Skip to content

Seradyn QMS Vancomycin reagents, Part Number: 0373589.… recallZ-1102-2008 · Seradyn, Inc. · initiated Dec 20, 2007

Terminated All recalls posted in April 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Seradyn, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

False negative test results due to Interfering substances (heterophile antibodies) in the patient's blood may produce erroneously low results.

FDA root cause category: Device Design

What the company is doing

Consignees were notified of the problem via Urgent Medical Device Correction letter dated 12/20/07. The letter provided information on the issues concerning the recall. The firm will revise the package insert to address the potential occurrence of false negative bias. If inconsistencies with Vancomycin level exist, confirm with an alternative test method. For further information please contact 1-800-428-4072.

Recall details

Recall number
Z-1102-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Dec 20, 2007
Posted by the FDA
Apr 29, 2008
Recalling firm
Seradyn, Inc.
Quantity
1,819 kits plus bulk material for 1,863 kits.
Product code
LEH
FDA submissions
K050419
Distribution
Worldwide Distribution --- USA including states of California and New York, and countries of Australia, Belgium, Canada, Finland, Germany, Ireland, Japan, Spain and United Kingdom.

Product description

Seradyn QMS Vancomycin reagents, Part Number: 0373589. Seradyn, Inc, Indianapolis, IN.

Code information (models, lots, serial numbers)

All lots (lots 501930, 502063, 502074, 600432, 601166, 601207, 602556, 603076, 603801, 603312, 700657 and 703027).

Other devices under the same FDA submission

1 device cleared under K050419 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page