Skip to content

HOUVA III Phototherapy System with DermaSense Technology… recallZ-1309-2008 · National Biological Corp · initiated Feb 19, 2008

Terminated All recalls posted in August 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from National Biological Corp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software allows operator to override "low line voltage" error warning and store light intensity value.

FDA root cause category: Software design

What the company is doing

National Biological sent an "Important safety reminders for your Houva DermaSense phototherapy device" letters, dated 2/11/08 to users. It states that the user should collect a "Profile" of the light energy and record it on a daily/weekly flow chart. The chart should be in plain view so each operator can clearly determine the units output. If an error message appears in the display panel, treatment should not be continued unless the phototherapy booth operator was able to identify and correct the cause of the error. Make sure both breakers are in the "ON" position before starting up the unit each morning and before giving treatment.

Recall details

Recall number
Z-1309-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Feb 19, 2008
Posted by the FDA
Aug 28, 2008
Recalling firm
National Biological Corp
Quantity
48 Units
Product code
FTC
FDA submissions
K041212
Distribution
Worldwide distribution --- USA including states of: CA, DE, FL, GA, ID, IL, IN, MA, MI, MN, NC, NH, NY, OH, OR, PA, TX, VA, WA and WI, and country of Canada.

Product description

HOUVA III Phototherapy System with DermaSense Technology comes on one skid within 10 cardboard boxes. A therapeutic product designed for individuals who require specific Ultraviolet radiation therapy for diagnosed skin disorders.

Code information (models, lots, serial numbers)

Serial #'s: 10134, 10145, 10147, 10148, 10105, 10151, 10112, 10113, 10114, 10101, 10150, 10140, 10103, 10123, 10135, 10135, 10138, 10137, 10122

Other devices under the same FDA submission

1 device cleared under K041212 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page