-0920-2008 · TerminatedStryker Modular Replacement System Curved Cemented Stem; MRS 11mmx203mm femoral stem. Catalog Number: 6485-3-311; Stryker Orthopaedics Mahwah, NJ 07430
- Z-0921-2008 · TerminatedStryker Modular Replacement System Curved Cemented Stem; MRS 12mmx203 Femoral Stem. Catalog Number: 6485-3-313; Stryker Orthopaedics Mahwah, NJ 07430
- Z-0922-2008 · TerminatedStryker Modular Replacement System Curved Cemented Stem; MRS 15mmx203mm Femoral Stem. Catalog Number: 6485-3-315; Stryker Orthopaedics Mahwah, NJ 07430
- Z-0923-2008 · TerminatedStryker Modular Replacement System Curved Cemented Stem; MRS 17mmx203 Femoral Stem. Catalog Number 6485-3-317; Stryker Orthopaedics Mahwah, NJ 07430
- Z-0924-2008 · TerminatedStryker Modular Replacement System Curved Cemented Stem; MRS 11x127 Curved Femoral Stem. Catalog Number: 6485-3-711; Stryker Orthopaedics Mahwah, NJ 07430
- Z-0925-2008 · TerminatedStryker Modular Replacement System Curved Cemented Stem; MRS 13x127 Curved Femoral Stem. Catalog Number: 6485-3-713; Stryker Orthopaedics Mahwah, NJ 07430
- Z-0927-2008 · TerminatedStryker Modular Replacement System Curved Cemented Stem; MRS 17x127 Curved Femoral Stem. Catalog Number: 6485-3-717; Stryker Orthopaedics Mahwah, NJ 07430
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page